Skip to content

AirLife™ 2K8004C2Recalls & reported problems

Airlife · Adult Manual Resuscitator •Oxygen Reservoir Bag, Adult Mask, CO2Detector · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 10, 2024Class IOpen, Classified

    Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

    Z-1060-2024 · Root cause: Process control · Recall details · FDA record ↗

Other Cardiopulmonary resuscitation aid kit devices

All Cardiopulmonary resuscitation aid kit devices →

Sources and updates