Skip to content

AirLife™ 2K8005Recalls & reported problems

Airlife · AirLife™ Manual Resuscitator Adult Manual Resuscitator 40 inch (1.0 M) Oxygen Reservoir Tubing, Adult Mask · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
1
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10190752114159), received between Jan 4, 2021 and Jun 21, 2024 — classified as 1 death, 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 10, 2024Open, Classified

    Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

    Z-1061-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem3
  • Gas/Air Leak1
  • Inflation Problem1

Patient problems

  • Cardiac Arrest2
  • Insufficient Information1
  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1
  • Respiratory Failure1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • DeathJun 21, 2024Mechanical Problem; Gas/Air Leak; Mechanical Problem; Gas/Air Leak

    RESUSCITATION BAG FAILED. PATIENT EXPIRED.

    Patient problems: Cardiac Arrest; Respiratory Failure

    MAUDE report 19588350 ↗
  • MalfunctionMay 10, 2024Mechanical Problem; Mechanical Problem

    RECIEVED A MEDWATCH REPORT FROM FDA STATING THAT THERE WAS A PRODUCT MALFUNCTION WITH A DEVICE THAT WAS INCLUDED IN A PRODUCT RECALL.

    Patient problems: Insufficient Information

    MAUDE report 19294142 ↗
  • InjuryMay 10, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE PATIENT EXPERIENCED A DROP IN OXYGEN LEVELS, PROMPTING THE USE OF AN BAG VALVE MASK FOR VENTILATION. HOWEVER, ATTEMPTS TO PROVIDE AIR FAILED EVEN AFTER REPOSITIONING THE PATIENT FOR BETTER AIRWAY ACCESS. SUBSEQUENTLY, THE PATIENT SUFFERED A CARDIAC ARREST, AND A NEW BAG WAS USED AND WORKED APPROPRIATELY. THE PROVIDER WAS ATTEMPTING TO TEST THE BAG, AND HEARD A "POP" WITH THE DEVICE, AND THEN IT FUNCTIONED AFTERWARD.

    Patient problems: Cardiac Arrest; Low Oxygen Saturation

    MAUDE report 16904110 ↗
  • InjuryJan 4, 2021Inflation Problem; Inflation Problem

    THE CUSTOMER REPORTED AN ISSUE ON AIRLIFE¿ ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT. PATIENT CODED POST PROCEDURE. THE NURSE STARTED TO BAG THE PATIENT AND THE MANUAL RESUSCITATOR FAILED TO RE-INFLATE AFTER VENTILATION. ANESTHESIA ARRIVED AND NOTED THE DEFECT AND THEY WERE ABLE TO ESTABLISH THE AIRWAY. THE PATIENT CONTINUED TO BE VENTILATED AND EVENTUALLY ESTABLISHED SPONTANEOUS RESPIRATIONS. SHE WAS ADMITTED OVER NIGHT FOR OBSERVATION AND WAS DISCHARGED HOME THE FOLLOWING DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11113405 ↗

Other Cardiopulmonary resuscitation aid kit devices

All Cardiopulmonary resuscitation aid kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page