AirLife™ 2K8005Recalls & reported problems
Airlife · AirLife™ Manual Resuscitator Adult Manual Resuscitator 40 inch (1.0 M) Oxygen Reservoir Tubing, Adult Mask · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10190752114159), received between Jan 4, 2021 and Jun 21, 2024 — classified as 1 death, 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Mechanical Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jan 10, 2024Open, Classified
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Z-1061-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem3
- Gas/Air Leak1
- Inflation Problem1
Patient problems
- Cardiac Arrest2
- Insufficient Information1
- Low Oxygen Saturation1
- No Clinical Signs, Symptoms or Conditions1
- Respiratory Failure1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathJun 21, 2024Mechanical Problem; Gas/Air Leak; Mechanical Problem; Gas/Air Leak
RESUSCITATION BAG FAILED. PATIENT EXPIRED.
Patient problems: Cardiac Arrest; Respiratory Failure
MAUDE report 19588350 ↗MalfunctionMay 10, 2024Mechanical Problem; Mechanical Problem
RECIEVED A MEDWATCH REPORT FROM FDA STATING THAT THERE WAS A PRODUCT MALFUNCTION WITH A DEVICE THAT WAS INCLUDED IN A PRODUCT RECALL.
Patient problems: Insufficient Information
MAUDE report 19294142 ↗InjuryMay 10, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE PATIENT EXPERIENCED A DROP IN OXYGEN LEVELS, PROMPTING THE USE OF AN BAG VALVE MASK FOR VENTILATION. HOWEVER, ATTEMPTS TO PROVIDE AIR FAILED EVEN AFTER REPOSITIONING THE PATIENT FOR BETTER AIRWAY ACCESS. SUBSEQUENTLY, THE PATIENT SUFFERED A CARDIAC ARREST, AND A NEW BAG WAS USED AND WORKED APPROPRIATELY. THE PROVIDER WAS ATTEMPTING TO TEST THE BAG, AND HEARD A "POP" WITH THE DEVICE, AND THEN IT FUNCTIONED AFTERWARD.
Patient problems: Cardiac Arrest; Low Oxygen Saturation
MAUDE report 16904110 ↗InjuryJan 4, 2021Inflation Problem; Inflation Problem
THE CUSTOMER REPORTED AN ISSUE ON AIRLIFE¿ ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT. PATIENT CODED POST PROCEDURE. THE NURSE STARTED TO BAG THE PATIENT AND THE MANUAL RESUSCITATOR FAILED TO RE-INFLATE AFTER VENTILATION. ANESTHESIA ARRIVED AND NOTED THE DEFECT AND THEY WERE ABLE TO ESTABLISH THE AIRWAY. THE PATIENT CONTINUED TO BE VENTILATED AND EVENTUALLY ESTABLISHED SPONTANEOUS RESPIRATIONS. SHE WAS ADMITTED OVER NIGHT FOR OBSERVATION AND WAS DISCHARGED HOME THE FOLLOWING DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11113405 ↗
Other Cardiopulmonary resuscitation aid kit devices
- Ambu® SPUR® II Pediatric Resuscitator 531600051Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall10 reports
- AirLife™ 2K8004Airlife · Cardiopulmonary resuscitation aid kit1 recall5 reports
- Ambu® SPUR® II Adult Resuscitator 523611051Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall3 reports
- AirLife™ 2K8017Airlife · Cardiopulmonary resuscitation aid kit1 recall1 death2 reports
- AirLife™ 2K8004FAirlife · Cardiopulmonary resuscitation aid kit1 recall1 report
- Ambu® SPUR® II Pediatric Resuscitator 530213000Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall0 reports
- Ambu® SPUR® II Adult Resuscitator 524611000Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall0 reports
- Ambu® SPUR® II Pediatric Resuscitator 530613831Ambu A/S · Cardiopulmonary resuscitation aid kit1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.