LTV 18888-001Recalls & reported problems

Vyaire Medical, Inc. · LTV 1200 Ventilator, English · Class 2

All models of LTV (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10190752115279), received between Sep 9, 2022 and Mar 20, 2026 — classified as 4 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem3
  • Unexpected Shutdown3
  • Failure to Power Up2
  • Calibration Problem1
  • Circuit Failure1
  • Erratic or Intermittent Display1
  • Failure to Cycle1
  • Failure to Run on Battery1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2026Tidal Volume Fluctuations

    IT WAS REPORTED THAT DURING USE ON A PATIENT, THE DEVICE ALARMED FOR LOW EXHALED TIDAL VOLUME. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24653560 ↗
  • MalfunctionMar 13, 2026Intermittent Loss of Power

    IT WAS REPORTED THAT THE DEVICE WAS INAPPROPRIATELY POWER CYCLING. THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590437 ↗
  • MalfunctionJan 9, 2026Failure to Power Up

    CUSTOMER STATED THAT THE UNIT WILL NOT TURN ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24020682 ↗
  • MalfunctionOct 29, 2025Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT THE VENTILATOR CIRCUIT DISCONNECTED WHILE ON A PATIENT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23412673 ↗
  • MalfunctionOct 2, 2025Failure to Cycle; Failure to Cycle

    IT WAS REPORTED THAT THE VENT IS NOT CYCLING, AND THE TURBINE IS NOT OPERATIONAL. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23199569 ↗
  • MalfunctionSep 15, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT BEFORE USE, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23059327 ↗
  • MalfunctionSep 11, 2025No Flow; No Flow

    IT WAS REPORTED THAT DURING A PATIENT TRANSPORT THE DEVICE "STOPPED" WHILE IN USE. THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23033548 ↗
  • MalfunctionJul 16, 2025Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THE DEVICE POWERED OFF AND TURNED BACK ON WHEN A "BUMP" WAS ENCOUNTERED DURING TWO PATIENT TRANSPORTS. COMPLAINANT REPORTED CLINICAL STAFF SUCCESSFULLY RETURNED THE DEVICE TO AN OPERATING STATE AND COMPLETED THE PATIENT TRANSPORTS. COMPLAINANT REPORTED NO ADVERSE EFFECTS OR PATIENT HARM OCCURRED DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22518634 ↗
  • MalfunctionJul 10, 2025Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22469896 ↗
  • MalfunctionJul 9, 2025Calibration Problem; Calibration Problem

    COMPLAINANT ALLEGED THAT BEFORE USE, THE DEVICE EXPERIENCED A HARDWARE FAULT ERROR. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22457652 ↗
  • MalfunctionJun 23, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    COMPLAINANT ALLEGED THAT THE DISPLAY IS FLICKERING AND APPEARS TO BE ATTEMPTING TO CYCLE POWER. THE DEVICE WAS TURNED OFF BY PRESSING AND HOLDING THE STANDBY/ON BUTTON FOR AN EXTENDED PERIOD. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT DURING THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 22312462 ↗
  • MalfunctionMay 30, 2025Failure to Run on Battery; Failure to Run on Battery

    COMPLAINANT ALLEGED THAT THE DEVICE WILL NOT RUN ON INTERNAL BATTERY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22115793 ↗
  • MalfunctionNov 20, 2024High Readings; High Readings

    IT WAS REPORTED THE VENT TIDAL VOLUMES RETURNING MUCH HIGHER THAN SET. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20734857 ↗
  • InjuryOct 11, 2024Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENT WAS ON A PATIENT AND SUDDENLY STOPPED WORKING / SHUT OFF COMPLETELY. THE BATTERY WAS FULLY CHARGED. THERE WAS NO PATIENT HARM REPORTED, AS STAFF WAS NEAR THE PATIENT'S BEDSIDE DURING INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20433368 ↗
  • MalfunctionJul 5, 2024Circuit Failure; Circuit Failure

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENTILATOR HAD INOP LIGHT ON WHILE DEVICE IS POWERED ON. DEVICE WAS NOT ON A PATIENT AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19679504 ↗
  • MalfunctionJun 18, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 1200 VENTILATOR'S INSP/EXP HOLD BUTTON WAS NOT WORKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19558097 ↗
  • InjuryJun 13, 2024Unintended System Motion; Unintended System Motion

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENTILATOR WAS AUTO-CYCLING IN SIMV/CPAP MODE 10/6 ON A PATIENT. THE PATIENT WAS TRANSFERRED TO A DIFFERENT DEVICE NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19527823 ↗
  • InjuryApr 26, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL THE LTV VENTILATOR ""SHUT DOWN"" ON A PATIENT. THE RT'S REPORTED THAT IT SHUT DOWN AND THEY STATE IT DID NOT ALARM. THEY STATED THE PATIENTS O2 SATS WENT DOWN TO 37% WHEN THEY NOTICED IT. THE PATIENT WAS TRANSFERRED TO ANOTHER VENTILATOR. NO PATIENT HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19196436 ↗
  • InjuryMar 28, 2024No Pressure; No Pressure

    IT WAS REPORTED TO VYAIRE MEDICAL THE PRESSURE SUPPORT IS "NOT WORKING" ON THIS UNIT. THE UNIT WAS ON A PATIENT, NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18992314 ↗
  • MalfunctionSep 24, 2022Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LAPTOP VENTILATOR SPRINTPACK BATTERY WAS NO LONGER CHARGING, AND THAT THE VENT INOPERABLE LIGHT WAS ON, THE CHARGE STATUS LIGHT WAS BLINKING ORANGE, AND NO LIGHTS CAME ON WHEN THE EXTERNAL BATTERY WAS TESTED. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15481440 ↗
  • MalfunctionSep 9, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 1200 VENTILATOR HAS BLENDER CLICKING WITH FIO2 ISSUE. THIS EVEN HAPPENED DURING PATIENT USE AND NO PATIENT HARM NOTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15397161 ↗
  • MalfunctionSep 9, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 1200 VENTILATOR HAS HARDWARE FAULT ALARM. THIS EVEN HAPPENED DURING PATIENT USE AND NO PATIENT HARM NOTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15395889 ↗

Other Ventilator, continuous, facility use devices

All Ventilator, continuous, facility use devices →