LTV 1150 VENTILATOR 18984-001 (UDI 10190752115644)Recalls & reported problems

Vyaire Medical, Inc. · LTV 1150 Ventilator, English · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10190752115644), received between Jun 1, 2022 and Jun 23, 2025 — classified as 3 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Display Difficult to Read1
  • Display or Visual Feedback Problem1
  • Failure to Cycle1
  • Failure to Power Up1
  • Inaccurate Flow Rate1
  • Infusion or Flow Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Display Difficult to Read; Display Difficult to Read

    COMPLAINANT ALLEGED THAT THE LED LIGHTS ARE GOING OUT AND BLANKING OUT ON THE DEVICE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22310285 ↗
  • MalfunctionApr 1, 2025Noise, Audible; Noise, Audible

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE WAS MAKING AN UNUSUAL NOISE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 21737621 ↗
  • MalfunctionJun 19, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE DISPLAY HAD A FAULTY LED. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19571280 ↗
  • InjuryJun 7, 2024Use of Incorrect Control/Treatment Settings; Inaccurate Flow Rate; Use of Incorrect Control/Treatment Settings; Inaccurate Flow Rate

    IT WAS REPORTED TO VYAIRE MEDICAL THE LTV VENTILATOR PRESENTED A LOW PRESSURE ALARM. THE PATIENT WENT INTO CARDIAC ARREST. CPR WAS INITIATED AND RETURN OF SPONTANEOUS CIRCULATION WAS ACHIEVED. THE PATIENT WAS SWITCHED OVER TO ANOTHER VENTILATOR. NO INJURY OR HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19488848 ↗
  • InjuryApr 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENTILATOR IS ON A PATIENT AND JUST STARTED BEEPING EVERY 20 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19105223 ↗
  • MalfunctionNov 28, 2023Failure to Cycle; Failure to Cycle

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP VENTILATOR 1150 HAS AUTO-CYCLING. THE CUSTOMER CONFIRM THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18223284 ↗
  • MalfunctionNov 14, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT HIGH RATES OCCURRED DURING USE ON A PATIENT ON THE LAP TOP VENTILATOR 1150.IT WOULD JUMP FROM 14 TO 20, 30 AND BACK DOWN. THE SETTINGS ARE A/C (ASSIST-CONTROL) RATE 14, VT (TIDAL VOLUME) 500 CC, PEEP (POSITIVE END-EXPIRATORY PRESSURE) 5 AND SENSITIVITY IS 1.0. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18132573 ↗
  • MalfunctionNov 1, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BPM (BREATHS PER MINUTE) READING IS 22 WHEN IT IS SET TO 12 ON THE LAP TOP VENTILATOR 1150.THE VCALC IS ALSO READING LOW TO WHAT THEY ARE USED TO SEEING 27 LPM. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18052229 ↗
  • MalfunctionAug 29, 2023Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THERE WAS A PATIENT ON A VERY HIGH SENSITIVITY AT 9, A/C, P/C RR 25, PEEP 8 VT 250ML ON THE LTV 1500 VENTILATOR. WHEN THEY MOVE THE PATIENT TO A STROLLER AND PLACE THE UNIT ON THE PATIENT, THEY NEED TO TURN DOWN THE SENSITIVITY TO ABOUT 7-8 SO SHE CAN TAKE A BREATH. THEY SWITCH OVER IS VERY QUICK MAYBE 30 SECONDS. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17644109 ↗
  • MalfunctionAug 18, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THERE WAS A VOLUME ISSUE ON THE LTV 1150 VENTILATOR. IT WAS CONFIRMED BY THE CUSTOMER THAT THE VOLUME WAS LOW. MOREOVER, THE FLOW RATE WAS OFF. THERE IS NO INFORMATION OF PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT AS OF THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17574186 ↗
  • MalfunctionJun 20, 2022Temperature Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THE LTV 1150 WAS RUNNING HOT WITH ODD SQUEAK PRESENT AND HAD A PLASTIC ODOR. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14749713 ↗
  • InjuryJun 1, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER REPORTED PROBLEM WITH LAP TOP VENTILATOR 1150 WHERE INTERNAL BATTERY FAILED CAUSING UNIT TO POWER DOWN WHILE TRANSPORTING A PATIENT FROM ER TO ICU. NO PATIENT HARM REPORTED. A STAFF IMMEDIATELY NOTED THE CONDITION AND BEGAN BAGGING THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14566854 ↗

Other Ventilator, continuous, facility use devices

All Ventilator, continuous, facility use devices →