LTV 2200 VENTILATOR 22690-001Recalls & reported problems
Vyaire Medical, Inc. · LTV2 2200 Ventilator · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
The LTV 2200 ventilator from Vyaire Medical is a portable machine used to keep patients breathing. The 27 reports mostly describe hardware and mechanical trouble — loud fan or turbine noise, false fault and "check circuit" alarms, gas leaks, and oxygen blender readings that drift or read low, plus display screen failures. Six reports involve injuries; no deaths are described. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10190752139220), received between Apr 12, 2021 and Nov 21, 2025 — classified as 6 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem16
- Incorrect Measurement4
- Connection Problem1
- Display or Visual Feedback Problem1
- Failure to Calibrate1
- Gas/Air Leak1
- Noise, Audible1
- Overfill1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING USE ON A PATIENT (AGE AND GENDER UNKNOWN), THE DEVICE EXPERIENCED A CONSISTENT HARDWARE FAULT ALARM. IT WAS CONFIRMED THAT THE DEVICE CONTINUED TO VENTILATE DURING THE REPORTED ALARM. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23623831 ↗MalfunctionMar 24, 2025Failure to Calibrate; Failure to Calibrate
COMPLAINANT ALLEGED THAT THE DEVICE IS OVER TRIGGERING WITH A "CHECK CIRCUIT " ALARM MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Patient problems: Insufficient Information
MAUDE report 21678934 ↗MalfunctionOct 16, 2024Overfill; Overfill
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE FAN AND TURBINE WAS REPLACED, BUT THE TURBINE SOUND IS STILL THERE. THE SOUND IS COMING FROM THE TURBINE INPUT AS A BACK FLOW. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20464627 ↗InjuryAug 28, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT SUDDENLY THE UNIT STARTED HAVING DEAD PIXELS ON THE LED. THE DEVICE WAS ON A PATIENT AT THE TIME OF THE ISSUE. NO PATIENT HARM WAS REPORTED. UNIT WAS REMOVED FROM PATIENT DUE TO ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20092000 ↗InjuryApr 26, 2024Noise, Audible; Noise, Audible
IT WAS REPORTED TO VYAIRE MEDICAL THAT "SUDDENLY THE UNIT STARTED HAVING HIGH FAN NOISE WHILE ON PATIENT WHICH WORRIED THE USER OF A FAULT OR OVERHEATING WHICH RESULTED IN REMOVING THE UNIT. THE UNIT IS NEW WITH MINIMUM USAGE." SWAPPED OUT DEVICE FOR DIFFERENT UNIT. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19193452 ↗MalfunctionApr 9, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV2 VENTILATOR AND THE BLENDER IS NOT WORKING PROPERLY ANYMORE. LOW FOI2 READINGS. WHEN SETTING O2 AT 100% THE UNIT GIVES MAX OF 70% FIO2 TOGETHER WITH THE IMT AND ALL THE OTHER READINGS ARE OUT. TRIED CALIBRATION OF THE O2 INLET BUT WE COULDN'T RESOLVE THE ISSUE. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19071890 ↗MalfunctionJan 12, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 2200 TIDAL VOLUMES ARE READING IN THE 2000 TO 3000 RANGE. AT THIS TIME, THERE WAS NO INFORMATION FOR PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18503923 ↗MalfunctionJan 11, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 2200 O2 BLENDING PERCENTAGE HAD FAILED AT 90%.THE RESULT WAS 96.1% WHILE THE MAXIMUM SPECIFICATION IS 95%. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18499832 ↗MalfunctionJan 11, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 2200 VTI (INHALED TIDAL VOLUME) TEST FAILED. THE CALCULATED MINIMUM SPECIFICATION WAS 265 BUT THE RESULT WAS 245. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18493701 ↗MalfunctionJan 4, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP VENTILATOR 220 HAS HIGH SIDE/LOW SIDE LEAK FAILURE. THE CUSTOMER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18446478 ↗MalfunctionNov 2, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT FRACTION OF INSPIRED OXYGEN VIA EXTERNAL ANALYZER WAS MUCH LOWER THAN THE FRACTION OF INSPIRED OXYGEN THAT WAS SET ON THE LAP TOP VENTILATOR 2200. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18055650 ↗MalfunctionNov 2, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL PROBLEM WITH THE LAPTOP 2200 VENTILATOR IN WHICH THE FIO2 (FRACTION OF INSPIRED OXYGEN) VIA EXTERNAL ANALYZER WAS MUCH LOWER THAN THE FIO2 THAT WAS SET ON THE VENT. FURTHERMORE, THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18055534 ↗InjuryNov 2, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL PROBLEM WITH THE LAPTOP 2200 VENTILATOR IN WHICH THE FIO2 (FRACTION OF INSPIRED OXYGEN) VIA EXTERNAL ANALYZER WAS MUCH LOWER THAN THE FIO2 THAT WAS SET ON THE VENT DURING PATIENT USE. FURTHERMORE, THERE WAS NO PATIENT HARM REPORTED ASSOCIATED WITH THE EVENT. UNIT WAS SWAPPED OUT FOR A DIFFERENT DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18055442 ↗MalfunctionNov 2, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT FRACTION OF INSPIRED OXYGEN VIA EXTERNAL ANALYZER WAS MUCH LOWER THAN THE FRACTION OF INSPIRED OXYGEN THAT WAS SET ON THE LAP TOP VENTILATOR 2200. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18055374 ↗InjuryNov 2, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 2200 VENTILATOR FIO2 VIA EXTERNAL ANALYZER WAS MUCH LOWER THAN THE FIO2 WAS SET. THIS HAPPENED DURING PATIENT USE. THE VENTILATOR WAS SWAPPED OUT FOR A DIFFERENT DEVICE. THERE WAS NO PATIENT HARM REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18055325 ↗MalfunctionNov 1, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT TIDAL VOLUME INSPIRED (VTI) WAS MEASURED TO BE BELOW THE ACCEPTABLE RANGE ON THE LAP TOP VENTILATOR 2200. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18049631 ↗MalfunctionNov 1, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LTV 2200 VTI (INHALED TIDAL VOLUME) WAS MEASURED AT 255 ML, WHICH WAS BELOW THE PASSING RANGE OF 265-335 ML. AT THIS TIME, THERE WAS NO INFORMATION FOR PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18049498 ↗MalfunctionNov 1, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP VENTILATOR 2200 VTI (INHALED TIDAL VOLUME) WAS MEASURED TO BE 252 ML, BELOW THE ACCEPTABLE RANGE OF 265-335 ML. AS OF THIS TIME, THERE IS NO INFORMATION OF PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18049329 ↗MalfunctionNov 1, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT TIDAL VOLUME INSPIRED (VTI) WAS MEASURED TO BE BELOW THE ACCEPTABLE RANGE ON THE LAP TOP VENTILATOR 2200. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18048991 ↗InjuryMay 4, 2022Pressure Problem; Pressure Problem
THE CUSTOMER REPORTED THE PRESSURE ON THIS UNIT KEPT RISING, THEN WOULD DUMP AND BACK AND FORTH DURING USE ON A PATIENT ON THE LAP TOP VENTILATOR 2200. THE CUSTOMER REPORTED THE UNIT WAS SWAPPED OUT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14281037 ↗InjuryApr 13, 2022Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED THAT THE PEEP (POSITIVE END-EXPIRATORY PRESSURE) KEPT DROPPING OUT AND THEN COMING BACK UP CONSTANTLY ON THE LTV 2200 LAP TOP VENTILATOR DURING PATIENT USE. THE DEVICE WAS REPLACED WITH ANOTHER VENTILATOR AND THERE WAS NO PATIENT HARM REPORTED ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14107468 ↗MalfunctionJan 10, 2022Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP 2200 VENTILATOR WAS HAVING CONTINUOUS CHECK CIRCUIT ISSUE. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13213186 ↗MalfunctionDec 30, 2021Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT LTV 2200 VENTILATOR IS CONTINUOUSLY ALARMING "CHECK CIRCUIT". THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT AND THE REPORTED ISSUE OCCURRED DURING SET-UP BEFORE PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13119452 ↗MalfunctionNov 23, 2021Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP VENTILATOR 2200 ALARMED CHK (CHECK) CIRCUIT. THE PEEP (POSITIVE END EXPIRATORY PRESSURE) AND INSPIRATORY PRESSURE ARE NOT ACCORDING CONTROL. THE CUSTOMER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12860559 ↗MalfunctionNov 23, 2021Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LAP TOP VENTILATOR 2200 ALARMED CHK (CHECK) CIRCUIT. THE PEEP (POSITIVE END EXPIRATORY PRESSURE) AND INSPIRATORY PRESSURE ARE NOT ACCORDING CONTROL. AS OF THIS TIME, THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12860522 ↗
Other Ventilator, continuous, facility use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page