VELA 16600 (UDI 10190752154230)Recalls & reported problems

Vyaire Medical, Inc. · VELA DIAMOND PACKAGE FOR KINDRED HEALTHCARE · Class 2

Quick summary

This VYAIRE VELA ventilator is a facility-use machine that delivers continuous breathing support. The 30 reports mostly describe calibration trouble, no output or no display, and low or missed breath volume, with the machine alarming or failing to cycle at times. Two injuries were reported and no deaths. No recalls name this device, and report numbers were the same in the last two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10190752154230), received between Dec 30, 2021 and Oct 2, 2024 — classified as 2 injuries, 28 malfunctions.

The most frequently reported specific device problem is “Failure to Calibrate” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 12 reportsMay 2024: 1 reportsJun 2024: 6 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Calibrate9
  • Failure to Recalibrate3
  • No Device Output3
  • No Display/Image3
  • Failure to Deliver2
  • Gas Output Problem2
  • Low Readings2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2024Low Readings; Low Readings

    IT WAS REPORTED TO VYAIRE MEDICAL THAT LOW VOLUME WHEN SET AT 500, BUT UNIT IS GIVING 227. UNKNOWN PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20352833 ↗
  • MalfunctionOct 2, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA HAD PERSISTENT XDCR (TRANSDUCER) FAULT. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20352477 ↗
  • MalfunctionSep 17, 2024Low Readings; Low Readings

    IT WAS REPORTED TO VYAIRE MEDICAL LOW VTE AND LEAK. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20237937 ↗
  • MalfunctionSep 13, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL THAT VENT INOP/MOTOR FAULT ALARMED, AND THE DEVICE WAS NOT VENTILATING. UNKNOWN PATIENT INVOLVEMENT

    Patient problems: Insufficient Information

    MAUDE report 20212904 ↗
  • InjurySep 11, 2024No Display/Image; No Display/Image

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENT HAD NO DISPLAY AND NOT VENTING. WHEN TURNING ON THE VENT, THE DISPLAY FLASHES ON THEN GOES BLACK. PATIENT INVOLVED. HOWEVER, NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20193057 ↗
  • MalfunctionJul 9, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL: A MOTOR FAULT DEFECTIVE MOTOR BOARD WAS NOTED ON THE VELA VENTILATOR. CURRENTLY, PATIENT INVOLVEMENT IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19701273 ↗
  • MalfunctionJun 20, 2024Gas Output Problem; Gas Output Problem

    IT WAS REPORTED TO VYAIRE MEDICAL, THE CUSTOMER WAS UNABLE TO ACHIEVE SUFFICIENT RESPIRATORY VOLUME; ALTHOUGH THE TURBINE WAS RUNNING, THE RESPIRATORY VOLUME WAS TOO LOW. NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19578241 ↗
  • MalfunctionJun 20, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL, THE CUSTOMER IS REQUESTING A PO FOR A MAIN BOARD AND TURBINE DUE TO TRANSDUCER FAULT CAUSING THE VELA VENTILATOR TO NOT CYCLE. ORDER WAS PROCESSED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19577090 ↗
  • MalfunctionJun 20, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL, THE EXHALATION DIFF TRANSDUCER WAS FAILING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19577016 ↗
  • MalfunctionJun 13, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL THAT ONSITE STAFF REPORTED MOTOR FAULT ALARM. CURRENTLY, NO PATIENT INVOLVEMENT WAS REPRORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19529167 ↗
  • MalfunctionJun 13, 2024Failure to Recalibrate; Failure to Recalibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THERE WAS AN ISSUE WITH THE VENTILATORS MAIN BOARD WHERE THE XDCR WAS FAILING CALIBRATION. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19527811 ↗
  • MalfunctionJun 11, 2024Gas Output Problem; Gas Output Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VENTILATOR WAS EXPERIENCING A LOW INLET PRESSURE ISSUE. THE INITIAL REPORTER WAS INFORMED THAT THE REGULATOR NEEDS TO BE ADJUSTED OR REPLACED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19509763 ↗
  • MalfunctionMay 1, 2024Failure to Recalibrate; Failure to Recalibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR WAS ALARMING VENT INOPERABLE AND MOTOR FAULT ALARM. THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19224724 ↗
  • MalfunctionApr 29, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE THAT A TRANSDUCER FAULT OCCURRED. NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19209047 ↗
  • MalfunctionApr 19, 2024Failure to Deliver; Failure to Deliver

    IT WAS REPORTED TO VYAIRE MEDICAL THAT 'VENT INOP' ALARM AND MOTOR FAULT ENTRY ON ERROR LOG. DETERMINED FAILURE OF MAIN BOARD. NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19146722 ↗
  • MalfunctionApr 18, 2024No Display/Image; No Display/Image

    IT WAS REPORTED TO VYAIRE MEDICAL THE VENT IS NOT VENTILATING WHEN BOOTING UP AND REPLACED A NEW POWER SUPPLY WHICH DID NOT RECTIFY THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139727 ↗
  • MalfunctionApr 18, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT A NEW MAIN BOARD WAS ORDERED DUE TO CONSTANT XDCR FAULT CAUSING VENT NOT TO CYCLE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19133521 ↗
  • MalfunctionApr 18, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT A NEW MAIN BOARD WAS ORDERED DUE TO CONSTANT XDCR FAULT CAUSING VENT NOT TO CYCLE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19133493 ↗
  • MalfunctionApr 16, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL "TRANSDUCER FAULT". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19120764 ↗
  • MalfunctionApr 16, 2024No Display/Image; No Display/Image

    IT WAS REPORTED TO VYAIRE MEDICAL THE DISPLAY ON THE VELA VENTILATOR IS BLANK. NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19116658 ↗
  • MalfunctionApr 11, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT A TRANSDUCER FAULT (A0801 FAILURE TO CALIBRATE) OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19092251 ↗
  • MalfunctionApr 11, 2024Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR PRODUCED A TRANSDUCER FAULT. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19088419 ↗
  • MalfunctionApr 10, 2024Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED TO VYAIRE THAT THE VELA CIRCUIT DISCONNECTED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19079809 ↗
  • MalfunctionApr 2, 2024Power Problem; Power Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENTILATOR INOP (INOPERABLE) ALARMED WHEN THE DEVICE WAS POWERED ON. A MOTOR FAULT ENTRY WAS FOUND ON THE ERROR LOG. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19024035 ↗
  • MalfunctionApr 2, 2024Failure to Recalibrate; Failure to Recalibrate

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE VELA VENT CAME TO THE BIOMED SHOP AND IS ALARMING O2 SENSOR FAILURE CONSTANTLY. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19020434 ↗

Other Ventilator, continuous, facility use devices

All Ventilator, continuous, facility use devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.