Avea 32029-001Recalls & reported problems
Vyaire Medical, Inc. · AVEA II Ventilator - Avea Refresh · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10190752186248), received between Jun 5, 2024 and Sep 13, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).
15 other recalls cover devices cleared under the same FDA submission (K103211); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement2
- Erratic or Intermittent Display1
- High Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 13, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT AFTER REPLACING THE O2 SENSOR, THE VENT IS ALARMING HIGH FIO2 WHILE THE VENT WAS ON A PATIENT. THE VENT WAS SWAPPED OUT. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20212490 ↗MalfunctionAug 28, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE AVEA HAD HIGH PPEAK ALARM DURING PM (PREVENTIVE MAINTENANCE). NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20088955 ↗MalfunctionAug 23, 2024High Readings; High Readings
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE AVEA SHOWED FIO2 ALARMS. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20050207 ↗MalfunctionJun 5, 2024Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE UIM (USER INTERFACE MODULE) IS FLICKERING, AND BUTTONS DON'T WORK. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19470346 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103211. One submission often covers many models, so check the product and lot details in each record.
- Z-1891-2016 · May 17, 2016 · TerminatedAVEA Comprehensive ventilator-refurbished, R17310-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering
- Z-1892-2016 · May 17, 2016 · TerminatedAVEA Standard ventilator-refurbished, R17311-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a pred
- Z-1894-2016 · May 17, 2016 · TerminatedTCA Board, 16542A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the brea
- Z-1895-2016 · May 17, 2016 · TerminatedGas Delivery Engine (GDE)-1st Generation, 16222-001-99 A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predeterm
- Z-1888-2016 · May 17, 2016 · TerminatedAVEA Comprehensive ventilator, 17310-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predeter
- Z-1889-2016 · May 17, 2016 · TerminatedAVEA Standard ventilator, 17311-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined
- Z-1896-2016 · May 17, 2016 · TerminatedGas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of ox
- Z-1898-2016 · May 17, 2016 · TerminatedAVEA GDE/UIM upgrade kit, 12283-PMN A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of
- Z-1890-2016 · May 17, 2016 · TerminatedAVEA Standard with Compressor ventilator, 17312-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering
- Z-1893-2016 · May 17, 2016 · TerminatedAVEA Standard with Compressor ventilator-refurbished, R17312-xx (xx = 0-14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by d
- Z-1899-2016 · May 17, 2016 · TerminatedAVEA GDE upgrade kit, 12284-PMN A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxyg
- Z-1897-2016 · May 17, 2016 · TerminatedGas Delivery Engine (GDE)-refurbished, R16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined per
- Z-1609-2015 · Apr 3, 2015 · TerminatedCareFusion AVEA ventilator, all models. Product Usage: A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predeterm
- Z-2273-2013 · Sep 5, 2013 · TerminatedProduct Name: AVEA Ventilator. All Models. The affected devices are cleared under K073069 pm 1/25/2008 and K103211 on 05/06/2011. This device is a Class II medical Device. A cont
- Z-0510-2012 · Sep 8, 2011 · TerminatedAVEA Ventilator (Catalog Code 17310, 17311, 17312). The AVEA is intended to provide continuous respiratory support in an institutional health care environment (e.g. hospitals). It
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page