Medline UROT1058Recalls & reported problems
Medline Industries, Inc. · 16FR 100% SIL TEMP CATH W/UM · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10193489604931), received between Aug 8, 2025 and Sep 16, 2026 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Connection Problem3
- Break1
- Disconnection1
- Fluid/Blood Leak1
- Incomplete or Inadequate Connection1
- Loose or Intermittent Connection1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
- Unspecified Mental, Emotional or Behavioral Problem1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Disconnection
Patient problems: Unspecified Mental, Emotional or Behavioral Problem
MAUDE report 50789026 ↗MalfunctionSep 9, 2026Fluid/Blood Leak; Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50694626 ↗MalfunctionAug 12, 2026Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50226665 ↗InjuryMar 25, 2026Connection Problem; Incomplete or Inadequate Connection
THE ALL-IN-ONE URINARY CATHETER KIT WAS IN USE IN THE PATIENT. IN THIS KIT, THE URINARY CATHETER, TUBING, AND BAG ALL COME PRE-CONNECTED AND THE JUNCTION BETWEEN THE CATHETER AND TUBING IS SECURED WITH A TAPE. IN THIS CASE, THE TUBING BECAME DISCONNECTED FROM THE CATHETER WHILE THE TAPE REMAINED INTACT. PERHAPS THE ADHESIVE BROKE DOWN, THOUGH THAT IS UNCLEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690393 ↗MalfunctionMar 2, 2026Break; Connection Problem; Material Integrity Problem
ACCORDING TO THE FACILITY, ON (B)(6), OBSERVED FOLEY IN PIECES ON THE BED, DRAIN INTACT IN PATIENT AND TUBING DISCONNECTED. REPLACED FOLEY.
Patient problems: Insufficient Information
MAUDE report 24485954 ↗InjuryNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE FACILITY ON (B)(6) 2025 "THE TUBING SLID AWAY FROM THE YELLOW TAPE, THE TAPE REMAINED INTACT AND CONNECTED TO THE CATHETER".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23493732 ↗InjuryNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE FACILITY ON (B)(6) 2025 "THE TUBING SLID AWAY FROM THE YELLOW TAPE, THE TAPE REMAINED INTACT AND CONNECTED TO THE CATHETER".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23490168 ↗InjuryNov 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE FACILITY ON 9/10/2025, "THE TUBING SLID AWAY FROM THE YELLOW TAPE, THE TAPE REMAINED INTACT AND CONNECTED TO THE CATHETER".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23489016 ↗InjuryNov 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE FACILITY ON (B)(6) 2025 "THE TUBING SLID AWAY FROM THE YELLOW TAPE, THE TAPE REMAINED INTACT AND CONNECTED TO THE CATHETER".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23451999 ↗MalfunctionAug 8, 2025Connection Problem; Connection Problem
FOLEY CATHETER DISCONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22754646 ↗
Other Catheter care tray devices
- Centurion Medical Products Corp CVI680CMedline Industries, Inc. · Catheter care tray2 recalls0 reports
- Medline Industries, Inc. DYNJ81597Medline Industries, Inc. · Catheter care tray2 recalls0 reports
- Bardia® Foley Catheter Insertion Tray 802010C. R. Bard, Inc. · Catheter care tray54 reports
- Medline DYND10160Medline Industries, Inc. · Catheter care tray1 recall4 reports
- Medline DYNC1816Medline Industries, Inc. · Catheter care tray1 recall2 reports
- SureStep Intermittent Catheter Tray INTAAIC. R. Bard, Inc. · Catheter care tray1 recall1 report
- Medline DYND10407Medline Industries, Inc. · Catheter care tray1 recall1 report
- MEDLINE URO4RR16CMedline Industries, Inc. · Catheter care tray1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page