Skip to content

MEDLINE DYNDTB1545Recalls & reported problems

Medline Industries, Inc. · IV ADMIN SET, 15DRP, 3 NF, 109 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10197344098644), received between Sep 21, 2026 and Sep 29, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Backflow” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 9 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 31, 2025Open, Classified

    Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.

    Z-1366-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Backflow5
  • Air/Gas in Device2
  • Connection Problem1
  • Fluid/Blood Leak1
  • Loss of or Failure to Bond1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Air/Gas in Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939292 ↗
  • MalfunctionSep 28, 2026Air/Gas in Device

    It was reported that "It collapses easier at the chamber and they get more air bubbles when using the port."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914711 ↗
  • MalfunctionSep 24, 2026Backflow

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50885960 ↗
  • MalfunctionSep 23, 2026Backflow

    It was reported that "They are having bad back flow problems".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872129 ↗
  • MalfunctionSep 22, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859552 ↗
  • MalfunctionSep 22, 2026Backflow

    It was reported that "Fluid is backing up in the tubing."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855688 ↗
  • MalfunctionSep 21, 2026Backflow; Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848004 ↗
  • MalfunctionSep 21, 2026Backflow; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50847252 ↗
  • MalfunctionSep 21, 2026Fluid/Blood Leak

    It was reported that ¿The new version leaks a lot from the part I circled in the pic. It seems flimsy where as the old one was solid and didn¿t leak.also the DYNDTB1545 seem to be flimsier.¿

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50843147 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →