Medline Industries IV Administration and Extension Sets recallZ-1366-2026 · Medline Industries, LP · initiated Dec 31, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
FDA root cause category: Process control
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 12/31/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-262 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1366-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 31, 2025
- Posted by the FDA
- Feb 13, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 926,227 units
- Product code
- FPA
- FDA submissions
- K242126
- Distribution
- Worldwide distribution - US Nationwide and the country of Bahamas.
Product description
Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1) DYNDTB0529, UDI-DI: 10197344105663(each), 20197344105660(case), Lot Number: All Lot Numbers; 2) DYNDTB1029, UDI-DI: 10197344076956(each), 20197344076953(case), Lot Number: All Lot Numbers; 3) DYNDTB1516, UDI-DI: 10197344105687(each), 20197344105684(case), Lot Number: All Lot Numbers; 4) DYNDTB1530, UDI-DI: 10197344105694(each), 20197344105691(case), Lot Number: All Lot Numbers; 5) DYNDTB1539, UDI-DI: 10197344008254(each), 20197344008251(case), Lot Number: All Lot Numbers; 6) DYNDTB1545, UDI-DI: 10197344098644(each), 20197344098641(case), Lot Number: All Lot Numbers; 7) DYNDTB2045, UDI-DI: 10197344077007(each), 20197344077004(case), Lot Number: All Lot Numbers; 8) DYNDTB0512, UDI-DI: 10197344105656(each), 20197344105653(case), Lot Number: All Lot Numbers; 9) DYNDTB0555, UDI-DI: 10197344105670(each), 20197344105677(case), Lot Number: All Lot Numbers; 10) DYNDTB1022, UDI-DI: 10197344076949(each), 20197344076946(case), Lot Number: All Lot Numbers; 11) DYNDTB1512, UDI-DI: 10197344098613(each), 20197344098610(case), Lot Number: All Lot Numbers; 12) DYNDTB1529, UDI-DI: 10197344098620(each), 20197344098627(case), Lot Number: All Lot Numbers; 13) DYNDTB1537, UDI-DI: 10197344008247(each), 20197344008244(case), Lot Number: All Lot Numbers; 14) DYNDTB1540, UDI-DI: 10197344098637(each), 20197344098634(case), Lot Number: All Lot Numbers; 15) DYNDTB2029, UDI-DI: 10197344076994(each), 20197344076991(case), Lot Number: All Lot Numbers; 16) DYNDTB5077, UDI-DI: 10197344098651(each), 20197344098658(case), Lot Number: All Lot Numbers; 17) DYNDTB1045, UDI-DI: 10197344076963(each), 20197344076960(case), Lot Number: All Lot Numbers; 18) DYNDTB1545D, UDI-DI: 10197344105700(each), 20197344105707(case), Lot Number: All Lot Numbers; 19) DYNDTB2524, UDI-DI: 10197344105717(each), 20197344105714(case), Lot Number: All Lot Numbers; 20) DYNDTB1012, UDI-DI: 10197344076932(each), 20197344076939(case), Lot Number: All Lot Numbers; 21) DYNDTB2012, UDI-DI: 10197344076970(each), 20197344076977(case), Lot Number: All Lot Numbers; 22) DYNDTB2022, UDI-DI: 10197344076987(each), 20197344076984(case), Lot Number: All Lot Numbers; 23) DYNDTB5081, UDI-DI: 10197344098668(each), 20197344098665(case), Lot Number: All Lot Numbers
Devices named in this recall
23 devices in the FDA device registry whose identifier appears in the recall's code information.
- MEDLINE DYNDTB1545Medline Industries, Inc. · Set, administration, intravascular1 recall9 reports
- MEDLINE DYNDTB1540Medline Industries, Inc. · Set, administration, intravascular1 recall5 reports
- MEDLINE DYNDTB5081Medline Industries, Inc. · Set, administration, intravascular1 recall3 reports
- MEDLINE DYNDTB5077Medline Industries, Inc. · Set, administration, intravascular1 recall2 reports
- MEDLINE DYNDTB1529Medline Industries, Inc. · Set, administration, intravascular1 recall1 report
- Medline DYNDTB1537Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- Medline DYNDTB1539Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1012Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1022Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1029Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1045Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- Medline DYNDTB2012Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB2022Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB2029Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB2045Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1512Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB0512Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB0529Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB0555Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1516Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1530Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB1545DMedline Industries, Inc. · Set, administration, intravascular1 recall0 reports
- MEDLINE DYNDTB2524Medline Industries, Inc. · Set, administration, intravascular1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page