Skip to content

MEDLINE DYNDTB5077Recalls & reported problems

Medline Industries, Inc. · IV EXT SET, 6.5, 1.1ML, M LUERLOCK, 1NF · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10197344098651), received between Sep 24, 2026 and Sep 25, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 31, 2025Open, Classified

    Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.

    Z-1366-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Connection Problem1
  • Detachment of Device or Device Component1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Connection Problem; Detachment of Device or Device Component

    It was reported that "The hook up is also smaller and more difficult to connect."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899080 ↗
  • MalfunctionSep 24, 2026Fluid/Blood Leak

    It was reported that Blood is suctioning back into the line and past the first port, it¿s leaking where she has marked the arrow on the photo attached.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884621 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page