MEDLINE URO4RR14CRecalls & reported problems
Medline Industries, Inc. · TRAY,ERASECAUTI,INTERMIT · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 4, 2026Open, Classified
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Z-2483-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Other Catheter care tray devices
- Centurion Medical Products Corp CVI680CMedline Industries, Inc. · Catheter care tray2 recalls0 reports
- Medline Industries, Inc. DYNJ81597Medline Industries, Inc. · Catheter care tray2 recalls0 reports
- Bardia® Foley Catheter Insertion Tray 802010C. R. Bard, Inc. · Catheter care tray54 reports
- Medline DYND10160Medline Industries, Inc. · Catheter care tray1 recall4 reports
- Medline DYNC1816Medline Industries, Inc. · Catheter care tray1 recall2 reports
- Medline DYND10407Medline Industries, Inc. · Catheter care tray1 recall1 report
- SureStep Intermittent Catheter Tray INTAAIC. R. Bard, Inc. · Catheter care tray1 recall1 report
- Medline Industries, Inc. DYNJ65199AMedline Industries, Inc. · Catheter care tray1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.