Medex MX693RRecalls & reported problems

Icu Medical, Inc. · Class 2

All models of Medex (98)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10351688505811), received between Apr 19, 2021 and Apr 19, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion2
  • Pressure Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 19, 2021Insufficient Flow or Under Infusion; Pressure Problem; Insufficient Flow or Under Infusion; Pressure Problem

    INFORMATION RECEIVED A INTERVENTIONAL IMAGING/MEDEX HIGH PRESSURE TUBING MALFUNCTIONED. THE REPORT INDICATED THAT ON TWO DIFFERENT OCCASIONS THE TUBING RUPTURED WHEN DOING PUMP PRESSURE INJECTION DURING A PROCEDURE. RATED FOR 900PSI, WERE USING 700-800 PSI. NO PATIENT INJURY DURING PROCEDURE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11685834 ↗
  • MalfunctionApr 19, 2021Insufficient Flow or Under Infusion; Pressure Problem; Insufficient Flow or Under Infusion; Pressure Problem

    INFORMATION RECEIVED A INTERVENTIONAL IMAGING/MEDEX HIGH PRESSURE TUBING MALFUNCTIONED. THE REPORT INDICATED THAT ON TWO DIFFERENT OCCASIONS THE TUBING RUPTURED WHEN DOING PUMP PRESSURE INJECTION DURING A PROCEDURE. RATED FOR 900PSI, WERE USING 700-800 PSI. NO PATIENT INJURY DURING PROCEDURE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11685527 ↗