Skip to content

Integra® Jarit® 505232Recalls & reported problems

Integra Lifesciences Production Corporation · Integra® Jarit® LARGE CERVICAL CONE FOR COHEN ACORN CANNULA · Class 2

All models of Integra Jarit (57)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780158370), received between Apr 28, 2026 and Apr 28, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Detachment of Device or Device Component

    THE FOLLOWING WAS REPORTED VIA MEDWATCH NUMBER (B)(4) THAT DURING A LAPAROSCOPIC OVARIAN CYSTECTOMY, THE COHEN ACORN CANNULA,2-1/8 CERV CONE,LG FOR 505-230 (505232) WAS "FOUND TO BE FUNCTIONAL; HOWEVER, PROPER ENGAGEMENT OF THE CONE TIP REQUIRES A 180-DEGREE TWIST FOR SECURE PLACEMENT. NO ONE NOTICED CONE MISSING FROM DEVICE UPON REMOVAL FROM USE. NO INJURY TO PATIENT, CONE PASSED SPONTANEOUSLY THROUGH VAGINA WHILE URINATING. THE CONE WAS NOT NOTICED TO BE MISSING IN THE OR. PHYSICIAN WAS MADE AWARE BY THE PATIENT AFTER DISCHARGE, WHO THEN ENSURED NO HARM OCCURRED. PATIENT RETURNED CONE TO PHYSICIAN¿S OFFICE DURING FOLLOW UP APPOINTMENT WHO INSPECTED CONE AND DETERMINED IT WAS INTACT. THE CONE WAS NOT RETURNED TO THE HOSPITAL." NO PATIENT INJURY OR DELAY REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25023781 ↗

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources and updates