Integra® Miltex® 33517Recalls & reported problems
Integra Lifesciences Production Corporation · Integra® Miltex® CryoSolutions® Cartridges 4 Pack (23.5g N2O each) · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10381780171966), received between Feb 27, 2024 and Aug 17, 2026 — classified as 5 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (12 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K091721); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Open, Classified
During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.
Z-1104-2025 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Leak/Splash12
- Insufficient Device Problem Information2
- Burst Container or Vessel1
- Explosion1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Caustic/Chemical Burns2
- Cryogenic Burn2
- Burn(s)1
- Numbness1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
OtherAug 17, 2026Nonstandard Device
Patient problems: Numbness
MAUDE report 50274694 ↗InjuryMar 16, 2026Insufficient Device Problem Information
MalfunctionSep 17, 2025Leak/Splash; Leak/Splash
MalfunctionSep 17, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT CRYOSOLUTIONS 4PK CARTRGE (33517) CONTINUES TO LEAK AFTER BEING REPAIRED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO INJURIES, DEATHS, OR SURGICAL DELAYS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23078374 ↗MalfunctionSep 16, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT CRYOSOLUTIONS 4PK CARTRGE (33517) CONTINUES TO LEAK AFTER BEING REPAIRED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO INJURIES, DEATHS, OR SURGICAL DELAYS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23063758 ↗MalfunctionJul 23, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT THE CRYOSOLUTIONS 4PK CARTRIDGE (33517) CANISTER LEAKED OR SPRAYED "CRYO" WHEN ATTACHED TO THE ADAPTER ON MORE THAN ONE OCCASION. THERE WAS NO PATIENT INVOLVEMENT; THUS, NO DEATHS, INJURIES, OR SURGICAL DELAYS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22575595 ↗MalfunctionJul 23, 2025Leak/Splash; Leak/Splash
MalfunctionJul 2, 2025Burst Container or Vessel; Burst Container or Vessel
MalfunctionMay 13, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT THEY WERE UNABLE TO SUCCESSFULLY ATTACH THE CRYOSOLUTIONS 4PK CARTRIDGE (33517) TO THE CONTROL MECHANISM WITHOUT GAS LEAKING OUT. THE O-RINGS APPEARED TO BE PROPERLY IN PLACE. THE EVENT OCCURRED BEFORE ATTEMPTING A PROCEDURE ON AN ANIMAL PATIENT. NO ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22013357 ↗MalfunctionApr 29, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT WHEN THEY "TRIED TO PUT IN THE CARTRIDGE/CRYOSOLUTIONS 4PK CARTRIDGE (33517) ON THE CANISTER, IT RELEASE A NITROUS OXIDE EVERYWHERE." IT IS UNKNOWN WHETHER THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO INJURY, DEATH, OR SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926899 ↗MalfunctionFeb 18, 2025Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT THE CARTRIDGE/CRYOSOLUTIONS 4PK CARTRIDGE (33517) SPEWED GAS WHEN TRYING TO ATTACH IT TO THE CRYOSOLUTIONS CONTROL MECHANISM BEFORE AND DURING PROCEDURES. THE DEVICE WAS NOT IN CONTACT WITH A PATIENT; THUS, NO INJURY, DEATH, OR SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21405128 ↗MalfunctionDec 29, 2024Leak/Splash; Leak/Splash
A FACILITY REPORTED THAT THERE WAS A "MASSIVE LEAK WHEN THE CARTRIDGES/CRYOSOLUTIONS CARTRIDGES (33517) WERE INSERTED IN 2 DEVICES." THERE WAS NO PATIENT INVOLVEMENT; THUS, NO INJURY OR SURGICAL DELAY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21032284 ↗InjuryAug 29, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MalfunctionJun 25, 2024Leak/Splash; Leak/Splash
InjuryMay 24, 2024Explosion; Explosion
A FACILITY REPORTED THAT WHEN THE USER PUSHED THE "TWIST" ON THE CARTRIDGE / CRYOSOLUTIONS 4PK CARTRIDGE (33517), IT EXPLODED. USER TRIED AGAIN WITH A SECOND CARTRIDGE, TWISTED ON IT AND IT EXPLODED WHEN TRYING TO REPLACE IT. IT WAS REPORTED THAT THE DOCTOR BURNED HIS FINGER FROM THE CARTRIDGE. NO USER OUTCOME, NOR MEDICAL INTERVENTION WAS NOT REPORTED.
Patient problems: Cryogenic Burn
MAUDE report 19391934 ↗InjuryFeb 27, 2024Leak/Splash; Leak/Splash
THIS REPORT IS 1 OF 2 FOR THE CARTRIDGE USED DURING THIS EVENT AND IS LINKED TO MFG NUMBER 3014334038-2024-00042. A FACILITY REPORTED THAT BEFORE A PROCEDURE, WHEN THE END USER STARTED SCREWING A CARTRIDGE OF THE CRYOSOLUTIONS 4PK CARTRIDGE (33517). GAS SPILLED OUT AND FROZE THE USER'S HANDS. THE USER EXPERIENCED REDNESS AND TENDERNESS TO HANDS. OBSERVATION WAS PERFORMED. THE CARTRIDGES WERE COLLECTED BY THE PRODUCT DISTRIBUTOR.
Patient problems: Caustic/Chemical Burns
MAUDE report 18794052 ↗InjuryFeb 27, 2024Leak/Splash; Leak/Splash
Recalls under the same FDA submission
These recalls involve devices cleared under K091721. One submission often covers many models, so check the product and lot details in each record.
- Z-1103-2025 · Dec 19, 2024 · Open, ClassifiedIntegra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;
Other UNIT, CRYOSURGICAL, ACCESSORIES devices
- IceFORCE® 2.1 CX 90° Cryoablation Needle FPRPR3602Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls4 deaths274 reports
- IceRod® 1.5 CX 90° Cryoablation Needle FPRPR3533Boston Scientific Corporation · Unit, Cryosurgical, Accessories1 recall168 reports
- IcePearl® 2.1 CX 90° Cryoablation Needle FPRPR3601Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls78 reports
- IceFORCE™ 2.1 CX L 90° Cryoablation Needle FPRPR3618Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls6 reports
- IceFORCE® 2.1 CX Cryoablation Needle FPRPR3604Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls3 reports
- IcePearl™ 2.1 CX L 90° Cryoablation Needle FPRPR3617Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls1 report
- IcePearl® 2.1 CX Prostate Cryoablation Kit / Visual-ICE® System FPRPR3606Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
- IceFORCE® 2.1 CX Prostate Cryoablation Kit / Visual-ICE® System FPRPR3605Boston Scientific Corporation · Unit, Cryosurgical, Accessories2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.