Skip to content

Integra® Miltex® 33518Recalls & reported problems

Integra Lifesciences Production Corporation · Integra® Miltex® CryoSolutions® Cartridges 1 Pack (23.5g N2O each) · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780171973), received between Sep 24, 2024 and Aug 11, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K091721); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2
  • Leak/Splash2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2025Leak/Splash; Leak/Splash

    A FACILITY REPORTED THAT WHEN THEY ATTACHED THE CRYOSOLUTIONS 1PK CARTRGE (33518) DURING A PROCEDURE, ALL OF THE AIR CAME OUT AT ONCE. NO INJURIES, DEATH, OR SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22770673 ↗
  • MalfunctionJul 2, 2025Leak/Splash; Leak/Splash

    A FACILITY REPORTED THAT THREE OF THE CRYOSOLUTIONS 1PK CARTRIDGE (33518) WERE LEAKING. IT IS UNKNOWN UNDER WHAT CIRCUMSTANCE THIS EVENT OCCURRED; HOWEVER, NO INJURY, DEATH OR SURGICAL DELAY WAS REPORTED. IT IS NOTED THAT THE REPORTING FACILITY IS A VETERINARY CENTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22391768 ↗
  • MalfunctionFeb 3, 2025Gas/Air Leak; Gas/Air Leak

    A FACILITY REPORTED THAT WHEN THEY WENT TO USE THE CRYOSOLUTIONS 1PK CARTRIDGE (33518) ALL THE GAS CAME OUT. THERE HAVE BEEN NO REPORTS OF PATIENT INVOLVEMENT, INJURY, OR SURGICAL DELAY ASSOCIATED WITH THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21297604 ↗
  • MalfunctionSep 24, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT DURING REPLACEMENT OF THE NEW CARTRIDGE/CRYOSOLUTIONS 1PK CARTRIDGE (33518), GAS STARTED SHOOTING OUT OF THE HOLES AT THE BASE OF THE HANDPIECE, COMPLETELY DRAINING THE NEW CARTRIDGE WITHIN A FEW SECONDS. IT WAS CONFIRMED THAT THE O-RING WAS STILL IN PLACE. IT WAS REPORTED THAT THE DEVICE WAS GOING TO BE USED ON AN ANIMAL PATIENT. NO INJURY, DEATH, OR SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297965 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091721. One submission often covers many models, so check the product and lot details in each record.

  • Z-1104-2025 · Dec 19, 2024 · Open, Classified
    Integra Miltex Cryosolutions Cartridges 4 Pack (23.5g N2O each), Model: C-CA-23, Item: 33517;
  • Z-1103-2025 · Dec 19, 2024 · Open, Classified
    Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

Other UNIT, CRYOSURGICAL, ACCESSORIES devices

All UNIT, CRYOSURGICAL, ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page