MAYFIELD® A2600MRecalls & reported problems
Integra Lifesciences Corporation · MAYFIELD® SPINE TABLE ADAPTOR · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780268062), received between Jun 16, 2021 and Mar 13, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem2
- Device Dislodged or Dislocated1
- Device Slipped1
- Mechanics Altered1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 13, 2024Device Slipped; Device Slipped
THIS IS 1 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2024-00033: A FACILITY REPORTED THAT DURING AN UNSPECIFIED CASE, THE MAYFIELD SPINE TABLE ADAPTOR METAL (A2600M) SLIPPED, RESULTING IN INJURY TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18894988 ↗MalfunctionDec 22, 2021Mechanics Altered; Mechanics Altered
THIS IS 2 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3004608878-2021-00706: A FACILITY REPORTED THAT WHEN PATIENT WAS PUT IN PINS WITH MAYFIELD SKULL CLAMP AND THE MAYFIELD SPINE TABLE ADAPTOR BASE UNIT (A2600M), THE PATIENT MOVED DURING THE PROCEDURE AS THE PATIENT¿S POSITIONING CHANGED. THE HANDLE¿S RELEASING LEVER IS VERY LOOSE, AND THOUGHT TO BE THE ISSUE. THE CUSTOMER SENT IN BOTH UNITS FOR FURTHER INSPECTION DUE TO THE INCIDENT. NO PATIENT INJURY OR SURGICAL DELAY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13061992 ↗MalfunctionJun 17, 2021Mechanical Problem; Device Dislodged or Dislocated; Mechanical Problem; Device Dislodged or Dislocated
A FACILITY REPORTED THAT THE SHOCK CUSHION ON THE MAYFIELD SPINE TABLE ADAPTOR (A2600M) WAS DISLODGED AND IT WOULD NOT LOCK SAFELY. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO PATIENT INJURY WAS REPORTED OR SURGICAL DELAY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12022003 ↗MalfunctionJun 16, 2021Mechanical Problem; Mechanical Problem
A FACILITY REPORTED THAT THE GOLD HANDLE ON THE MAYFIELD SPINE TABLE ADAPTOR METAL (A2600M) WOULD NOT LOCK SECURELY. IT IS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO PATIENT INJURY OR SURGICAL DELAY HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12009243 ↗
Other Operating Room Accessories Table Tray devices
- MAYFIELD® A2101Integra Lifesciences Corporation · Operating Room Accessories Table Tray101 reports
- NA 210080Mako Surgical Corp. · Operating Room Accessories Table Tray74 reports
- ALLEN 104002Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall8 reports
- N/A 7210569Smith & Nephew, Inc. · Accessories, operating-room, table (kit)57 reports
- ALLEN 107004Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall5 reports
- ALLEN 108001Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall3 reports
- NA 7210551Smith & Nephew, Inc. · Accessories, operating-room, table (kit)41 reports
- MAYFIELD® A1018Integra Lifesciences Corporation · Operating Room Accessories Table Tray36 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.