Skip to content

RELI® SBCN-R22G101Recalls & reported problems

Myco Medical Supplies Inc. · RELI® Safety Blood Collection Needle 22G x 1”, Sterile · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10386120009350), received between Jul 10, 2024 and Jul 10, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fitting Problem1
  • Improper or Incorrect Procedure or Method1
  • Lack of Effect1
  • Leak/Splash1

Patient problems

  • Needle Stick/Puncture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2024Break; Leak/Splash; Improper or Incorrect Procedure or Method; Fitting Problem; Lack of Effect; Break; Leak/Splash; Improper or Incorrect Procedure or Method; Fitting Problem; Lack of Effect

    8 TEAMMATES COMMENTS THAT THE "TUBES POP" OFF AND WHEN REAPPLYING TUBES ARE NOT FILLING PROPERLY. 4 TEAMMATES COMMENTS THAT THE DEVICE IS DIFFICULT TO SCREW ONTO THE ADAPTERS THAT IT CROSS THREADS. 3 TEAMMATES THAT THEY HAVE HAD THE SAFETY FEATURECOMPLETELY BREAK OFF WHEN APPLYING ONTO THE HUB·MULTIPLE TEAM HAVE SAID THE NEEDLES ONCE SET UP THEY HAVE HAD DIFFICULTYREMOVING THE CAP TO EXPOSE THE COLLECTION NEEDLE ·2 HAVE HAD THE RUBBER THAT CLOSES OFF THE BLOOD FROM ENTERING THE TUBECOME OFF OR PARTIALLY COME OFF DURING COLLECTION RESULTING IN BLOOD SPILLAGE. " DEVICE SEEMS MORE BULKIER, AND CUMBERSOMEBECAUSE THE SAFETY SHIELD GETS IN THE WAY AND WHEN REPOSITIONED THE SAFETY FEATURE BECOMES WEAKENED AND HAS BROKEN " "HAVE HAD SOME THAT THE SAFETY GUARD WERE ALREADY DISLODGED PRIOR TO SET UP" "SAFETY DEVICE HAD NOT CLOSED PROPERLY SEVERALTIMES, CONCERNED ABOUT GETTING ACCIDENTLY STUCK"·"SAFETY SHIED DOES NOT OPEN FULLY TO VIEW COLLECTION SITE PRECISELY"·WE RECENTLY HAD A LONG-TIME EMPLOYEE HAVE AN ACCIDENTAL DIRTY NEEDLE STICK DUE TO SAFETY SHIELD WOULD NOT OPEN FULLY FOR VIEW.

    Patient problems: Needle Stick/Puncture

    MAUDE report 19709360 ↗

Other Needle, Hypodermic, Single Lumen devices

All Needle, Hypodermic, Single Lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page