SELF CENTERING 1035-56-000Recalls & reported problems

Depuy Orthopaedics, Inc. · SELF CENTERING BI-POLAR HEAD 28mm ID 56mm OD · Class 2 · Implant

All models of SELF CENTERING (29)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295003465), received between Apr 6, 2018 and Feb 27, 2026 — classified as 14 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Device Dislodged or Dislocated6
  • Appropriate Device Problem Term/Code Not Available1
  • Device Expiration Issue1
  • Naturally Worn1

Patient problems

  • Joint Dislocation5
  • Joint Laxity4
  • Unspecified Infection3
  • No Code Available2
  • Pain2
  • Bone Fracture(s)1
  • Fall1
  • Insufficient Information1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD AN INFECTED RIGHT HIP. BIPOLAR HEAD AND 28 HEAD EXCHANGE. DOI: (B)(6) 2026. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 24472991 ↗
  • InjuryDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS INFECTED AND DOCTOR PERFORMED I&D AND CHANGED BIPOLAR HEAD. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 23864855 ↗
  • InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REVISED DUE TO PAIN. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Pain

    MAUDE report 22793289 ↗
  • InjuryJun 27, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD A PROSTALAC SPACER PLACED BUT DISLOCATED POSTOPERATIVELY. THE DOCTOR BROUGHT THE PATIENT BACK TO REVISE. HE BELIEVED THAT THE VERSION AND OFFSET NEEDED TO BE CHANGED. THE PROSTALAC STEM WAS REMOVED. A NEW HIGH OFFSET SIZE 1 PROSTALAC WAS IMPLANTED WITH A 56MM BIPOLAR HEAD AND 28 +8.5 FEMORAL HEAD. AFFECTED SIDE : LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 22344760 ↗
  • InjuryMar 14, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO THE PATIENT HAD FELL AND DISLOCATED THE HIP. THE SURGEON DECIDED TO REMOVE THE BIPOLAR SHELL AND REPLACE WITH A CUP AND CONSTRAINED LINER. THERE WAS NO PROBLEM WITH OUR IMPLANTS AND THERE WAS NO EXTENDED TIME AS WELL. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 21608261 ↗
  • InjuryOct 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS UNSTABLE. THE STEM VERSION AND OFFSET WERE CHANGED. NO DELAYS. NO HARM TO PATIENT. NO FURTHER INFORMATION IS AVAILABLE. DOI- (B)(6) 2024. DOR- (B)(6) 2024. AFFECTED SIDE- LEFT HIP.

    Patient problems: Joint Laxity

    MAUDE report 20344157 ↗
  • InjurySep 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO UNSTABLE HIP. NO POLY WEAR. DO NOT KNOW CAUSE OF INSTABILITY. NO DELAY AND NO FURTHER INFORMATION IS AVAILABLE. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Laxity

    MAUDE report 20281922 ↗
  • InjurySep 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED: PRIMARY DATE UNKNOWN. NO DELAY, INSTABILITY, NO HARM.

    Patient problems: Joint Laxity

    MAUDE report 20156943 ↗
  • InjuryJan 23, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S HIP WAS REVISED DUE TO DISLOCATION. THE HIP COULD NOT BE REDUCED UNDER SEDATION, SO THE SURGEON PERFORMED AN OPEN REDUCTION PROCEDURE. DURING THE PROCEDURE, THE SURGEON REMOVED AND REPLACED THE PATIENT'S ORIGINAL BI-POLAR HEAD CONSTRUCT WITH A LONGER UNI POLAR HEAD CONSTRUCT, TO INCREASE THE PATIENT'S LEG LENGTH AND FEMORAL OFFSET FOR INCREASED STABILITY. UNKNOWN SURGICAL DELAY. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation; Joint Laxity; Physical Asymmetry

    MAUDE report 18565040 ↗
  • InjuryAug 8, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO BIPOLAR LINER WAS DISASSOCIATING FROM METAL SHELL DURING FINAL RANGE OF MOTION TEST. DOI: UNKNOWN; DOR: (B)(6) 2023; AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 17481808 ↗
  • InjuryJan 6, 2022Naturally Worn; Naturally Worn

    MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS PATIENT WAS REVISED WAS DUE TO PAINFUL HEMI ARTHROPLASTY WITH ARTHRITIS OF LEFT HIP AND OBVIOUS WEAR OF HIS BIPOLAR IMPLANT. DOI: (B)(6) 2017, DOR: (B)(6) 2019, LEFT HIP 2ND REVISION. PLEASE SEE (B)(4) LEFT HIP 1ST REVISION.

    Patient problems: Pain

    MAUDE report 13175264 ↗
  • InjuryOct 18, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS REVISED TO ADDRESS CHRONIC DISLOCATION, POSTERIOR ACETABULAR WALL COLLAPSE . GIRDLESTONE PROCEDURE. DOI: (B)(6) 2019; DOR: (B)(6) 2019; RIGHT HIP.

    Patient problems: Bone Fracture(s); Joint Dislocation; No Code Available

    MAUDE report 9209379 ↗
  • InjuryFeb 27, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES INFECTION AND BONE FRACTURE AFTER FIRST REVISION. DOI: (B)(6) 2014; DOR: (B)(6) 2014; (LEFT HIP).

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 8374029 ↗
  • InjuryNov 8, 2018Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available; Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available

    REVISED DUE TO DISASSOCIATION.

    Patient problems: Joint Dislocation

    MAUDE report 8051595 ↗
  • MalfunctionApr 6, 2018Device Expiration Issue; Device Expiration Issue

    DER STATES A BI-POLAR HEAD WAS IMPLANTED IN THIS PATIENT THAT HAD AN EXPIRATION DATE OF (B)(6) 2018. SALES REP MISSED THE DATE, AS DID THE CIRCULATOR AND THE SCRUB NURSE. I WAS JUST NOTIFIED ABOUT THE MISTAKE. THE FACILITY IS REQUESTING A RESPONSE LETTER, STATING THE IMPLANT AND PATIENT ARE SAFE. DOE: (B)(6) 2018 UNKNOWN AFFECTED SIDE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7402790 ↗

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →