S-ROM 1365-17-500Recalls & reported problems
Depuy (Ireland) · S-ROM M METAL ON METAL FEMORAL HEAD Diameter 28mm +3 11/13 TAPER ULTAMET · Class 2 · Implant
All models of S-ROM (185)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295033158), received between Feb 1, 2018 and Oct 13, 2020 — classified as 6 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Device Dislodged or Dislocated2
Patient problems
- No Code Available3
- Joint Dislocation2
- Not Applicable1
- Osteolysis1
- Pain1
- Tissue Damage1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO POLYETHYLENE WEAR. LEG LENGTH WAS SHORT. BOTH THE HEAD AND LINER WERE EXCHANGED. HEAD WAS INCREASED FROM A 28 +3 TO A 36+9. LINER CHANGED FROM 56 X 28 WITH +4 10 DEGREE TO 56 X 36 +4 10 DEGREE. DOI: UNKNOWN, DOR: (B)(6) 2020, AFFECTED SIDE: LEFT HIP.
Patient problems: No Code Available
MAUDE report 10671564 ↗InjuryMay 21, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED DUE TO DISLOCATING HIS HIP ON MULTIPLE OCCASIONS. THE SURGEON REVISED THE ACETABULAR COMPONENTS TO IMPROVE STABILITY. THE PINNACLE CUP WAS EXPLANTED AND A ZIMMER BIOMET CUP WAS IMPLANTED. THE S ROM STEM WAS LEFT IMPLANTED BUT THE HEAD WAS REVISED TO A 28MM +9 METAL S ROM COMPONENT. DOI: UNKNOWN. DOR: (B)(6) 2020. LEFT HIP.
Patient problems: Joint Dislocation; No Code Available
MAUDE report 10080730 ↗InjuryJan 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PINNACLE CLAIM FORM AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF MEDICAL RECORDS, IT WAS STATED THAT THE PATIENT WAS REVISED TO ADDRESS PAIN AND INFECTION. ANTIBIOTIC SPACERS WERE THEN IMPLANTED. RE-REVISION TO REMOVED THE SPACERS AND RE-IMPLANTATION OF DEFINITE IMPLANTS OCCURRED ON (B)(6) 2011. DOI: (B)(6) 2008 (CUP, SCREW, STEM, SLEEVE); DOI: (B)(6) 2010 (HEAD, LINER); DOR: (B)(6) 2011; RIGHT HIP.
Patient problems: No Code Available; Pain; Unspecified Infection
MAUDE report 9643399 ↗InjuryOct 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PPF AND STICKER SHEETS RECEIVED. PPF ALLEGES ABDUCTOR MUSCLE REPAIR. DOI: (B)(6) 2005; DOR: NONE REPORTED; LEFT HIP.
Patient problems: Tissue Damage
MAUDE report 9202167 ↗InjurySep 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A GREATER TROCHANTER FRACTURE CAUSED BY OSTEOLYSIS. COBALT LEVELS WERE 1.9. ALL IMPLANTS WERE STILL WELL FIXED. THE DURALOC CUP WAS REMOVED AND REPLACED WITH A MUCH LARGER COMPETITOR CUP WITH A DUAL MOBILITY CONSTRUCT. THE STEM WAS LEFT AND A NEW HEAD WAS IMPLANTED. ORIGINAL IMPLANT DATE UNKNOWN. DOI: UNKNOWN - DOR: (B)(6) 2018; LEFT HIP.
Patient problems: Not Applicable; Osteolysis
MAUDE report 7864179 ↗InjuryFeb 1, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED TO ADDRESS DISLOCATION. SURGEON SAID THAT PATIENT HAS BEEN A CHRONIC DISLOCATOR SINCE 2012 AND PATIENT HAD A PRIOR REVISION 2 YEARS AGO. DOI: (B)(6) 2016; DOR: (B)(6) 2018; RIGHT HIP.
Patient problems: Joint Dislocation
MAUDE report 7233590 ↗
Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page