Skip to content

SIGMA 2011-03-003Recalls & reported problems

Depuy (Ireland) · SIGMA HP REVISION INSTRUMENTS 4 IN 1 CUTTING BLOCK SIZE 4 · Class 1

All models of SIGMA (380)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295081241), received between Aug 8, 2017 and Jun 8, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Connection Problem1
  • Difficult to Remove1
  • Disassembly1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Information1
  • Not Applicable1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE TRYING TO REMOVE THE PIN OUT OF THE SIZE 4 TC3 FEMORAL CUTTING BLOCK, THE PIN GOT STUCK AND CAUSED DAMAGE TO THE INSIDE HOLE OF THE CUTTING BLOCK. CASE WAS FINISHED OKAY AND WITHOUT ANY PROBLEMS OR DELAYS. JUST NEED BLOCK REPLACED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11958828 ↗
  • MalfunctionJul 22, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREW WILL NOT PASS THROUGH SCREW HOLE IN SIZE 4, 4-IN-1 CUTTING BLOCK AND NOTCH CUTTING GUIDE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10309034 ↗
  • MalfunctionMar 12, 2019Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE SURGEON WAS DRIVING THE PIN INTO THE TC3 SIZE 4 CUTTING BLOCK AND THROUGH THE AUGMENT WHEN THE PIN BENT AND GOT BOUND UP BREAKING THE PIN HEAD OFF INTO THE PIN DRIVER AND BENDING THE PIN AT THE SAME TIME.

    Patient problems: Not Applicable

    MAUDE report 8412684 ↗
  • MalfunctionAug 8, 2017Disassembly; Difficult to Remove; Connection Problem; Disassembly; Difficult to Remove; Connection Problem

    IT WAS REPORTED THAT A HEADED PIN GOT STUCK INSIDE A SIZE 4 HP REVISION 4 IN 1 CUTTING BLOCK WITH 8 MM AUGMENT ON IT. PULLED PIN FROM BONE WITH PLYERS.

    Patient problems: No Information

    MAUDE report 6773200 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page