Skip to content

GLOBAL AP 2130-20-000Recalls & reported problems

Depuy (Ireland) · GLOBAL AP OSTEOTOME GUIDE PIN 3.2mm LONG · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295090809), received between May 10, 2018 and Jan 23, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation4
  • No Apparent Adverse Event2
  • Component Missing1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Material Deformation

    IT WAS REPORTED THAT THE 3.2MM OSTEOTOMY GUIDE PIN LONG BENT IN USE DURING SURGERY. NO DELAY TO SURGERY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24172890 ↗
  • MalfunctionApr 29, 2025Material Deformation; Material Deformation

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE EVENT OCCURRED DURING SURGERY. NO PATIENT HARM REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21926382 ↗
  • MalfunctionApr 29, 2025Material Deformation; Material Deformation

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE EVENT OCCURRED DURING SURGERY. NO PATIENT HARM REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21926363 ↗
  • MalfunctionApr 16, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THIS INSTRUMENT IS COMPOSED OF 1 PART EACH. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER. ADDITIONAL EVENT INFORMATION: THE PHOTO INVESTIGATION REVEALED THAT ' 213020000, GL AP 3.2MM OSTEO GD PIN LONG¿ HAD SCRATCHES ON THE SURFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19121851 ↗
  • MalfunctionApr 16, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THIS INSTRUMENT IS COMPOSED OF 1 PART EACH. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER. ADDITIONAL EVENT INFORMATION THE PHOTO INVESTIGATION REVEALED THAT 213020000, GL AP 3.2MM OSTEO GD PIN LONG¿ HAD SCRATCHES ON THE SURFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19121840 ↗
  • MalfunctionNov 30, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE GUIDE PIN WAS FOUND BENT IN SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12895772 ↗
  • MalfunctionMay 10, 2018Component Missing; Manufacturing, Packaging or Shipping Problem; Component Missing; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT WHEN THE ITEM ARRIVED AT THE OFFICE, THE PACKAGE WHICH SHOULD HAVE HELD A GLOBAL OSTEOTOME GUIDE PIN, WAS EMPTY. A HOSPITAL AND SURGERY NOT INVOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7503361 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page