Skip to content

NA 2178-65-140Recalls & reported problems

Depuy (Ireland) · M.B.T. REVISION TIBIAL VIEW PLATE SIZE 4 · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy M.B.T. Revision Tibial View Plate is a trial plate used to check fit during knee replacement surgery. The 64 reports mostly describe the plate snapping in half, cracking, or wearing out — many noticed when technicians inspected loaner sets rather than during surgery — with no deaths and no injuries. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 10603295096689), received between Jun 29, 2017 and Sep 4, 2025 — classified as 64 malfunctions.

The most frequently reported specific device problem is “Break” (55 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break55
  • Crack8
  • Naturally Worn2
  • Dull, Blunt1
  • Material Deformation1
  • Scratched Material1

Patient problems

  • No Consequences Or Impact To Patient40
  • No Information15
  • No Clinical Signs, Symptoms or Conditions8
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION BY THE LOAN KIT TECHNICIAN IT WAS IDENTIFIED THAT THE MBT REV TIBIAL VIEW PLT SZ4 HAS BROKEN IN HALF. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22965909 ↗
  • MalfunctionApr 25, 2023Break; Break

    IT WAS REPORTED THAT TRIAL IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16806705 ↗
  • MalfunctionAug 1, 2022Scratched Material; Scratched Material

    IT WAS REPORTED THAT PLATE WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15141943 ↗
  • MalfunctionJun 17, 2022Break; Break

    IT WAS REPORTED THAT THE TIBIAL SIZER WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14734643 ↗
  • MalfunctionMay 12, 2022Break; Break

    TIBIAL VIEW PLATE HAS SNAPPED IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14368197 ↗
  • MalfunctionApr 29, 2022Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE TABS ON 14X75 PILOT SHAFT WAS BROKE. ALL PIECES WERE RETRIEVED. MBT SIZE 4 CLEAR TRIAL BROKE W/WEAR. NO PIECES. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14251770 ↗
  • MalfunctionFeb 11, 2021Break; Break

    BOTH TIBIAL VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11313311 ↗
  • MalfunctionDec 15, 2020Break; Break

    BOTH TIBIAL VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11007863 ↗
  • MalfunctionAug 6, 2020Break; Break

    IT WAS REPORTED THAT THE WRENCH IS ROUNDED OFF AND WAS SLIPPING WHILE TIGHTENING THE STEMMED IMPLANT. THE SZ 4 VIEW PLATE BROKE DURING THE SURGERY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10374330 ↗
  • MalfunctionJan 16, 2020Break; Break

    ALL FOUR MBT TIBIAL VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9596032 ↗
  • MalfunctionOct 18, 2019Break; Break

    IT WAS REPORTED THAT THE PLATE WAS BROKEN DURING CASE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9209060 ↗
  • MalfunctionAug 12, 2019Break; Break

    IT WAS REPORTED THAT THE PIECE WAS BROKEN IN TRAY BEFORE CASE SIZE4 CLEAR PIECE TIBIA TRIAL. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8885549 ↗
  • MalfunctionAug 6, 2019Break; Break

    BOTH TIBIAL VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8864677 ↗
  • MalfunctionJul 9, 2019Break; Break

    SIZE 4 MBT REVISION SIZING GUIDE BROKE INTRAOPERATIVELY. PATIENT STATUS/ OUTCOME / CONSEQUENCES? NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8770411 ↗
  • MalfunctionJul 9, 2019Break; Naturally Worn; Break; Naturally Worn

    ORTHO TECH REPORTED THE INSTRUMENT AS BROKEN FOLLOWING DECONTAMINATION FOLLOWING A CASE. REPLACEMENT INSTRUMENT REQUIRED FOR LTC KIT. IF OTHER, DESCRIBE, UNSURE IF BREAKAGE OCCURRED INTRA-OP OR POST-OP DUE TO GENERAL WEAR AND TEAR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8770405 ↗
  • MalfunctionJun 5, 2019Break; Break

    DURING INSPECTION, IT WAS NOTICED THAT PRODUCT WAS DAMAGED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8670428 ↗
  • MalfunctionJun 5, 2019Break; Break

    DURING INSPECTION, IT WAS NOTICED THAT PRODUCT WAS DAMAGED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8670415 ↗
  • MalfunctionMay 31, 2019Break; Break

    IT WAS REPORTED THAT THE MBT REV TIBIAL VIEW PLT SZ4 CLEAR PLASTIC SIZER WAS SPLIT IN TWO. EVERYTHING WAS RETRIEVED FROM THE PATIENT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8659602 ↗
  • MalfunctionApr 26, 2019Break; Break

    SZ4 TIBIAL VIEW PLATE HAS SNAPPED IN HALF. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8556108 ↗
  • MalfunctionMar 22, 2019Crack; Crack

    SURGEON WAS USING THE VIEW PLATE TO REAM THROUGH TO MAKE SURE HE WAS CENTERED. REAMED UP AND THEN DIDN'T REMOVE IT BEFORE THE PROXIMAL REAMER AND IT CRACKED AND BROKE THE VIEW PLATE. ALSO, SPD NOTICED WHEN THEY WERE PUTTING THE SET BACK TOGETHER THAT THE OTHER PLATE WAS CRACKED. UNSURE OF WHEN THAT HAPPENED WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? --> NO, ACTION TAKEN WHEN EVENT OCCURRED? --> REMOVED BROKEN 2 PIECES AND MOVED ON, WAS PROCEDURE SUCCESSFULLY COMPLETED? --> YES, WERE FRAGMENTS GENERATED? --> NO, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? --> YES, IF NO, EXPLAIN --> PERFECT BREAK IN TWO, PATIENT STATUS/ OUTCOME / CONSEQUENCES --> NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: --> UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY --> NO, ACTIVITY LEVEL --> UNKNOWN, (B)(4) DEVICE PROPERTY OF -->HOSPITAL, DEVICE IN POSSESSION OF -->(B)(6), (B)(4). DEVICE PROPERTY OF -->HOSPITAL, DEVICE IN POSSESSION OF -->(B)(4) BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8444833 ↗
  • MalfunctionFeb 22, 2019Crack; Crack

    IT WAS REPORTED THAT VIEW PLATES WERE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8361294 ↗
  • MalfunctionFeb 22, 2019Crack; Crack

    TRIALS CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8361280 ↗
  • MalfunctionFeb 20, 2019Break

    THE INSTRUMENT WAS REPORTED FOR UNKNOWN REASON. OFFICE SETS, DID NOT EFFECT SURGERY NOR PATIENT. (B)(6) 2019 EXAMINATION OF THE RETURNED DEVICE REVEALED DEVICE BREAKAGE. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8353445 ↗
  • MalfunctionDec 20, 2018Crack; Crack

    IT WAS REPORTED THAT THE TRIALS ARE CRACKED. HUDSON WILL NOT LOCK ONTO REAMERS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8184795 ↗
  • MalfunctionNov 29, 2018Break; Break

    BOTH VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8118433 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →