NA 2178-65-150Recalls & reported problems

Depuy (Ireland) · M.B.T. REVISION TIBIAL VIEW PLATE SIZE 5 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy M.B.T. revision tibial view plate is an orthopedic manual surgical instrument. The 41 reports mostly describe the plate or related instruments breaking or cracking, sometimes with broken pieces recovered. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 10603295096696), received between Jun 29, 2017 and Feb 11, 2026 — classified as 41 malfunctions.

The most frequently reported specific device problem is “Break” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break32
  • Crack10
  • Material Deformation1
  • No Apparent Adverse Event1
  • Scratched Material1

Patient problems

  • No Consequences Or Impact To Patient28
  • No Information8
  • No Clinical Signs, Symptoms or Conditions5
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE INSTRUMENTS WERE WORN AND NEED TO BE REPLACED. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316843 ↗
  • MalfunctionMay 20, 2024Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE SHIM TEMPLATE CRACKED. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19355769 ↗
  • MalfunctionJan 4, 2024Crack; Crack

    THIS INSTRUMENT IS COMPOSED OF 1 PART. THERE WAS A CRACK ON THE MBT REV TIBIAL TRIAL. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446275 ↗
  • MalfunctionFeb 16, 2022Crack; Crack

    IT WAS REPORTED THAT RECEIVED DAMAGED INSTRUMENT BY SPD TECH ON (B)(6) 2022. THE TIBIAL SIZER WAS CRACKED AND LEFT OUT OF THE TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13540978 ↗
  • MalfunctionDec 10, 2020Break; Break

    IT WAS REPORTED THAT THE PLASTIC TIP OF THE TORQUE WRENCH BROKE OFF PREVENTING THEM FROM BEING ABLE TO TIGHTEN THE ADAPTOR BOLT. ALL PIECES WERE RETRIEVED FROM THE PATIENT, HOWEVER, THE PLASTIC PIECE WAS THROWN AWAY AND WILL NOT BE BEING RETURNED. THE WRENCH PORTION WILL BE BEING RETURNED. IT WAS ALSO REPORTED THAT WHILE SIZING THE TIBIA, THE VIEW PLATE BROKE IN HALF. BOTH PIECES WERE RECOVERED FROM PT AND ARE BEING RETURNED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10984740 ↗
  • MalfunctionJul 22, 2020Scratched Material; Scratched Material

    SCUFFED INSTRUMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10308482 ↗
  • MalfunctionApr 2, 2020Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENTS WERE FOUND BROKEN IN SPD AND WORN OUT/STRIPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9917084 ↗
  • MalfunctionJan 16, 2020Break; Break

    ALL FOUR MBT TIBIAL VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9596050 ↗
  • MalfunctionJan 2, 2020Break; Crack; Break; Crack

    IT WAS REPORTED THAT THE SZ 5, 3 AND 2.5 VIEW PLATES WERE CRACKED, CHIPPED. DAMAGED FOUND WHEN THEY RETURNED TO THE OFFICE FOR INSPECTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9540620 ↗
  • MalfunctionAug 9, 2019Break; Break

    PIECE OF VIEW PLATE BROKE OFF DURING SURGERY, RESULTED IN NO DELAY. ITEM DID BREAK INTO 2 PIECES, BOTH RETRIEVED. PART THAT BROKE WAS AROUND CENTRAL HOLE. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? --> NO, WAS PROCEDURE SUCCESSFULLY COMPLETED? --> YES, WERE FRAGMENTS GENERATED? --> YES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8880177 ↗
  • MalfunctionJul 30, 2019Break; Break

    IT WAS REPORTED THAT WE LOST BUSINESS A LOT OF BUSINESS HERE IN TOWN. ALL INSTRUMENTS HAVE BEEN PULLED TO RE-ALLOCATE WITHIN OUR TERRITORY. BROKEN INSTRUMENTS WERE IN DESIGNATED BOX AT FACILITY LABELED AS "BROKEN" OR WERE FOUND IN STERILE WRAPPED SETS THAT WERE RETURNED TO THE OFFICE. NO PATIENT WAS AFFECTED. NO SURGERY WAS AFFECTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8841170 ↗
  • MalfunctionFeb 22, 2019Crack; Crack

    TRIALS CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8361279 ↗
  • MalfunctionFeb 18, 2019Crack; Crack

    IT WAS REPORTED THAT WHILE TRIALING, THE KNEE WAS PUT THROUGH RANGE OF MOTION AND THE TRIAL WAS CRACKED IN HALF. THE TRIAL WAS THEN SIZED UP TO A 6 VIEW PLATE WHICH WAS ALSO CRACKED DURING TRIALING. NO TIME WAS ADDED TO THE CASE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8347314 ↗
  • MalfunctionDec 20, 2018Crack; Crack

    IT WAS REPORTED THAT THE TRIALS ARE CRACKED. HUDSON WILL NOT LOCK ONTO REAMERS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8184768 ↗
  • MalfunctionDec 7, 2018Break; Break

    INSTRUMENT BROKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8143275 ↗
  • MalfunctionNov 29, 2018Break; Break

    BOTH VIEW PLATES HAVE SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8118431 ↗
  • MalfunctionOct 23, 2018Crack; Crack

    IT WAS REPORTED THAT THE SIZING PLATE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7996451 ↗
  • MalfunctionAug 13, 2018Break; Break

    VIEW PLATE HAS SNAPPED IN HALF. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7776203 ↗
  • MalfunctionJun 11, 2018Break; Break

    VIEW PLATE HAS SNAPPED IN HALF. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7588969 ↗
  • MalfunctionJun 8, 2018Break; Break

    IT WAS REPORTED THAT UPON PLACING THE TIBIAL SIZING VIEW PLATE ON THE TIBIA AND WHEN IT WAS REMOVED, IT WAS IN TWO PIECES. WHEN THE SCRUB TECH WAS GOING TO LIFT THE UPPER TRAY OUT OF THE PAN TO GET TO THE BOTTOM, IT BROKE IN HALF. ALL PIECES ARE BEING RETURNED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7582478 ↗
  • MalfunctionJun 8, 2018Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS FOUND BROKEN IN SPD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7582039 ↗
  • MalfunctionJun 7, 2018Break; Break

    VIEW PLATES HAVE ALL SNAPPED IN HALF. THIS CASE HAS BEEN REPORTED BY THE LOAN KIT TECHNICIAN DURING LOAN KIT INSPECTION. IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7577433 ↗
  • MalfunctionMay 25, 2018Break; Break

    VIEW PLATE HAS SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7544693 ↗
  • MalfunctionMay 24, 2018Break; Break

    IT WAS REPORTED THAT TIBIAL SIZING CLEAR DISC SNAPPED IN HALF. SURGERY DELAY OF 5 MINS.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7541376 ↗
  • MalfunctionMay 15, 2018Break; Break

    VIEW PLATE HAS SNAPPED IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7514363 ↗