Skip to content

PINNACLE 2217-00-020Recalls & reported problems

Depuy (Ireland) · PINNACLE ALTERNATE BEARING SUCTION CUP INSERTER · Class 1

All models of PINNACLE (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle alternate bearing suction cup inserter is a surgical instrument used to place implant components during hip replacement. The 43 reports mostly describe the plastic suction tip cracking, splitting, or coming apart, sometimes before or during surgery, with wear after sterilization and a few cases of pieces missing. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 10603295097235), received between Aug 28, 2018 and May 21, 2026 — classified as 43 malfunctions.

The most frequently reported specific device problem is “Break” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break26
  • Component Missing6
  • Naturally Worn5
  • Device-Device Incompatibility4
  • Material Deformation3
  • Material Twisted/Bent2
  • Unintended Movement2
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions30
  • No Consequences Or Impact To Patient13
  • No Code Available1

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 21, 2026Break

    IT WAS REPORTED THAT THE PINNACE SUCTION CUP INSERTER WAS BROKEN AND WORN. THERE WAS NO PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY IN SURGERY. THE LOT NUMBER WAS NOT AVAILABLE AND THE DEVICE WAS THROWN OUT AND THEREFORE, NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25246733 ↗
  • MalfunctionDec 23, 2025Device-Device Incompatibility; Device-Device Incompatibility

    THE SUCTION CUP ON THE INSERT HOLDER IS DEFECTIVE; IT NO LONGER STAYS ON THE HANDLE AND CAN NO LONGER PROPERLY HOLD (SUCTION) AND INSERT AFTER THE EQUIPMENT HAS BEEN STERILIZED. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? NO. IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? NO. PATIENT STATUS/ OUTCOME / CONSEQUENCES NO. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: NO. IS THE PATIENT PART OF A CLINICAL STUDY: NO. IP, (B)(6). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST: TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877199 ↗
  • MalfunctionJul 31, 2025Break; Break

    IT WAS REPORTED THAT THE PLASTIC PIECE AT THE TIP OF THE INSTRUMENT IS RIPPED. AN ALTERNATE BEARING SUCTION CUP INSERTER WAS USED. THE EVENT DID NOT HAPPEN DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22680015 ↗
  • MalfunctionJul 24, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE FOUND WITHIN A BROKEN INSTRUMENTS TUB AT HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22588623 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT THE SUCTION END OF THE INSERTER SPLIT IN TWO. THERE WAS NO DELAY TO CASE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925458 ↗
  • MalfunctionOct 7, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT DURING SURGERY, THE PLASTIC TIP/INSERT SUCTION CUP WAS DAMAGED. IT WAS WORN, STRIPPED AND DID NOT WORK ITS PURPOSE. THERE WAS A SURGICAL DELAY BECAUSE THE SURGEON HAD TO SWITCH TO ANOTHER TOOL THAT HE DISLIKED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20383451 ↗
  • MalfunctionSep 4, 2024Break; Unintended Movement; Break; Unintended Movement

    IT WAS REPORTED THAT THE SILICON SUCTION TIP CAME OFF IN STERILIZATION PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136102 ↗
  • MalfunctionSep 4, 2024Break; Unintended Movement; Break; Unintended Movement

    IT WAS REPORTED THAT THE SILICON SUCTION TIP CAME OFF IN STERILIZATION PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136100 ↗
  • MalfunctionAug 18, 2024Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    SUCTION CUP ON INSERT HOLDER IS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19999339 ↗
  • MalfunctionJun 19, 2024Break; Break

    IT WAS REPORTED THAT THE SUCTION CUP INSERTER FOUND BROKE INTO 2 PIECES. ALL FRAGMENTS WERE REMOVED. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19566872 ↗
  • MalfunctionJan 16, 2024Break; Break

    IT WAS REPORTED THAT THE SUCTION CUP IS TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18519965 ↗
  • MalfunctionJan 16, 2024Break; Break

    IT WAS REPORTED THAT THE SUCTION CUP IS TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18518501 ↗
  • MalfunctionJan 3, 2024Break; Break

    IT WAS REPORTED THAT THE SUCTION CUP FROM INSERTION HANDLE IS MISSING. IT HAPPENED BETWEEN DECONTAMINATION AND CLEAN SIDE OF CENTRAL STERILE, NOT IN SURGERY. NO ADVERSE CONSEQUENCES, AND NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18440934 ↗
  • MalfunctionFeb 28, 2023Break; Break

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON. DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16451422 ↗
  • MalfunctionJan 24, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THE INSTRUMENTS ARE WORN. DISCOVERY MADE BY STERILIZATION. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16233834 ↗
  • MalfunctionDec 7, 2022Break; Naturally Worn; Break; Naturally Worn

    WEAR AND TEAR BREAKAGE OF INSERT INSERTER - NO LONGER USABLE FOR SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15934379 ↗
  • MalfunctionDec 6, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT BOX OSTEOTOME IS DULL AND JAGGED. SUCTION CUP DOESN¿T STICK. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15924317 ↗
  • MalfunctionDec 2, 2022Break; Break

    IT WAS REPORTED THAT THE AB SUCTION CUP INSERTER TIP BROKE OFF OF THE HANDLE. THE HANDLE WAS BEING RETURNED. NO CASE TIME ADDED TO AS A RESULT. PATIENT OUTCOME NOT INPACTED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15902413 ↗
  • MalfunctionNov 8, 2022Break; Break

    IT WAS REPORTED THAT THE SUCTION CUP RIPPED IN WASH. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15756104 ↗
  • MalfunctionJul 25, 2022Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS ON THE PINNACLE INSTRUMENT SET BROKE. NO HARM TO THE PATIENT AND NO DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15094700 ↗
  • MalfunctionApr 5, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACETABULAR LINER SUCTION CUP WILL NOT HOLD LINER ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14013223 ↗
  • MalfunctionMar 2, 2022Break; Break

    IT WAS REPORTED THAT THE CERAMIC SUCCION CUP INSERTER (221700020) HAS BROKEN AND DOESN¿T SUCCION ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13651616 ↗
  • MalfunctionFeb 22, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THESE INSTRUMENTS ARE BEING REPORTED FOR NATURAL WEAR AND TEAR. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13577109 ↗
  • MalfunctionFeb 3, 2022Break; Break

    IT WAS REPORTED THAT THE RUBBER SUCKER FOR THE EXTRACTOR HAS A SPLIT IN IT. NEEDS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13440425 ↗
  • MalfunctionSep 24, 2021Break; Break

    THE STRIKE PLATE END OF THE HANDLE IS BROKEN OFF AND NEEDS TO BE REPLACED. IF OTHER, DESCRIBE: WAS NOTIFIED THROUGH MDRD DURING STERILIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12528448 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page