Skip to content

ESC PINNACLE 2218-00-010Recalls & reported problems

Depuy (Ireland) · PINNACLE ESC LINER EXTRACTOR TIP · Class 1

All models of ESC PINNACLE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle liner extractor tip is a surgical tool used to remove hip implant liners. Its 147 reports describe the tip breaking, bending, or deforming, and most report no effect on patients. The reports include no injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
151
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 151 adverse event reports that list this exact device (DI 10603295099505), received between Sep 8, 2017 and Sep 29, 2026 — classified as 151 malfunctions.

The most frequently reported specific device problem is “Break” (72 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 5 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 5 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 7 reportsDec 2025: 3 reportsJan 2026: 4 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break72
  • Material Deformation62
  • Material Twisted/Bent17
  • Bent9
  • Naturally Worn4
  • Crack2
  • Material Integrity Problem2
  • No Apparent Adverse Event2

Patient problems

  • No Clinical Signs, Symptoms or Conditions90
  • No Consequences Or Impact To Patient47
  • Insufficient Information10
  • No Information7
  • No Code Available3

Recent adverse event reports

The 25 most recent of 151 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936426 ↗
  • MalfunctionSep 11, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735152 ↗
  • MalfunctionSep 2, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50555087 ↗
  • MalfunctionSep 2, 2026Material Deformation

    It was reported that 2X ESC liner tips were bent while extracting a metal liner.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50555085 ↗
  • MalfunctionMay 26, 2026Break

    IT WAS REPORTED THAT DURING INSPECTION OF THE DEVICE IT WAS OBSERVED THAT THE TIP OF THE TOOL WAS BROKEN OFF AND UNABLE TO BE USED. IT WAS REPORTED THAT THE EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25279573 ↗
  • MalfunctionApr 9, 2026No Apparent Adverse Event

    IT WAS REPORTED THE DEVICE WAS FOUND BROKEN DURING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24832625 ↗
  • MalfunctionApr 7, 2026Break; Material Deformation

    IT WAS REPORTED THAT THE TIP OF AN EXTRACTOR WAS BENT AND BROKEN. THE EVENT DID NOT OCCUR DURING SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24802718 ↗
  • MalfunctionApr 1, 2026Break

    IT WAS REPORTED THAT THE TIP OF LINER REMOVAL TOOL WAS BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24756140 ↗
  • MalfunctionFeb 24, 2026Break

    IT WAS REPORTED THAT PINNACLE CONSTRAINED SPIKE TOOL BROKE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431133 ↗
  • MalfunctionFeb 16, 2026Break

    IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24343474 ↗
  • MalfunctionJan 23, 2026Material Deformation

    IT WAS REPORTED THAT THE TIP OF THE PINNACLE LINER EXTRACTOR IS BENT AND UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24164526 ↗
  • MalfunctionJan 16, 2026Material Deformation

    IT WAS REPORTED THAT TIP WAS BENT. CUSTOMER REQUESTED TO REPLACE THE PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24094634 ↗
  • MalfunctionJan 16, 2026Material Deformation

    IT WAS REPORTED THAT INSTRUMENT TIP WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093862 ↗
  • MalfunctionJan 6, 2026Break; Material Deformation

    THE REPORTED THAT THE PINN ESC LINER EXTRACTOR TIP WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23976169 ↗
  • MalfunctionDec 15, 2025Break; Break

    ADDITIONAL INFORMATION STATES 887907959411.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802903 ↗
  • MalfunctionDec 10, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS NOTICED TO BE BROKEN WHILE OPENING FOR THE CASE. IT DID NOT HAPPEN DURING THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23766886 ↗
  • MalfunctionDec 3, 2025Material Deformation; Material Deformation

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE TIP WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23703795 ↗
  • MalfunctionNov 19, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT WHILE TRYING TO REMOVE THE LINER, THE CONSTRAINED LINER REMOVAL TOOL WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23597023 ↗
  • MalfunctionNov 18, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TIPS ON BOTH INSTRUMENTS WERE BENT AND WORN. THE WAS NO HOSPITAL INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23581084 ↗
  • MalfunctionNov 18, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TIPS ON BOTH INSTRUMENTS WERE BENT AND WORN. THE WAS NO HOSPITAL INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23581058 ↗
  • MalfunctionNov 10, 2025Break; Break

    IT WAS REPORTED THAT TIP OF LINER EXTRACTOR WAS BROKEN OFF. DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512675 ↗
  • MalfunctionNov 5, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT THE EVENT DID NOT HAPPEN IN SURGERY. IT WAS NOTICED DURING MANUFACTURERS INVESTIGATION THAT THE INSTRUMENT WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478622 ↗
  • MalfunctionNov 5, 2025Break; Break

    IT WAS REPORTED THAT WHILE TRYING TO REMOVE THE CERAMIC LINER FROM THE PATIENT, THE TIP OF THE REMOVAL PIECE WAS BROKEN. THIS DID NOT IMPACT THE LENGTH OF THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476069 ↗
  • MalfunctionNov 5, 2025Break; Break

    IT WAS REPORTED THAT WHILE TRYING TO REMOVE THE CERAMIC LINER FROM THE PATIENT, THE TIP OF THE REMOVAL PIECE WAS BROKEN. THIS DID NOT IMPACT THE LENGTH OF THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476062 ↗
  • MalfunctionOct 28, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT 2 LINER EXTRACTION TIPS WERE BENT OR BROKEN DURING SURGERY. ALL PIECES WERE RETRIEVED AND THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23402689 ↗

Other EXTRACTOR devices

All EXTRACTOR devices →

Sources and updates