Skip to content

RECLAIM 2975-00-635Recalls & reported problems

Depuy (Ireland) · RECLAIM ASSEMBLY UPPER PULL ROD · Class 1

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Reclaim pull rod is part of a surgical extractor tool. Many reports describe parts bending, breaking, or jamming, and threads stripping. The 156 reports include no deaths and 2 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
157
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 157 adverse event reports that list this exact device (DI 10603295156017), received between Jul 26, 2017 and Aug 11, 2026 — classified as 2 injuries, 155 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (47 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility47
  • Naturally Worn31
  • Break30
  • Material Deformation21
  • Mechanical Jam20
  • Material Twisted/Bent18
  • Appropriate Device Problem Term/Code Not Available8
  • Crack4

Patient problems

  • No Clinical Signs, Symptoms or Conditions85
  • No Consequences Or Impact To Patient38
  • Insufficient Information14
  • No Information10
  • Not Applicable10
  • No Code Available8

Recent adverse event reports

The 25 most recent of 157 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2026Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214351 ↗
  • MalfunctionJun 16, 2026Material Deformation

    IT WAS REPORTED THAT THREADS ARE STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25519839 ↗
  • MalfunctionApr 15, 2026Break

    ASSEMBLY TOOL ON RECLAIM TENSIONER WAS BROKEN. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887772 ↗
  • MalfunctionApr 1, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE PART ARRIVED DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24756490 ↗
  • MalfunctionMar 30, 2026Mechanical Jam

    IT WAS REPORTED THAT THE FITTINGS OF THE PIECES WERE STUCK AFTER USE OF THE RECLAIM DEVICES AND COULD NOT BE UNCLIPPED. THE EVENT OCCURRED POST-SURGERY. THERE WERE NO COMPLICATIONS OR PATIENT HARM. THERE WAS NO SIGNIFICANT DELAY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24717465 ↗
  • MalfunctionMar 13, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE ROD WAS BENT AND IT WON¿T CAPTURE THE EYE GLASS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590967 ↗
  • MalfunctionMar 10, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSERTER FOR RECLAIM IS STICKING TOGETHER, AND UNABLE TO PROPERLY PUT TOGETHER OR TAKE APART. IT DID NOT HAPPENED IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557771 ↗
  • MalfunctionJan 14, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREW TOP OF THE ASSEMBLY TOOL BECAME COLD-WELDED AND WOULD NO LONGER TURN AS IT IS SUPPOSED TO. THERE WAS NO ADVERSE CONSEQUENCES TO THE PATIENT AND SURGICAL TIME WAS NOT EXTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24075819 ↗
  • MalfunctionDec 9, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE RECLAIM ASSEMBLY PULL ROD THREAD WAS NOT WORKING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756096 ↗
  • MalfunctionNov 5, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE THREADS WERE STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476000 ↗
  • MalfunctionOct 27, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED STATES THAT NOTHING BROKE SPECIFICALLY, JUST STUCK TOGETHER. INSTRUMENTS CANNOT BE USED IF NOT TAKEN A PART SO THEY WERE TAKEN OUT OF THE ROTATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23391636 ↗
  • MalfunctionOct 23, 2025Break; Break

    IT WAS REPORTED THAT THE RECLAIM ASSEMBLY TOOL BROKEN IN HALF. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23370586 ↗
  • MalfunctionOct 22, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT WHILE TAKING APART THE RECLAIM ASSEMBLY TOOL PULL ROD AFTER THE CASE FOR CLEANING THE THREADS WERE STUCK AND WAS UNABLE TO SCREW BACK TO ORIGINAL STARTING POINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23357748 ↗
  • MalfunctionOct 17, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE THREADS ON THE TENSIONING DEVICE WERE UNUSABLE. TEAM WAS ONLY ABLE TO IMPACT THE PROXIMAL BODY ONTO THE STEM BEFORE PUTTING THE LOCKING BOLT IN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23323782 ↗
  • MalfunctionSep 9, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT PATIENT UNDERWENT HIP SURGERY. THE ASSEMBLY TOOL AND ASSEMBLY PULL ROD KEPT COMING APART WHEN TRYING TO BE USED DURING SURGERY, AND THEN WOULDN'T COME APART AFTER THE CASE; BELIEVED TO BE CROSS THREADED. THERE WAS NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004329 ↗
  • MalfunctionAug 25, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN RECLAIM ASSEMBLY TOOL. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882816 ↗
  • MalfunctionJul 22, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE STAR DRIVER TIP IS STRIPPED. THE RELCAIM PIECE'S NUT IS STUCK AND WON'T RELEASE. THE ARTICULATING SURFACE HAS A BROKEN SPRING. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22566382 ↗
  • MalfunctionMay 22, 2025Material Deformation; Mechanical Jam; Material Deformation; Mechanical Jam

    IT WAS REPORTED THAT THE ASSEMBLY PULL ROD WAS BENT, THREADED CAP JAMMED IN THE TOP POSITION. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22072199 ↗
  • MalfunctionMay 8, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE DEVICE WAS STRIPPED DURING SURGERY AND WILL NOT COME APART AT THE TOP. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21988463 ↗
  • MalfunctionApr 21, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE RECLAIM DISASSEMBLY INSTRUMENT COULD NOT BE DISCONNECTED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21871764 ↗
  • MalfunctionMar 27, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE THREADS ON THE PRODUCT STRIPPED DURING THE CASE. THERE WAS NO PATIENT HARM OR SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21711016 ↗
  • MalfunctionMar 6, 2025Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT THE DEVICE WAS JAMMED UP. IT WOULDN¿T FULLY SCREW IN TO ASSEMBLE FINAL PROXIMAL BODY.

    Patient problems: Insufficient Information

    MAUDE report 21537824 ↗
  • MalfunctionJan 22, 2025Material Deformation; Mechanical Jam; Material Deformation; Mechanical Jam

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO DELAY AND NO ADVERSE CONSEQUENCES.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21208447 ↗
  • MalfunctionJan 17, 2025Mechanical Jam; Mechanical Jam

    THE PIECE OF INSTRUMENTATION ON WHICH THE HOURGLASS IS MOUNTED FOR THE COUPLING TORQUE OF THE STEM WITH THE PROXIMAL BODY DOES NOT WORK, THE THREADED CAP IS COMPLETELY STUCK AND CANNOT BE UNSCREWED, THE TORQUE CANNOT BE APPLIED AS IT SHOULD BE.

    Patient problems: Insufficient Information

    MAUDE report 21180771 ↗
  • MalfunctionNov 14, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT ASSEMBLY TOOL UPPER PULL ROD HAD DAMAGED THREADS AND WOULD NOT THREAD INTO DISTAL STEM DURING IMPLANT ASSEMBLY. TAPER WAS IMPACTED AND LOCKING BOLT ADDED. THERE WAS NO DELAY IN PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20689356 ↗

Other EXTRACTOR devices

All EXTRACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page