P.F.C. 86-5226Recalls & reported problems
Depuy (Ireland) · P.F.C. PRESS FIT REAMER EXTRACTOR · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy P.F.C. extractor is a surgical instrument. Reports mostly describe it bending or breaking, sometimes during use, with no deaths and one injury mentioned. There are 80 adverse event reports, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 80
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 80 adverse event reports that list this exact device (DI 10603295216766), received between Jul 11, 2017 and Jan 5, 2026 — classified as 1 injury, 79 malfunctions.
The most frequently reported specific device problem is “Break” (41 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break41
- Material Deformation13
- Material Twisted/Bent11
- Bent9
- Appropriate Device Problem Term/Code Not Available2
- Connection Problem2
- Naturally Worn2
- Crack1
Patient problems
- No Consequences Or Impact To Patient37
- No Clinical Signs, Symptoms or Conditions22
- No Information10
- Not Applicable7
- No Code Available4
- Insufficient Information3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 80 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 5, 2026Material Deformation
IT WAS REPORTED THAT THE PFC STEM TRIAL EXTRACT WAS BENT BUT DID NOT BREAK INTO TWO PIECES. THE EVENT DID NOT CAUSE ANY DELAY TO THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23961790 ↗MalfunctionJun 9, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ITEM WAS BENT AT THE TIP WHERE THE THREADS ARE LOCATED, THEREFORE WAS NOT THREADING WITH A STEM TRIAL. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22166504 ↗MalfunctionNov 5, 2024Break; Break
IT WAS REPORTED THAT A PART OF THE EXTRACTION INSTRUMENT BROKE APART. SURGICAL DELAY WAS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 20610146 ↗MalfunctionOct 16, 2024Break; Break
THE THREADED END HAS BROKEN OFF AND DAMAGED INTO THE STEM TRIAL.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20461975 ↗MalfunctionAug 22, 2024Material Deformation; Material Deformation
THIS INSTRUMENT IS COMPOSED OF 1 PART. THE KNOB OF THE INSTRUMENT IS BENT. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20030280 ↗MalfunctionAug 9, 2024Material Deformation; Material Deformation
THIS INSTRUMENT IS COMPOSED OF 1 PART. THE TOP OF THE INSTRUMENT IS BENT. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 19948241 ↗MalfunctionJul 3, 2024Material Deformation; Material Deformation
THIS INSTRUMENT IS COMPOSED OF 1 PART. THE TOP PART OF THE INSTRUMENT HAS A VISIBLE BENT NEAR THREADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19666288 ↗MalfunctionJul 1, 2024Break; Break
IT WAS REPORTED THAT THE SCREWDRIVER TIP BROKE OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19643994 ↗MalfunctionJun 14, 2024No Apparent Adverse Event; No Apparent Adverse Event
IT WAS REPORTED THAT THE INSTRUMENT HAS AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN DURING SURGERY. ADDITIONAL EVENT INFORMATION: VISUAL INSPECTION OF THE RETURNED SAMPLE FOUND THAT TIP OF THE DEVICE IS BROKEN OFF. BROKEN FRAGMENT WAS FOUND STUCK INSIDE A PFC* FLUT TIB ROD TRL 75 X 14.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19543479 ↗MalfunctionMar 15, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT THE TRIAL STEM EXTRACTOR TIP IS BENT. THE ALLEGATION DID NOT HAPPEN IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18916259 ↗MalfunctionFeb 16, 2024Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT THE INSTRUMENT IS BENT AND TIP OF INSTRUMENT IS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18720357 ↗MalfunctionDec 16, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT THE PFC STEM TRIAL EXTRACT WAS BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18341690 ↗MalfunctionSep 28, 2023Material Deformation; Material Deformation
THIS INSTRUMENT IS COMPOSED OF THE 1 PART. THE TIP OF THE INSTRUMENT HAS BEEN BENT. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17832854 ↗MalfunctionMar 17, 2023Break; Break
IT WAS REPORTED THAT SHAFT HAS BECOME DISSOCIATED FROM HANDLE WHERE THE WELDING CONNECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16563616 ↗MalfunctionMar 15, 2023Break; Break
IT WAS REPORTED THAT INSTRUMENT IS BROKEN. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16550699 ↗MalfunctionJan 12, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT THE INSTRUMENT IS BENT. NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16146863 ↗MalfunctionDec 22, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE SHAFT DISCONNECTED FROM GRIP BLASTED SHAFT HOLDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16033793 ↗MalfunctionSep 8, 2022Break; Break
IT WAS REPORTED THAT THE FLUTED ROD EXTRACTOR TIP BROKEN OFF FLUTED STEM END OF ROD EXTRACTOR BROKE INSIDE STEM. TIME ADDED ON ABOUT 20 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15380036 ↗MalfunctionDec 15, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE THREADS ARE STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13009691 ↗MalfunctionOct 19, 2021Break; Break
DURING KNEE REVISION SURGERY TODAY THE ITEMS WERE DAMAGED TRYING TO REMOVE A STEM TRIAL STUCK IN THE TIBIA. THE END OF THE EXTRACTOR THAT SCREWS INTO THE STEM TRIAL BROKE OFF INSIDE THE STEM TRIAL. NO SIGNIFICANT DELAY TO SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12662652 ↗MalfunctionOct 11, 2021Break; Break
IT WAS REPORTED THAT THE TIP OF STEM EXTRACTOR WAS BROKEN WHILE TRYING TO REMOVE THE 16 MM STEM TRIAL. ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605551 ↗MalfunctionApr 29, 2021Material Deformation; Material Deformation
ADDITIONAL INFORMATION RECEIVED WAS RECEIVED INDICATED THAT THE SLEEVE TAPER EXTRACTOR HAD STRIPPED THREADS. THE STEM WRENCH FUNCTION HAD NOT BEEN LUBRICATED APPROPRIATELY MAKING THE ACTION STIFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11749428 ↗MalfunctionMar 8, 2021Break; Break
IT WAS REPORTED THAT UPON INSPECTION OF INSTRUMENTS, IT WAS FOUND OUT THAT THE STEM TRIAL EXTRACTOR WAS BROKEN. THE INSTRUMENT WAS NOT USED IN THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11435461 ↗MalfunctionMar 4, 2021Crack; Crack
EMAIL RECEIVED FROM CSSD DEPARTMENT AT (B)(6) HOSPITAL WITH A LIST OF BROKEN INSTRUMENTS. ALL INSTRUMENTS REMAIN IN ONE PIECE. DAMAGE IS THROUGH CRACKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11413368 ↗MalfunctionOct 28, 2020Break; Break
STEM EXTRACTION TOOL TIP BROKE OFF INSIDE TRIAL STEM. ALL PIECES RECOVERED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10747984 ↗
FDA clearance and classification
The FDA regulates product code HWB (extractor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the P.F.C. 86-5226
Has the P.F.C. 86-5226 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10603295216766), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the P.F.C. 86-5226?
The FDA's MAUDE database lists 80 adverse event reports for this device, received between Jul 11, 2017 and Jan 5, 2026. In those reports, the most common specific device problems are “Break” (41), “Material Deformation” (13) and “Material Twisted/Bent” (11); the most common patient problem is “Not Applicable” (7). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the P.F.C. 86-5226?
None of the reports are classified as deaths, 1 as an injury and 79 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the P.F.C. 86-5226 FDA approved?
Devices of this type (product code HWB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).
Who makes the P.F.C. 86-5226, and is it still sold?
The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand P.F.C., model 86-5226). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR390 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR242 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR156 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR150 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR150 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR129 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR69 reports
- LPS 2987-72-030Depuy (Ireland) · EXTRACTOR66 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page