GLOBAL APG+ 2230-00-023Recalls & reported problems
Depuy (Ireland) · GLOBAL APG+ CURVED HANDLE · Class 1
All models of GLOBAL APG+ (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295102960), received between Jul 20, 2017 and May 10, 2023 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility7
- Appropriate Device Problem Term/Code Not Available3
- Naturally Worn3
- Break2
- Unintended Movement2
- Fitting Problem1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Consequences Or Impact To Patient3
- No Information2
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 10, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE BLACK HANDLE AND 48+0 GUIDE WERE BOTH BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16903911 ↗MalfunctionJan 30, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE CURVED HANDLE IS WORN OUT AT THE DISTAL TIP WHERE IT CONNECTS TO THE SIZER PIN GUIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16263420 ↗MalfunctionDec 15, 2022Structural Problem; Unintended Movement; Structural Problem; Unintended Movement
IT WAS REPORTED THAT THE SURGEON REALIZED THAT THE LOCKING MECHANISM OF THE APG+ CURBED HANDLE WAS NOT WORKING WHEN PUTTING THE PIN SIZER GUIDES ON THE HANDLE. SURGEON MAKE IT WORK AND WOBBLE LESS TO PIN HIS LOCATION. SURGERY WAS NOT AFFECTED AND CASE WENT AS PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15987014 ↗MalfunctionDec 5, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE HANDLE AND GUIDES ARE STRIPPED AND GUIDE WON¿T STAY ON. NEEDS REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15916550 ↗MalfunctionAug 25, 2022Device-Device Incompatibility; Device-Device Incompatibility
GLENOID GUIDE BROKE OFF INSIDE INSERTER HANDLE. SURGEON WAS FINISHED USING DEVICE SO IT DIDN'T IMPACT CASE BUT RENDERED THE HANDLE AND 52+3 GUIDE UNUSABLE FOR FUTURE CASES. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15295001 ↗MalfunctionJul 6, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE APG+ CURVED HANDLE DID NOT HOLD SIZER PIN GUIDES DURING CASE. ALSO SIZER PIN GUIDES 52+0 WAS WORN DOWN AND WOULD NOT STAY ON THE HANDLE. ALL INSTRUMENTS WERE DISPOSED OF BEFORE BEING ABLE TO RETURN BACK. ALL PIECES WERE RECOVERED. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14931164 ↗MalfunctionJul 5, 2022Naturally Worn; Unintended Movement; Naturally Worn; Unintended Movement
IT WAS REPORTED THAT THE HANDLE WAS VERY LOOSE ON THE SIZING DISK. A FEW DIFFERENT SIZE DISKS WERE TRIED AFTER THE CASE AND IT IS THE CURVED HANDLE THAT'S WORN/ BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14914495 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE APG CURVED HANDLE IS NO LONGER HOLDING APG SIZER PIN GUIDES. SURGERY WAS NOT DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14647956 ↗MalfunctionJul 30, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE HANDLE DID NOT HOLD ON TO THE SIZER DISK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12252452 ↗MalfunctionMay 13, 2021Break; Break
IT WAS REPORTED THAT WHILE PUTTING IN THE GUIDE WIRE AND WHEN HE WENT TO PULL THE TRAIL OFF THE TIP OF THE TRIAL BROKE IN THE HANDLE. ALL PARTS WERE REMOVED. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11820323 ↗MalfunctionMay 17, 2019Fitting Problem; Device-Device Incompatibility; Fitting Problem; Device-Device Incompatibility
IT WAS REPORTED THAT THE TRIAL WON'T STAY ON HANDLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8619643 ↗MalfunctionAug 9, 2018Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN GUIDE AND CURVED HANDLE STUCK TOGETHER. NOTHING LEFT IN PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7766943 ↗MalfunctionDec 14, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE 48 SIZING DISK IS NOT CONNECTING PROPERLY TO HANDLE. MAY BE DUE TO WEAR. NO SURGICAL DELAY; DOE: (B)(6) 2017.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7116699 ↗MalfunctionOct 27, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT INSIDE OF THE APG+ CURVED HANDLE WAS WORN AND WAS NOT HOLDING THE GUIDE CORRECTLY.
Patient problems: No Information
MAUDE report 6983584 ↗MalfunctionJul 20, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE HANDLE WAS WORN DOWN AND WOULD NOT HOLD WITHOUT SPINNING.
Patient problems: No Information
MAUDE report 6728209 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page