QUICKSET 2274-25-500Recalls & reported problems

Depuy Orthopaedics, Inc. · QUICKSET 3.8mm 1 PC FLEX DRILL BIT 25mm LGT · Class 1

All models of QUICKSET (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy QuickSet flexible drill bit is used in orthopaedic surgery. Most reports describe the bit breaking during use, and others mention it bending, twisting or failing to cut, with some reports noting that broken pieces were retrieved. The records include 1,999 reports, 9 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,978
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,978 adverse event reports that list this exact device (DI 10603295109129), received between Dec 17, 2015 and Sep 18, 2026 — classified as 9 injuries, 1,969 malfunctions.

The most frequently reported specific device problem is “Break” (1,320 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 12 reportsJan 2024: 6 reportsFeb 2024: 18 reportsMar 2024: 13 reportsApr 2024: 15 reportsMay 2024: 15 reportsJun 2024: 19 reportsJul 2024: 11 reportsAug 2024: 15 reportsSep 2024: 11 reportsOct 2024: 21 reportsNov 2024: 19 reportsDec 2024: 3 reportsJan 2025: 23 reportsFeb 2025: 10 reportsMar 2025: 15 reportsApr 2025: 7 reportsMay 2025: 23 reportsJun 2025: 11 reportsJul 2025: 21 reportsAug 2025: 20 reportsSep 2025: 12 reportsOct 2025: 17 reportsNov 2025: 18 reportsDec 2025: 32 reportsJan 2026: 24 reportsFeb 2026: 18 reportsMar 2026: 23 reportsApr 2026: 25 reportsMay 2026: 36 reportsJun 2026: 14 reportsJul 2026: 7 reportsAug 2026: 7 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202636
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1,320
  • Material Deformation296
  • Material Twisted/Bent157
  • Failure to Cut110
  • Dull, Blunt50
  • Entrapment of Device50
  • Bent43
  • Corroded31

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,085
  • No Consequences Or Impact To Patient621
  • Insufficient Information160
  • No Information114
  • No Code Available42
  • Not Applicable30
  • Foreign Body In Patient19
  • No Known Impact Or Consequence To Patient5

Recent adverse event reports

The 25 most recent of 1,978 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Break

    It was reported that during a total hip replacement surgery a drill bit broke while drilling. It broke into two pieces. The broken piece was found and removed from the patient. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819256 ↗
  • MalfunctionSep 16, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790684 ↗
  • MalfunctionSep 16, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790681 ↗
  • MalfunctionAug 14, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261375 ↗
  • MalfunctionAug 11, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214953 ↗
  • MalfunctionAug 11, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214600 ↗
  • MalfunctionAug 11, 2026Material Deformation

    It was reported that the 25mm drill bent during case and needs replaced. No adverse consequences no time added.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50213496 ↗
  • MalfunctionAug 7, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50174998 ↗
  • MalfunctionAug 7, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50174991 ↗
  • MalfunctionAug 7, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50174335 ↗
  • MalfunctionJul 15, 2026Material Deformation

    IT WAS REPORTED THE TIP OF THE ACTIS THREADED INSERTER WAS DEFORMED. THIS WAS OBSERVED DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25826471 ↗
  • MalfunctionJul 10, 2026Material Deformation

    IT WAS REPORTED THE DRILL BIT BENT AT THE COIL AREA AND FRAYED. ALL PIECES WERE REMOVED; THE PROCEDURE WAS COMPLETED WITH NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25777848 ↗
  • MalfunctionJul 9, 2026Break

    IT WAS REPORTED THE SURGEON DRILLED TO PREPARE FOR A SCREW AND THE DRILL BIT BROKE. NO ADVERSE EVENTS OR DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25764888 ↗
  • MalfunctionJul 1, 2026Material Deformation

    IT WAS REPORTED THAT X4 QUICKSET FLEXIBLE BITS 25 MM ARE BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681948 ↗
  • MalfunctionJul 1, 2026Material Deformation

    IT WAS REPORTED THAT X4 QUICKSET FLEXIBLE BITS 25 MM ARE BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681945 ↗
  • MalfunctionJul 1, 2026Material Deformation

    IT WAS REPORTED THAT X4 QUICKSET FLEXIBLE BITS 25 MM ARE BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681944 ↗
  • MalfunctionJul 1, 2026Crack

    IT WAS REPORTED THAT THE DEVICE CRACKED WHILE DRILLING FOR SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679344 ↗
  • MalfunctionJun 26, 2026Material Deformation

    IT WAS REPORTED THAT THE ISSUE DID NOT OCCUR DURING SURGERY; SPD NOTIFIED THAT TWO 25MM PINNACLE DRILL BITS WERE DISCARDED DUE TO BEING BENT AT THE TIP, WITH NOTHING FURTHER TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25638977 ↗
  • MalfunctionJun 24, 2026Material Deformation

    IT WAS REPORTED THAT 25MM DRILL WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25608798 ↗
  • MalfunctionJun 19, 2026Break

    IT WAS REPORTED THAT THE DRILL BIT BROKE INTO TWO PIECES DURING USE. THE BROKEN TIP WAS RETRIEVED AND NOTHING REMAINED IN THE PATIENT. THE PROCEDURE WAS FINISHED WITH ANOTHER DRILL BIT. THERE WAS NO DELAY OR EXTENDED SURGICAL TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25563150 ↗
  • MalfunctionJun 19, 2026Break

    IT WAS REPORTED THAT DRILL BIT BROKE. SILVER KNOB ON THE BACK OF THE PINNACLE IMPACTOR BROKE OFF. IT WAS BROKEN INTO TWO PIECES, ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25562965 ↗
  • MalfunctionJun 18, 2026Break

    IT WAS REPORTED THAT TWO PINNACLE DRILL BITS BROKE DURING SURGERY. THERE WAS NO SURGICAL DELAY AND ALL OF THE PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25547135 ↗
  • MalfunctionJun 16, 2026Material Deformation

    IT WAS REPORTED THAT A FLEXIBLE 25MM DRILL BIT BENT WHEN ATTEMPTING TO DRILL FOR A SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25521939 ↗
  • MalfunctionJun 16, 2026Break

    IT WAS REPORTED THAT FLEXIBLE SHAFT OF DRILL BIT STARTED TO FRAY; NO LOOSE PIECES WERE GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25518080 ↗
  • MalfunctionJun 16, 2026Break

    IT WAS REPORTED THE DRILL BIT BROKE DURING STERILIZATION; THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25517948 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the QUICKSET 2274-25-500

Has the QUICKSET 2274-25-500 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295109129), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the QUICKSET 2274-25-500?

The FDA's MAUDE database lists 1,978 adverse event reports for this device, received between Dec 17, 2015 and Sep 18, 2026. In those reports, the most common specific device problems are “Break” (1,320), “Material Deformation” (296) and “Material Twisted/Bent” (157); the most common patient problems are “Not Applicable” (30) and “Foreign Body In Patient” (19). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the QUICKSET 2274-25-500?

None of the reports are classified as deaths, 9 as injuries and 1,969 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the QUICKSET 2274-25-500 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy Orthopaedics, Inc.

Who makes the QUICKSET 2274-25-500, and is it still sold?

The FDA's device database lists Depuy Orthopaedics, Inc. as the company responsible for it (brand QUICKSET, model 2274-25-500). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BIT, DRILL devices

All BIT, DRILL devices →