Skip to content

LCS 2366-83-000Recalls & reported problems

Depuy (Ireland) · LCS HIP DRILL 45 DEGREES ANGLE · Class 1

All models of LCS (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy LCS angled hip drill has 205 malfunction reports, with no deaths and no injuries. Many reports describe the drill tip or bit breaking off or not staying in the driver, and some describe parts not fitting together. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
204
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 204 adverse event reports that list this exact device (DI 10603295120179), received between Jul 21, 2017 and Aug 21, 2026 — classified as 204 malfunctions.

The most frequently reported specific device problem is “Break” (93 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 9 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 5 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break93
  • Device-Device Incompatibility64
  • Mechanical Jam36
  • Naturally Worn14
  • Physical Resistance/Sticking9
  • Material Deformation6
  • Appropriate Device Problem Term/Code Not Available5
  • Difficult to Advance3

Patient problems

  • No Clinical Signs, Symptoms or Conditions127
  • No Consequences Or Impact To Patient49
  • Insufficient Information20
  • No Information6
  • No Code Available5
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 204 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2026No Apparent Adverse Event

    It was reported that the instrument had unknown allegation. It did not happen in surgery.Upon manufacturers investigation it was noted that instrument was deformed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50373502 ↗
  • MalfunctionJun 12, 2026Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT THE DRILL HAS LOOSENED AND NO LONGER HOLDS A DRILL BIT PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25486196 ↗
  • MalfunctionMay 19, 2026Break

    THE TOP SLIDE PIECE WAS BROKEN OFF COMPLETELY. 45 DEG ANGLE DRILL (236683000) : SIDE AFFECTED: NOT APPLICABLE ## REASON PRODUCT IS NOT AVAILABLE: AVAILABLE. 45 DEG ANGLE DRILL (236683000) : LOT/BATCH NUMBER: [REDACTED]. WAS THERE ANY REPORTED PATIENT OR USER HARM? NO . WAS THERE A SIGNIFICANT DELAY? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25221624 ↗
  • MalfunctionApr 21, 2026Break

    IT WAS REPORTED THAT WHEN THE TRAY WAS OPENED IT WAS OBSERVED THAT THE TIP WAS BROKEN OFF, NO OTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942560 ↗
  • MalfunctionApr 13, 2026Break

    IT WAS REPORTED THAT TIP OF DRILL POPPED OFF WHILE DRILLING FOR SCREWS. THROWN AWAY AND REQUESTED A NEW ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24862239 ↗
  • MalfunctionApr 10, 2026Break

    IT WAS REPORTED THAT THE DRILL BIT WAS BROKE INSIDE DRILL, AND THERE WAS NO PATIENT INVOLVEMENT RELATED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24839630 ↗
  • MalfunctionMar 24, 2026Break; Difficult to Advance

    IT WAS REPORTED THAT 45 DEG DRILL IS BROKEN. DRILL BIT WILL NOT SEAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675873 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE DRILL SHAFT WAS BROKEN. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24637015 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE DRILL SHAFT WAS BROKEN. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24637013 ↗
  • MalfunctionMar 16, 2026Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS FOUND BROKEN WHEN THE TRAY WAS OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24604030 ↗
  • MalfunctionMar 11, 2026Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING THE TOTAL HIP PROCEDURE. SURGEON HAD BACKUPS, SO THERE WAS NO DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24576018 ↗
  • MalfunctionJan 23, 2026Break

    IT WAS REPORTED THAT DURALOC DRILL AND ALSO T HANDLE BROKE DURING TOTAL HIP. T HANDLE WHITE HANDLE CRACKED, AND DURALOC DRILL PIECE THAT HOLDS DRILL BIT ALSO BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24159633 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT THE DURALOC ANGLED DRILL BROKEN. REQUIRED REPLACEMENT ASAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850408 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT THE DURALOC ANGLED DRILL WAS BROKEN. REQUIRED REPLACEMENT ASAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850273 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT AFTER THE SURGERY WHILE REASSEMBLING TRAYS IN STERILE PROCESSING DEPARTMENT THE DRILL DRIVER SHAFT CAME APART. ALL PARTS ARE ACCOUNTED FOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23705176 ↗
  • MalfunctionNov 10, 2025Break; Break

    IT WAS REPORTED THAT WHILE USING THE FIXED ANGLED DRILL SHAFT THE CENTER PIECE THAT HOLDS THE MODULAR DRILL BIT BROKE OFF. ALL PIECES WERE RECOVERED NO PATIENT HARM OR ADDED TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23510720 ↗
  • MalfunctionAug 29, 2025Break; Break

    IT WAS REPORTED THAT THE GUIDE POSTS FOR DRILL BIT ON DRILL ADAPTER PIECE WAS BROKEN/ NOT SECURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22923863 ↗
  • MalfunctionJul 21, 2025Break; Break

    IT WAS REPORTED THAT THERE WAS A BROKEN ANGLED DRILL AND DRILL BIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22556001 ↗
  • MalfunctionJul 7, 2025No Apparent Adverse Event; No Apparent Adverse Event

    THE ITEM WAS REPORTED FOR UNKNOWN REASON. THE EVENT DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22414143 ↗
  • MalfunctionJun 30, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22358462 ↗
  • MalfunctionJun 27, 2025Break; Break

    IT WAS REPORTED THAT WHEN THE DOCTOR WAS DRILLING FOR SCREWS DURING A TOTAL HIP REPLACEMENT, THE END PART OF THE DRILL BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22344574 ↗
  • MalfunctionJun 25, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT ON (B)(6) 2025. WHILE THE SURGEON WAS USING THE DRILL SHAFT, EVERY DRILL BIT THAT WAS PLACED IN THERE WAS WOBBLING. THE CONNECTOR WOULD NOT ALLOW THE DRILL BIT TO ADVANCE STRAIGHT INTO THE PELVIS. THE WAS NO SURGICAL DELAY AND ALL PIECES WERE RECOVERED FROM THE FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332182 ↗
  • MalfunctionJun 25, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT ON (B)(6) 2025. WHILE THE SURGEON WAS USING THE DRILL SHAFT, EVERY DRILL BIT THAT WAS PLACED IN THERE WAS WOBBLING. THE CONNECTOR WOULD NOT ALLOW THE DRILL BIT TO ADVANCE STRAIGHT INTO THE PELVIS. THE WAS NO SURGICAL DELAY AND ALL PIECES WERE RECOVERED FROM THE FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22331253 ↗
  • MalfunctionApr 25, 2025No Apparent Adverse Event; No Apparent Adverse Event

    Patient problems: Insufficient Information

    MAUDE report 21911630 ↗
  • MalfunctionFeb 25, 2025Break; Break

    IT WAS REPORTED THAT 3.8 DRILL BIT BROKE AND WAS LEFT IN PATIENT. 45 DEG ANGLE DRILL BROKE AND NEEDS TO BE REPLACED, STRAIGHT SCREWDRIVER WORN, NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21457500 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LCS 2366-83-000

Has the LCS 2366-83-000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295120179), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LCS 2366-83-000?

The FDA's MAUDE database lists 204 adverse event reports for this device, received between Jul 21, 2017 and Aug 21, 2026. In those reports, the most common specific device problems are “Break” (93), “Device-Device Incompatibility” (64) and “Mechanical Jam” (36); the most common patient problem is “Not Applicable” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LCS 2366-83-000?

None of the reports are classified as deaths, 0 as injuries and 204 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LCS 2366-83-000 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the LCS 2366-83-000, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand LCS, model 2366-83-000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page