NA 323.062Recalls & reported problems

Synthes GmbH · 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes drill bit with depth marks is a surgical bit used to drill holes in bone. The 310 reports mostly describe bits that break during drilling, fail to cut or get stuck in the bone, sometimes leaving fragments in the patient. There are no deaths and 80 injuries reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
308
Deaths reported
0
Injuries reported
80
Recalls naming it
0

Summary

The FDA's MAUDE database contains 308 adverse event reports that list this exact device (DI 10886982190758), received between Feb 15, 2017 and May 21, 2026 — classified as 80 injuries, 228 malfunctions.

The most frequently reported specific device problem is “Break” (195 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 5 reportsApr 2024: 0 reportsMay 2024: 6 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 5 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 6 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break195
  • Failure to Cut74
  • Entrapment of Device51
  • Dull, Blunt25
  • Material Fragmentation11
  • Material Deformation7
  • Device Operates Differently Than Expected6
  • Device-Device Incompatibility4

Patient problems

  • No Clinical Signs, Symptoms or Conditions121
  • Foreign Body In Patient76
  • No Consequences Or Impact To Patient54
  • Insufficient Information23
  • No Code Available21
  • No Known Impact Or Consequence To Patient10
  • Device Embedded In Tissue or Plaque9
  • No Patient Involvement6

Recent adverse event reports

The 25 most recent of 308 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 21, 2026Break

    THE CUSTOMER REPORTS THAT THE HANDLE BROKE DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241372 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THAT A 2.0MM DRILL BIT WITH DOUBLE MARKINGS BROKE WHILE DRILLING INTO THE BONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592384 ↗
  • MalfunctionMar 10, 2026Break

    IT WAS REPORTED THAT DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3 AND K-WIRE Ø1.6 W/DRILL TIP L200 WERE INVOLVED IN AN INTRAOPERATIVE EVENT. DURING THE REVIEW OF IMPLANTS AFTER SURGERY, THE K-WIRE WAS OBSERVED TO BE DOUBLED, AND THE DRILL BIT BROKE DURING THE PROCEDURE. THE DETACHED PART WAS REMOVED, AND THERE WAS NO IMPACT OR CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24558888 ↗
  • MalfunctionMar 9, 2026Break; Entrapment of Device

    IT WAS REPORTED THAT STAGBEETLE FORCEPS, RATCHET LOCK, LENGTH 125 MM, AND DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3 WERE INVOLVED IN AN INTRAOPERATIVE EVENT. DURING SURGERY, THE DRILL BIT BROKE WHILE DRILLING TO PLACE THE LAST SCREW, AND THE BROKEN FRAGMENT REMAINED INTRAMEDULLARY AS IT COULD NOT BE REMOVED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT CHANGES TO THE SCHEDULE, AND IT WAS INDICATED THAT THE RETAINED FRAGMENT WOULD NOT CAUSE PROBLEMS. IN A SEPARATE INCIDENT DURING THE SAME PROCEDURE, THE STAGBEETLE FORCEPS BROKE INTO SEVERAL PIECES WHILE REDUCING AN ELBOW FRAGMENT. THE BROKEN FORCEPS WERE IMMEDIATELY REPLACED WITH ANOTHER PAIR FROM THE SURGICAL KIT, ALLOWING THE PROCEDURE TO CONTINUE WITHOUT DELAY OR PATIENT IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24544971 ↗
  • MalfunctionMar 2, 2026Break

    THE DRILL ARRIVED WITHIN THE SET BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 24483017 ↗
  • MalfunctionJan 16, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THE DRILL BIT WAS SKEWED AND IT WAS DIFFICULT TO PASS THROUGH THE TOWER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24097045 ↗
  • MalfunctionJan 9, 2026Break; Entrapment of Device

    IT WAS REPORTED THAT TWO (2) 2.0 DRILL BITS BROKE IN HUMERUS DURING PROCEDURE. THE BROKEN DRILL BIT WAS LEFT IN THE PATIENT AND THE SURGEON DECIDED TO TAKE NO ACTION ON BROKEN DRILL BIT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24027986 ↗
  • MalfunctionJan 9, 2026Break; Entrapment of Device

    IT WAS REPORTED THAT TWO (2) 2.0 DRILL BITS BROKE IN HUMERUS DURING PROCEDURE. THE BROKEN DRILL BIT WAS LEFT IN THE PATIENT AND THE SURGEON DECIDED TO TAKE NO ACTION ON BROKEN DRILL BIT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24027984 ↗
  • MalfunctionDec 23, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED DURING THE PROCEDURE, WHEN DRILLING TO PLACE ONE OF THE 2.7 LOCKING SCREWS, THE 2.0 BIT BROKE, WITHOUT FINDING THE TIP. WE PROCEED TO TAKE X-RAY IMAGES, SHOWING THAT THE FRAGMENT IS INSIDE THE PATIENT'S BONE. DR. TRIES TO REMOVE THIS FRAGMENT, BUT IT WAS NOT POSSIBLE AND DECIDES TO LEAVE THIS TIP BECAUSE OF THE DIFFICULT EXTRACTION. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY.

    Patient problems: Foreign Body In Patient

    MAUDE report 23877930 ↗
  • MalfunctionDec 17, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3, DRILL BIT Ø2/1.15 CANN L150/48 3FLUTE F/, AND K-WIRE Ø1 L150 SST WERE INVOLVED IN AN INTRAOPERATIVE EVENT. DURING THE PROCEDURE, SHORTAGES OF A 2.0 DRILL BIT AND A 1.0 K-NEEDLE WERE IDENTIFIED UPON OPENING THE EQUIPMENT. LATER, THE 2.0 DRILL BIT BROKE, LEAVING ITS TIP EMBEDDED IN THE PATIENT'S BONE. THE FRAGMENT COULD NOT BE REMOVED AND WAS INTENTIONALLY LEFT IN PLACE, AS IT WAS DETERMINED THERE WOULD BE NO ADVERSE CONSEQUENCE. THE REMAINING PORTION OF THE DRILL BIT WAS REMOVED, AND A REPLACEMENT DRILL BIT WAS USED TO COMPLETE THE SURGERY WITHOUT DELAY OR DISCOMFORT TO THE SPECIALIST.

    Patient problems: Foreign Body In Patient

    MAUDE report 23829243 ↗
  • MalfunctionDec 12, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THE DRILL WAS COMPLETELY BENT. IT WAS NOTED DURING SURGERY. NO IMPACT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23788021 ↗
  • MalfunctionDec 10, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE 2.0 MM DRILL BITS BROKE DURING SURGERY; ONE PIECE REMAINED INSIDE THE PATIENT'S BONE, WHILE THE OTHER IS SENT INTACT. SCREWS WITH DAMAGED HEADS ARE NOT USED IN SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 23771259 ↗
  • MalfunctionDec 10, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE 2.0 MM DRILL BITS BROKE DURING SURGERY; ONE PIECE REMAINED INSIDE THE PATIENT'S BONE, WHILE THE OTHER IS SENT INTACT. SCREWS WITH DAMAGED HEADS ARE NOT USED IN SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 23771249 ↗
  • MalfunctionDec 4, 2025Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE DOCTOR WAS DRILLING INTO THE DISTAL TIBIA. THE PATIENT HAD SCREWS MADE OF A MATERIAL THAT BROKE WHEN THE PREVIOUS PLATE WAS REMOVED. THE DRILL BIT STRUCK ONE OF THESE SCREWS AND BROKE OFF AT THE TIP. THE BROKEN PIECE WAS REMOVED WITHOUT AFFECTING THE PATIENT AND REPLACED WITH ANOTHER DRILL BIT OF THE SAME TYPE, ALSO INCLUDED IN THE KIT. THE SURGERY WAS THEN SUCCESSFULLY COMPLETED WITHOUT DISCOMFORT FOR THE SPECIALIST, WITHOUT AFFECTING THE PATIENT, AND WITHOUT DELAYS TO THE SURGICAL SCHEDULE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23714320 ↗
  • MalfunctionNov 11, 2025Break; Break

    IT WAS REPORTED THAT WHEN CLEANING THE DRILL BIT IT HAD SNAPPED OFF THE DRILL. THE DRILL BIT ON THE TRAY WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23524258 ↗
  • MalfunctionSep 15, 2025Break; Break

    IT WAS REPORTED THAT THE 2.0MM DRILL BIT WAS FRACTURED DURING THE PROCEDURE. THE SURGERY WAS SUCCESSFULLY COMPLETED. THERE WAS NO DAMAGE TO THE PATIENT, ALL FRAGMENTS WERE RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23058083 ↗
  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT TWO DRILL BITS OF 2.0MM REFERENCE WERE BROKEN INTRA-OPERATIVELY; ONE OF THE PARTS OF THE DRILL BIT REMAINED IN THE PATIENT, DID NOT AFFECT THE PATIENT OR THE SURGICAL PROCEDURE WAS PROLONGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22963362 ↗
  • MalfunctionSep 4, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT TWO DRILL BITS OF 2.0MM REFERENCE WERE BROKEN INTRA-OPERATIVELY; ONE OF THE PARTS OF THE DRILL BIT REMAINED IN THE PATIENT, DID NOT AFFECT THE PATIENT OR THE SURGICAL PROCEDURE WAS PROLONGED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22963314 ↗
  • MalfunctionSep 4, 2025Break; Break

    IT WAS REPORTED THAT THE 2.0 BIT WAS BROKEN DURING A PROCEDURE. ANOTHER DRILL BIT WAS USED. THIS DID NOT AFFECT THE PATIENT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22963240 ↗
  • MalfunctionSep 2, 2025Break; Break

    IT WAS REPORTED THAT DURING THE PREOPERATIVE PERIOD, THE DRILL BIT BROKE IN TWO WHEN THE SCREWS WERE FITTED. THE TWO PIECES OF THE DRILL BIT WERE RECOVERED DIRECTLY. NO CONSEQUENCE FOR THE PATIENT BECAUSE THE DRILL BIT HAD NOT REMAINED IN SITU. RISK OF METAL BREAKAGE IN THE SURGICAL SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22939538 ↗
  • MalfunctionAug 28, 2025Break; Break

    IT WAS REPORTED THAT DURING THE PROCEDURE A 2.0MM DRILL BT BROKE WHILE DRILLING. THE BIT TOGETHER WITH ITS FRAGMENTS WAS REMOVED AND ANOTHER DRILL BIT OF THE SAME CHARACTERISTICS WAS USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT IMPACT AND NO ALTERATIONS IN THE SURGICAL TIMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22917437 ↗
  • MalfunctionAug 18, 2025Break; Entrapment of Device; Break; Entrapment of Device

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE SURGERY WAS COMPLETED SUCCESSFULLY AND THERE WAS NO SURGICAL DELAY REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22832570 ↗
  • MalfunctionJul 1, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT WAS OPERATED ON FOR TRAUMA OF THE LEFT UPPER LIMB FOLLOWING A FALL FROM A BICYCLE. DURING THE INTERVENTION (HUMERAL OSTEOSYNTHESIS) A DRILL BIT OF THE BROKE IN THE PATIENT. THE PIECES HAVE BEEN RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22370346 ↗
  • MalfunctionJun 30, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING SURGERY, THE SURGEON NOTICED THAT THE DRILL BIT WAS VERY DULL AND WAS NOT FUNCTIONING ADEQUATELY. THE ITEM WAS DEEMED TO BE NO LONGER FUNCTIONAL AND WAS SUBSEQUENTLY DISCARDED. THE SURGERY WAS NOT DELAYED, AND THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22358795 ↗
  • MalfunctionMay 13, 2025Break; Break

    IT WAS REPORTED THAT THERE IS A BROKEN DRILL BIT.

    Patient problems: Insufficient Information

    MAUDE report 22009404 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 323.062

Has the NA 323.062 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886982190758), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 323.062?

The FDA's MAUDE database lists 308 adverse event reports for this device, received between Feb 15, 2017 and May 21, 2026. In those reports, the most common specific device problems are “Break” (195), “Failure to Cut” (74) and “Entrapment of Device” (51); the most common patient problems are “Foreign Body In Patient” (76) and “Device Embedded In Tissue or Plaque” (9). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 323.062?

None of the reports are classified as deaths, 80 as injuries and 228 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 323.062 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.

Who makes the NA 323.062, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 323.062). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.