Skip to content

DELTA XTEND 2307-71-000Recalls & reported problems

Depuy (Ireland) · DELTA Xtend ORIENTATION PIN · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10603295116073), received between Nov 3, 2017 and Nov 5, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation10
  • Break3
  • Material Twisted/Bent3
  • Device-Device Incompatibility2
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information3
  • No Consequences Or Impact To Patient3
  • No Information1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A TOTAL SHOULDER REPLACEMENT PROCEDURE, IT WAS DISCOVERED THAT A PIECE OF THE ORIENTATION PIN WAS BENT. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WERE NO FRAGMENTS GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476436 ↗
  • MalfunctionOct 20, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE DELTA XTEND ORIENTATION PIN WAS BENT AND WILL NOT FIT THROUGH THE CUTTING GUIDE HANDLE HOLE OR TRIAL/IMPLANT/INTRODUCER. WITHOUT THE ORIENTATION PIN, THIS MAKES IT IMPOSSIBLE FOR THE SURGEON TO KNOW WHETHER HE HAS PLACED THE INSTRUMENTS IN THE CORRECT ORIENTATION. THIS WILL BE DETRIMENTAL TO THE FINAL POSITION AND IMPLANTATION AND MAY CAUSE IMPLANT FAILURE. THIS DID NOT DELAY SURGERY OR HARM THE PATIENT AS HE WAS ABLE TO USE THE 2.5MM GUIDE PIN INSTEAD, BEING CAUTIOUS OF THE POINTED TIP HOWEVER, SO AS NOT TO DAMAGE SURROUNDING SOFT TISSUE OR NERVES. DUE TO ONGOING USE OVER TIME, THE END OF THE DELTA XTEND LOCKING SCREWDRIVER INTERNAL ROD HAS BECOME ROUNDED AND MIS-SHAPED, MAKING IT VERY DIFFICULT ENGAGING AND INSERTING THE INTERNAL LOCKING SCREW. THIS DELAYED SURGERY BY ONLY 10 MINUTES UNTIL A SECOND INSTRUMENT TRAY WAS OPENED AND USED IN PLACE OF THE WORN INSTRUMENT. THERE WAS NO HARM TO THE PATIENT. DUE TO ONGOING USE OVER TIME, THE END OF THE DELTA XTEND 3.5MM CANNULATED HEX SCREWDRIVER HAS BECOME ROUNDED AND MIS-SHAPED, MAKING IT VERY DIFFICULT ENGAGING AND INSERTING THE NON-LOCKING LOCKING SCREW. THIS DID NOT DELAY SURGERY AS A SECOND INSTRUMENT TRAY W

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335649 ↗
  • MalfunctionMay 12, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ORIENTATION PIN (VERSION ROD) ON DELTA EXTEND WAS FOUND TO BE BENT ON SET INSPECTION PRIOR TO CASE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22004017 ↗
  • MalfunctionMar 27, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025. WHEN OPENING THE SETS PRIOR TO SURGERY, IT WAS DETERMINED THAT THE DEPTH GAUGE AND VERSION ROD WERE BOTH BENT.

    Patient problems: Insufficient Information

    MAUDE report 21708055 ↗
  • MalfunctionNov 18, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THE ORIENTATION GUIDE PIN BENT. IT WAS NOTED THE HOSPITAL DISCARDED IT. THERE WAS PATIENT CONSEQUENCES NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20708934 ↗
  • MalfunctionFeb 29, 2024Material Deformation; Material Deformation

    ADDITIONAL INFORMATION WAS RECEIVED AND IT WAS STATED THAT THE OPERATION WAS ON THE (B)(6) AND THAT THERE WAS NO DELAY TO THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18812818 ↗
  • MalfunctionFeb 9, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT PIN IS BENT. NO PIECES WERE LEFT INSIDE PATIENT. SURGICAL DELAY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18675506 ↗
  • MalfunctionFeb 6, 2024Break; Break

    IT WAS REPORTED THAT THE PIN BROKE NO PIECES WERE LEFT IN PATIENT. IT WAS UNKNOWN ABOUT SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18650953 ↗
  • MalfunctionJan 17, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT ORIENTATION PIN BENT. PROXIMAL REAMING GUIDE HANDLE WAS LOOSE AND WOULD NOT HOLD.

    Patient problems: Insufficient Information

    MAUDE report 18524958 ↗
  • MalfunctionJun 26, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ORIENTATION PIN IS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199736 ↗
  • MalfunctionDec 21, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ORIENTATION PIN WAS BENT DURING THE CASE AND WAS NOT ABLE TO BE USED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16023647 ↗
  • MalfunctionMay 18, 2022Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE INSTRUMENTS ARE BROKEN AND DAMAGED FROM WASH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14428809 ↗
  • MalfunctionJan 28, 2020Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE BLUE ROD GOT BENT MAKING IT DIFFICULT TO ALIGN THE INSTRUMENTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9637701 ↗
  • MalfunctionOct 25, 2018Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE ORIENTATION PIN WAS BENT DURING INSERTION OF STEM TRIAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8006379 ↗
  • MalfunctionAug 14, 2018Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE VERSION ROD WAS BENT AND DISCARDED BY THE HOSPITAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7781039 ↗
  • MalfunctionNov 3, 2017Break; Break

    IT WAS REPORTED THAT BLUE PIN GUIDE CRACKED AND BROKE IN TWO PIECES.

    Patient problems: No Information

    MAUDE report 6998911 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →