DELTA XTEND 2307-90-004Recalls & reported problems

Depuy (Ireland) · DELTA XTEND SCREW GUIDE PIN 1.2 · Class 1

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295116530), received between May 28, 2020 and Dec 4, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Entrapment of Device2
  • Component Missing1
  • Device-Device Incompatibility1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE GLENOSPHERE GUIDE WIRE SNAPPED OFF AND THE SURGEON WAS ABLE TO RETRIEVE IT. FRAGMENTS WERE GENERATED. IT WAS FURTHER REPORTED THAT THE DISTAL PORTION OF A CANNULATED 3.5-MM HEX SCREWDRIVER WAS DAMAGED (CROSS-THREADED AND BENT). THERE WAS TEN-MINUTE DELAY TO THE SURGICAL PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23719931 ↗
  • MalfunctionJul 21, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE INSERTING SCREWS, ONE OF GUIDE PINS BECAME STUCK IN DRILL HOLE. WHILE ATTEMPTING TO REMOVE IT, THE PIN WAS BROKE IN HALF. AFTER AN ATTEMPT TO REMOVE THE PIECE THAT WAS STILL IN THE DRILL HOLE, IT WAS DETERMINED BY THE SURGEON TO PROCEED WITH THE CASE, AS IT WOULD POSE MORE OF AN ISSUE TO CONTINUE TO ATTEMPT TO REMOVE PIN. THE REMAINDER OF THE PIN WAS LEFT IN THE PATIENT AND THE PROCEDURE WAS FINISHED. THE PIN WAS THEN REMOVED ON JULY 1 WITHOUT ANOTHER ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22554246 ↗
  • MalfunctionJan 27, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2025 DURING A RSA SURGERY PERFORMED ON SHOULDER FOR CTA, A PART OF SCREW GUIDE PIN BROKE AND REMAINED INSIDE THE BODY (IN THE BONE). DETAILS WERE AS FOLLOWS. AFTER THE METAGLENN WAS PLACED, A DRILL AND A DRILL GUIDE WERE USED TO CREATE THE PRE-HOLE. THE SCREW GUIDE PIN IN QUESTION WAS PLACED IN THE BONE HOLE AND A LOCKING SCREW WAS INSERTED. THE SURGEON THEN ATTEMPTED TO REMOVE THE SCREW GUIDE PIN IN QUESTION BUT COULD NOT DO SO MANUALLY. THE SCREW GUIDE PIN IN QUESTION COULD NOT BE REMOVED EVEN WITH A PIN REMOVER, IT WAS FINALLY REMOVED BY GRASPING IT WITH THE JACOBS CHUCK OF THE HANDPIECE AND ROTATING IT. HOWEVER, A POSTOPERATIVE X-RAY SHOWED THAT A BROKEN SCREW GUIDE PIN REMAINED AT THE TIP OF THE SCREW. THE SURGEON USED THE IMAGE TO CHECK AT DIFFERENT ANGLES, CONFIRMED THAT THE SCREW GUIDE PIN WAS LIKELY INSIDE THE BONE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 21243654 ↗
  • MalfunctionMay 28, 2020Break; Entrapment of Device; Component Missing; Material Twisted/Bent; Break; Entrapment of Device; Component Missing; Material Twisted/Bent

    IT WAS REPORTED THAT SURGEON WAS ON 4TH SCREW WITH SAME GUIDE PIN. GUIDE PIN WENT IN AND SCREW WAS SEATED AND PIN WAS BOUND. USED K. WIRE DRIVER TO REMOVE AND APPROX 1.5 INCH WAS MISSING, TWISTED OFF. USED X-RAY TO SEARCH, LOCATED REMAIN BUT NOT REACHABLE PER DOCTOR. NOTHING FURTHER.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 10097389 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates