NA 2307-96-000Recalls & reported problems
Depuy (Ireland) · GLENO GUIDE PIN D1.5 LG 300 MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295116608), received between Jun 19, 2020 and Feb 20, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Device-Device Incompatibility1
- Entrapment of Device1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE GLENOSPHERE WAS INSERTED OVER GLENOSPHERE GUIDE WIRE AND WHEN THE GLENOSPHERE SCREW WAS ENGAGED INTO METAGLENE. WHEN ATTEMPTED TO PULL THE GUIDE PIN, IT WAS STUCK IN THE GLENOSPHERE. SURGEON STRUGGLED, BUT WAS ABLE TO REMOVE GUIDEWIRE. THERE WAS A 10 MINUTES OF SURGICAL DELAY. DOE: (B)(6) 2023. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16403863 ↗MalfunctionMar 15, 2022Material Deformation; Material Deformation
THE PINS ARE BENT IN THIS KIT. THE DRILL IS COMPLETELY DULL. ISSUES WERE DISCOVERED WHEN INSTRUMENTS ARRIVED TO THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13771424 ↗MalfunctionJun 19, 2020Break; Entrapment of Device; Break; Entrapment of Device
RESISTANCE WAS FELT WHEN THE SURGEON WAS ATTEMPTING TO REMOVE THE GLENOSPHERE GUIDEPIN BY HAND. UPON SUCCESSFUL REMOVAL OF THE GUIDEPIN USING THE TRS MODULAR DRIVE, IT WAS NOTED THAT THE REMOVED PIN IS SHORTER THAN IT¿S ORIGINAL LENGTH. UNSCREWED GLENOSPHERE TO ATTEMPT TO FIND THE BROKEN PART OF THE BIT, HOWEVER, UNABLE TO LOCATE IT. INTERVENTIONAL IMAGING DONE AND PIN SEEN ON IMAGING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10176130 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page