PROSTALAC 2541-43-000Recalls & reported problems
Depuy Orthopaedics, Inc. · PROSTALAC QUICK RELEASE NUT · Class 1
All models of PROSTALAC (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295129554), received between Jan 15, 2020 and Jan 18, 2022 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility7
- Break4
- Mechanical Problem3
- Material Twisted/Bent2
- Naturally Worn2
Patient problems
- No Consequences Or Impact To Patient9
- No Code Available4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 18, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE TWO OF THE PROSTALAC MOLD QUICK RELEASE NUTS ARE STRIPPED AND NO LONGER THREAD ONTO THE MOLD BOLTS. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13283944 ↗MalfunctionJan 18, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE TWO OF THE PROSTALAC MOLD QUICK RELEASE NUTS ARE STRIPPED AND NO LONGER THREAD ONTO THE MOLD BOLTS. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13283800 ↗MalfunctionOct 15, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PART THAT WAS BROKEN WAS THE THREADED PART OF THE ASSEMBLY. THE BROKEN PART STAYED INSIDE OF THE STEM. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10684287 ↗MalfunctionOct 15, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PART THAT WAS BROKEN WAS THE THREADED PART OF THE ASSEMBLY. THE BROKEN PART STAYED INSIDE OF THE STEM. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10684276 ↗MalfunctionOct 15, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PART THAT WAS BROKEN WAS THE THREADED PART OF THE ASSEMBLY. THE BROKEN PART STAYED INSIDE OF THE STEM. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10684264 ↗MalfunctionOct 15, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PART THAT WAS BROKEN WAS THE THREADED PART OF THE ASSEMBLY. THE BROKEN PART STAYED INSIDE OF THE STEM. NO BROKEN PIECES WERE LEFT INSIDE THE PATIENT.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10684253 ↗MalfunctionApr 8, 2020Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT THE PROSTALAC QK REL NUT WAS STRIPPED ON THE INSIDE AND WOULD NOT SCREW TIGHTLY ONTO THE MOLDS. THERE WAS NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9934242 ↗MalfunctionApr 8, 2020Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT THE PROSTALAC QK REL NUT WAS STRIPPED ON THE INSIDE AND WOULD NOT SCREW TIGHTLY ONTO THE MOLDS. THERE WAS NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9934216 ↗MalfunctionJan 15, 2020Mechanical Problem; Device-Device Incompatibility; Mechanical Problem; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED THAT THE PROSTALAC NUTS DON¿T LOCK AND COULD NEVER GET THEM TO TIGHTEN DOWN CORRECTLY. THEY WOULD JUST SPIN ON THE THREADS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9592202 ↗MalfunctionJan 15, 2020Mechanical Problem; Device-Device Incompatibility; Mechanical Problem; Device-Device Incompatibility
THEY WERE THROWN AWAY BY THE ACCOUNT. THEY DON¿T LOCK BUT WE COULD NEVER GET THEM TO TIGHTEN DOWN CORRECTLY. THEY WOULD JUST SPIN ON THE THREADS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9592196 ↗MalfunctionJan 15, 2020Mechanical Problem; Device-Device Incompatibility; Mechanical Problem; Device-Device Incompatibility
THEY WERE THROWN AWAY BY THE ACCOUNT. THEY DON¿T LOCK BUT WE COULD NEVER GET THEM TO TIGHTEN DOWN CORRECTLY. THEY WOULD JUST SPIN ON THE THREADS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9592195 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page