ATTUNE 2544-00-111Recalls & reported problems
Depuy (Ireland) · ATTUNE PINNING SYSTEM · Class 1
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy ATTUNE pinning system is used to hold bone in place during orthopedic surgery, such as knee procedures. The 35 reports mostly describe pins that do not fit or slide properly into the matching DePuy pin driver, pins bending, sticking or snapping, and cutting jig parts breaking or jamming during surgery, sometimes complicating the operation. Four reports mention injuries and no deaths. There are no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 42
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 10603295129943), received between Jun 20, 2018 and Sep 10, 2026 — classified as 4 injuries, 38 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility26
- Break15
- Adverse Event Without Identified Device or Use Problem4
- Mechanical Jam2
- Physical Resistance/Sticking2
- Defective Device1
- Device Contamination with Chemical or Other Material1
- Device Damaged Prior to Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions31
- Insufficient Information4
- No Code Available3
- No Consequences Or Impact To Patient3
- Unspecified Infection2
- Hypersensitivity/Allergic reaction1
- Joint Laxity1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713260 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713243 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713234 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713222 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713218 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713215 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713213 ↗MalfunctionSep 10, 2026Device-Device Incompatibility
It was reported that some pins would not fit in the pin driver, others were tight, and some were fine. It was unknown if there was a delay in surgery, successful completion of the surgery. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713182 ↗MalfunctionJun 25, 2026Device-Device Incompatibility
IT WAS REPORTED THE PINS IN THE THREADED HEAD DID NOT FIT THE ATTUNE PIN DRIVER. A NEW PACK OF PINS WAS OPENED AND THEY FIT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25618962 ↗MalfunctionJun 25, 2026Device-Device Incompatibility
IT WAS REPORTED THE PINS IN THE THREADED HEAD DID NOT FIT THE ATTUNE PIN DRIVER. A NEW PACK OF PINS WAS OPENED AND THEY FIT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25618958 ↗MalfunctionJun 23, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN DRIVERS WERE RETAINING PINS IMPROPERLY CAUSING PINS TO PULL OUT AND STICK. AS A RESULT ONE PIN WAS COLD WELDED TO AN LEFT CUT BLOCK. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25598117 ↗MalfunctionJun 23, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN DRIVERS WERE RETAINING PINS IMPROPERLY CAUSING PINS TO PULL OUT AND STICK. AS A RESULT, ONE PIN WAS COLD WELDED TO AN LEFT CUT BLOCK. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25598110 ↗MalfunctionApr 8, 2026Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE SCREW SNAPPED INSIDE SCREW HOLE OF DISTAL CUTTING JIG, UNABLE TO GET THE SCREW OUT AND THE CUT WAS ALREADY MADE SO NO ADDITIONAL MEASURES HAD TO BE TAKEN. REPLACED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24816914 ↗MalfunctionSep 22, 2025Device-Device Incompatibility; Device-Device Incompatibility
PATIENT UNDERWENT KNEE SURGERY ON (B)(6) 2025. INTRAOPERATIVELY, A PIN FROM THE ATTUNE PINNING SYSTEM REPEATEDLY GOT STUCK IN THE CUTTING BLOCKS (BENT, METAL NOSE, OR BOTH). OTHER PINS HAVE BEEN USED ON THE BLOCKS AND WORKED PERFECTLY WELL. THIS COMPLICATED THE SURGERY BUT HAD NO IMPACT ON THE OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23109643 ↗MalfunctionJan 8, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
THREADED HEADED ATTUNE PIN HEAD SNAPPED OFF INSIDE THE HP PIN DRIVER. IT WAS UNKNOWN IF THE SURGERY WAS COMPLETED SUCCESSFULLY OR DELAYED. IT WAS ALSO UNKNOWN IF THERE WERE FRAGMENTS GENERATED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21103147 ↗MalfunctionDec 13, 2024Break; Break
IT WAS REPORTED THAT THE PIN BROKE OFF INTO TWO PARTS WHEN FIRST TIME IT WAS SCREWED INTO THE FEMUR. ALL PARTS HAVE BEEN RECOVERED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. NO OTHER MEDICAL INTERVENTION WAS REQUIRED. THERE WAS NO PATIENT HARM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20921912 ↗InjuryOct 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6), STUDY NO: (B)(6). CLINICAL ADVERSE EVENTS RECEIVED FOR POSITIVE TISSUE BIOPSY X2 KNEE SYNOVIUM. DEVICE AND PROCEDURE (RELATEDNESS): DEVICE RELATED: DEFINITELY NOT, PROCEDURE RELATED: POSSIBLY, DATE OF EVENT: 15 JAN 2019, DATE OF IMPLANT: (B)(6) 2019, DATE OF REVISION: NO INFORMATION PROVIDED, DEVICE LOCATION: RIGHT. TREATMENT/IMPACT: INJECTED MEDICATION.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 20507018 ↗MalfunctionJul 3, 2024Break; Break
IT WAS REPORTED THAT BROKEN PIN WOULDN'T FREE FROM CUTTING BLOCK. CHECK AGAINST OTHER PINS ¿ THIS REVEALED THAT NONE OF THE PINS WOULD FIT CORRECTLY AND LIKELY THE METAL HOLE HAD PERHAPS WARPED OUT OF SHAPE AND WOULD NOT FUNCTION CORRECTLY. THE SAME PINS WERE COMPARED TO THE RIGHT CUTTING BLOCK ON THIS TRAY AND PERFORMED AS INTENDED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19664517 ↗MalfunctionJan 16, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
FIXATION PIN BROKE IN THE CUTTING BLOCK AND IS JAMMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18520035 ↗MalfunctionNov 30, 2023Break; Break
TINY SHARDS OF ATTUNE THREADED PINS BROKE OFF INTO THE WOUND WHEN USING THE ATTUNE PS CUTTING BLOCKS DURING 2 X TOTAL KNEE REPLACEMENTS. SURGEON REMOVED THE FRAGMENTS FROM THE WOUND. SURGERY DELAYED LESS THAN 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18236882 ↗MalfunctionNov 30, 2023Break; Break
TINY SHARDS OF ATTUNE THREADED PINS BROKE OFF INTO THE WOUND WHEN USING THE ATTUNE PS CUTTING BLOCKS DURING 2 X TOTAL KNEE REPLACEMENTS. SURGEON REMOVED THE FRAGMENTS FROM THE WOUND. SURGERY DELAYED LESS THAN 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18236880 ↗MalfunctionNov 29, 2023Break; Break
TINY SHARDS OF ATTUNE THREADED PINS BROKE OFF INTO THE WOUND WHEN USING THE ATTUNE PS CUTTING BLOCKS DURING 2 X TOTAL KNEE REPLACEMENTS. SURGEON REMOVED THE FRAGMENTS FROM THE WOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18227127 ↗MalfunctionNov 29, 2023Break; Break
TINY SHARDS OF ATTUNE THREADED PINS BROKE OFF INTO THE WOUND WHEN USING THE ATTUNE PS CUTTING BLOCKS DURING 2 X TOTAL KNEE REPLACEMENTS. SURGEON REMOVED THE FRAGMENTS FROM THE WOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18227108 ↗MalfunctionSep 5, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT INTRAOPERATIVELY THE PINS SEEMED TO BE TOO THICK AND DID NOT FIT. NO SURGICAL DELAY, NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17686352 ↗MalfunctionAug 7, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ONE OF THE PIN HOLES DID NOT FIT THE ATTUNE PINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17469171 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page