SUMMIT 2570-03-300Recalls & reported problems
Depuy (Ireland) · SUMMIT NECK SEGMENT SIZE 6/7 STD · Class 1
All models of SUMMIT (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295143116), received between Jul 5, 2018 and Aug 20, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Break2
- Material Twisted/Bent2
- Corroded1
- Detachment of Device or Device Component1
- Illegible Information1
- Material Discolored1
- Naturally Worn1
Patient problems
- No Consequences Or Impact To Patient4
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 20, 2024Break; Device-Device Incompatibility; Illegible Information; Break; Device-Device Incompatibility; Illegible Information
IT WAS REPORTED THAT NECK TRIAL DOES NOT SIT FLUSH ON BROACH. CHIPPED AT BOTTOM. FADED AND UNABLE TO READ LETTERING. REQUESTED NEW ONE. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 20000856 ↗MalfunctionJul 2, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE NECK TRIAL DID NOT SEAT FLUSH ON 6 BROACH. BROACH NUB SEEMED NOT DAMAGED.
Patient problems: Insufficient Information
MAUDE report 19654209 ↗MalfunctionJun 12, 2024Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
NECK TRIAL LOCKING MECHANISM CAME OFF AND HEAD TRIAL NO LONGER LOCKS ON.
Patient problems: Insufficient Information
MAUDE report 19518664 ↗MalfunctionDec 19, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE NECK TRIAL NO LONGER ATTACHES TO BROACHES. NO PIECES WERE BROKEN OFF. NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16009467 ↗MalfunctionAug 3, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE LOCKING RING WAS MISSING FROM THE TRIAL NECK SEGMENT. NO DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268170 ↗MalfunctionMar 12, 2020Material Twisted/Bent; Material Twisted/Bent
MalfunctionMar 12, 2020Material Twisted/Bent; Material Twisted/Bent
MalfunctionJul 5, 2018Corroded; Corroded
THE MATERIALS PRESENTED OXIDATION. (B)(6). D257003150 NECK 2/3 HI LOT MT47132, D258003400 NECK 8/9 ST LOT MT447377, D2577003050 NECK 0/1 HI LOT MT40595, D267003200 NECK 4/5 ST LOT MT47047, D257003300 NECK ST 6/7 LOT MT47294, D24400549 STRAWBERRY 49 LOT S02019713, D24400542 STRAWBERRY 42 LOT S02015959.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7663626 ↗MalfunctionJul 5, 2018Material Discolored; Material Discolored
PRODUCTS PRESENTED OXIDATION. IN THE EMAIL RECEIVED, CAME A LIST OF PRODUCTS SEPARATED BY BOX. THIS COMPLAINT REFERS TO THE PRODUCTS OF (B)(6) IN BOX 1. (B)(6). PRODUCT: 257003050, LOT: MT40595. PRODUCT: 257003250, LOT: MT47299. PRODUCT: 257003350, LOT: MT45800. PRODUCT: 257003200, LOT: MT47074. PRODUCT: 257003300, LOT: MT47294.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7663420 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page