Skip to content

RECLAIM 2975-00-745Recalls & reported problems

Depuy (Ireland) · RECLAIM DISASSEMBLY TORQUE SHAFT 95mm · Class 1

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295156123), received between Dec 4, 2019 and May 12, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility3
  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1
  • Not Applicable1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT INSTRUMENTS BROKE DURING EXPLANT. THE PROXIMAL BODY OF RECLAIM WAS COLD WELDED ONTO THE DISTAL STEM. THE DISTAL STEM DID NOT COME OUT. PATIENT WAS NOT HARMED. NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25148567 ↗
  • MalfunctionNov 9, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PUSH ROD TIP WAS BENT, FORCED INTO TRQ SHFT 95, AND NOW STUCK TOGETHER. LOCKING TIP MISSING FROM BODY COLLET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12779227 ↗
  • MalfunctionMay 12, 2021Break; Break

    IT WAS REPORTED THAT THE HEAD WAS REVISED FOR INFECTION. ATTEMPTS WERE MADE TO REVISE RECLAIM PROXIMAL BUT THE DISASSEMBLY TOOL BROKE IN DIFFERENT PIECES. NO PIECES WERE LEFT IN THE PATIENT. SURGERY TIME WAS EXTENDED FOR 1 HR. DOI: (B)(6) 2020, DOR: (B)(6) 2021, RIGHT HIP.

    Patient problems: Insufficient Information

    MAUDE report 11813789 ↗
  • MalfunctionAug 14, 2020Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS BROKE WHILE TRYING TO DISENGAGE PROXIMAL BODY. THERE WAS AN (1) ONE HOUR OF SURGICAL DELAY.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 10410755 ↗
  • MalfunctionDec 4, 2019Device-Device Incompatibility; Device-Device Incompatibility

    THE SCRUB TECH WAS TRYING TO GRIP THE IMPLANT WITH THE DISASSEMBLY TOOL AND THE TOOL WAS NOT GRIPPING (HOLDING ONTO) THE IMPLANT CORRECTLY. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES. IF YES, NUMBER OF MINUTES: 10. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? NO.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9411713 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page