S-ROM 53-4153Recalls & reported problems

Depuy (Ireland) · S-ROM HIP PILOT SHAFT 13mm · Class 1

All models of S-ROM (185)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295171751), received between Nov 28, 2017 and Apr 24, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn2
  • No Apparent Adverse Event2
  • Appropriate Device Problem Term/Code Not Available1
  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Hip Fracture1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THESE INSTRUMENTS HAVE REACHED END OF LIFE. THE SURGEONS HAD CONTINUALLY COMPLAINED ABOUT WEAR AND TEAR OF INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24981745 ↗
  • MalfunctionApr 24, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THESE PRODUCT HAVE REACHED END OF LIFE DUE TO WEAR AND TEAR. THE SURGEONS ARE COMPLAINING ABOUT INSTRUMENT BEING DULL, ILLEGIBLE, AND GENERALLY WORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24981334 ↗
  • MalfunctionJul 20, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE SROM 18X13 PILOT SHAFT STEM TRIAL THREADS ARE WORN. NO ADDITIONAL SURGERY TIME REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15064503 ↗
  • MalfunctionJul 20, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE SROM 18X13 PILOT SHAFT STEM TRIAL THREADS ARE WORN. NO ADDITIONAL SURGERY TIME REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15064502 ↗
  • MalfunctionFeb 5, 2018Component Missing; Component Missing

    AT STERILIZATION, THEY NOTICED THAT A PIECE WAS MISSING, THE PIN THAT HOLDS THE SCREWED END IN PLACE. PATIENT CONSEQUENCE? :NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7240464 ↗
  • InjuryNov 28, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    PROCEDURE: TOTAL HIP REPLACEMENT. DURING THE PROCEDURE, THE S-ROM TRIANGLE MILLER HANDLE WAS BROKEN BY THE SURGEON DURING ATTEMPTS TO REMOVE A MILLER SHELL AND SHAFT PILOT THAT HAD BECOME "STUCK" INSIDE THE FEMORAL CANAL. IT IS UNCLEAR AS TO WHY THE MILLER SHELL AND SHAFT PILOT HAD BECOME STUCK, BUT A SMALL FEMORAL FRACTURE OCCURRED WHILST ATTEMPTING TO REMOVE THE "STUCK" INSTRUMENTS. THIS WAS RECTIFIED INTRA-OPERATIVELY BY THE SURGEON BY PLACING A CERCLAGE CABLE AROUND PROXIMAL PORTION OF THE FRACTURED FEMUR. FEMALE PATIENT ID (B)(6), INITIALS (B)(6). ACTION TAKEN: A SLAP HAMMER EXTRACTION DEVICE WAS USED TO REMOVE THE "STUCK" INSTRUMENTS, FOLLOWING WHICH THE CASE WAS COMPLETED. IT WAS ALSO NOTED DAMAGE TO ONE OF THE 20X15 S-ROM SHAFT PILOTS (REPORTED ABOVE) INCLUDED IN THIS KIT, BUT IT IS UNCLEAR AS TO WHEN OR HOW THAT DAMAGE OCCURRED?

    Patient problems: Hip Fracture

    MAUDE report 7065947 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices