Skip to content

S-ROM 53-6450Recalls & reported problems

Depuy (Ireland) · S-ROM STEMSLEEVE SEPARATOR · Class 1

All models of S-ROM (185)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy S-ROM separator is an orthopedic manual surgical instrument used with hip stems. Reports mostly describe the tip bending, chipping, cracking or breaking off, often noticed during kit inspection without patient impact. There are 25 adverse event reports that name its identifier, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 10603295172901), received between Sep 26, 2017 and Jul 28, 2025 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation11
  • Break7
  • Material Integrity Problem4
  • Material Twisted/Bent4
  • Naturally Worn2
  • Scratched Material2
  • Crack1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Consequences Or Impact To Patient8
  • Insufficient Information2
  • No Code Available1
  • No Information1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT SPD DID NOT ACCEPT THIS INSTRUMENT AS IT WAS TOO BADLY DAMAGED. THE TIP WAS BENT TO ONE SIDE. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22632302 ↗
  • MalfunctionFeb 3, 2025Crack; Material Integrity Problem; Crack; Material Integrity Problem

    IT WAS REPORTED THAT SOME INSTRUMENTS WERE CRACKED AND DAMAGED. NO IMPACT TO PATIENT; NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21295203 ↗
  • MalfunctionJan 29, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENT BENT DURING USE. SURGERY WAS NOT DELAYED DUE TO THE REPORTED EVENT. IT WAS UNKNOWN IF THEY PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21263392 ↗
  • MalfunctionDec 26, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A TOTAL HIP REPLACEMENT PROCEDURE ON (B)(6) 2024 , THE HEAD DISTRACTION TIP WAS BENT DURING USE FOR HEAD REMOVAL. THERE WAS NO ADDITIONAL SURGERY TIME WAS REQUIRED. BACKUP INSTRUMENT WAS ON HAND.

    Patient problems: Insufficient Information

    MAUDE report 21021033 ↗
  • MalfunctionOct 14, 2024Break; Break

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION ON OCTOBER 9, 2024, THE TIP OF THE INSTRUMENT WAS CHIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20439982 ↗
  • MalfunctionOct 3, 2024Break; Break

    IT WAS REPORTED THAT THE SROM STEM SLEEVE SEPARATOR IMPACTOR IS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 20363650 ↗
  • MalfunctionSep 13, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT UPON INSPECTION THE FOLLOWING PARTS WERE FOUND. S-ROM*STEM/SLEVE SEPARTOR - BENT TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20213146 ↗
  • MalfunctionJan 16, 2024Break; Break

    THIS INSTRUMENT IS COMPOSED OF 1 PART. THE INSTRUMENT HAS SNAPPED OFF. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18517200 ↗
  • MalfunctionJan 5, 2024Break; Material Deformation; Scratched Material; Break; Material Deformation; Scratched Material

    IT WAS REPORTED THAT THE END WAS BENT AND SNAPPED OFF. PATIENT STATUS/ OUTCOME / CONSEQUENCES: WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? UNKNOWN. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18453450 ↗
  • MalfunctionJun 27, 2023Break; Break

    IT WAS REPORTED THAT THE STEM/SLEEVE SEPARATOR WAS CHIPPED OFF DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17210948 ↗
  • MalfunctionMay 10, 2023Break; Material Deformation; Break; Material Deformation

    IT REPORTED THAT IT WAS NOTICED THAT THESE INSTRUMENT WERE DAMAGED OR BEAT UP. INSTRUMENTS ARE CHIPPED OR BENT. NOTHING LEFT IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16902408 ↗
  • MalfunctionSep 6, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TIP IS BENT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15364225 ↗
  • MalfunctionNov 15, 2021Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE TIP IS DULL AND FLATTENED OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12814019 ↗
  • MalfunctionAug 5, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT WHILE USING THE CHISEL TO BREAK THE TAPER, A DEFORMITY WAS CREATED IN THE FLAT TIP END.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12280030 ↗
  • MalfunctionJun 25, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED AS "DAMAGED TRYING TO GET STEM OUT DURING A REVISION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12066000 ↗
  • MalfunctionJun 25, 2021Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED AS "DAMAGED TRYING TO GET STEM OUT DURING A REVISION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12065999 ↗
  • MalfunctionMar 22, 2021Material Integrity Problem; Material Integrity Problem

    STOM STEM SEPARATOR DAMAGED. NO REPORTED SURGICAL DELAYS OR PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11541304 ↗
  • MalfunctionOct 13, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE LPS EXTRACTOR ADAPTER WAS STUCK ON LPS INSERTER HANDLE. SROM OSTEOTOME IS WARPED.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10671044 ↗
  • MalfunctionApr 3, 2020Material Twisted/Bent; Scratched Material; Illegible Information; Material Twisted/Bent; Scratched Material; Illegible Information

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SCRATCHED UP OR DAMAGED AND SIZES CAN¿T BE READ OR IDENTIFIED EASILY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9919388 ↗
  • MalfunctionOct 1, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TRIAL 15MM STEM AND THE 36+8 NECK SEGMENT ARE BOTH STRIPPED AND WOULD NOT ENGAGE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7923841 ↗
  • MalfunctionAug 30, 2018Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE STEM/SLEEVE SEPARATOR BENT DURING USE. NO ADDITIONAL INFORMATION. PLEASE REPLACE TO OUR (B)(6) HOLD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7832503 ↗
  • MalfunctionAug 23, 2018Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE INSTRUMENTS WERE WORN OUT AND BENT AND NEEDS TO BE REPLACED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7809266 ↗
  • MalfunctionAug 20, 2018Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE IMPACTION END IS DULL AND FLAT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7798891 ↗
  • MalfunctionAug 20, 2018Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT IMPACTION IN IS BEND AND DULL. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7796850 ↗
  • MalfunctionSep 26, 2017Break; Break

    COMPLAINT DESCRIPTION: IT WAS REPORTED THAT THE SET OF SROM HIP INSTRUMENTS WERE PUT TOGETHER AND NOTICED TIP OF SLEEVE EXTRACTOR WAS SEVERELY DAMAGED.

    Patient problems: No Consequences Or Impact To Patient; No Information

    MAUDE report 6894673 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page