Skip to content

SIGMA LCS 9505-01-203Recalls & reported problems

Depuy (Ireland) · SIGMA LCS HIGH PERFORMANCE IM TIBIA JIG · Class 1

All models of SIGMA LCS (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy SIGMA LCS high performance tibial jig is a surgical alignment guide used to set the angle of the tibia bone during knee replacement. The 33 reports mostly describe metal parts breaking, bending, wearing out or jamming, with settings that drift or would not hold and a dial that would not turn. No deaths and no injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 10603295225119), received between Aug 18, 2017 and Mar 24, 2025 — classified as 32 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Appropriate Device Problem Term/Code Not Available4
  • Material Deformation4
  • Device-Device Incompatibility3
  • Mechanical Jam3
  • Naturally Worn3
  • Physical Resistance/Sticking3
  • Incorrect Measurement2

Patient problems

  • No Consequences Or Impact To Patient14
  • No Clinical Signs, Symptoms or Conditions11
  • Not Applicable3
  • Insufficient Information2
  • No Information2

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE THREADS ON THE DEVICE WERE STRIPPED.

    Patient problems: Insufficient Information

    MAUDE report 21678452 ↗
  • MalfunctionJan 31, 2025Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT IM CUTTING GUIDE DID NOT ALLOW FOR SLOPE ADJUSTMENTS. ANOTHER WAS AVAILABLE AND THE CASE PROCEEDED WITHOUT DELAY. THERE WAS NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21288521 ↗
  • MalfunctionJan 16, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE BLUE DIAL ON THE JIG WILL NOT ROTATE AND NEEDS REPLACING.

    Patient problems: Insufficient Information

    MAUDE report 21167142 ↗
  • MalfunctionJul 21, 2023Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE IM TIBIAL JIG POSTERIOR SLOPE ADJUSTMENT HAS TOO MUCH ¿PLAY¿ WHEN ATTACHED TO THE OUTRIGGER, IMPEDING THE ABILITY TO GET A PREDICTABLE AND PRECISE SETTING FOR TIBIAL POSTERIOR SLOPE. NO SURGICAL DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17370564 ↗
  • MalfunctionJan 11, 2023Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT SIGMA IM TIBIAL JIG WILL NOT HOLD TIBIAL CUTTING BLOCK. POSSIBLY THE TWO POST THAT HOLD THE BLOCK ARE BENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16138208 ↗
  • MalfunctionNov 7, 2022Break; Break

    IT WAS REPORTED THAT THE IMPACTOR BROKE DURING FINAL INSERTION. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15742802 ↗
  • MalfunctionAug 22, 2022Device-Device Incompatibility; Material Deformation; Material Integrity Problem; Device-Device Incompatibility; Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT THE L SHAPED PIECE ON THE SIGMA IM TIBIA GUIDE WAS DAMAGED AND THE ROD COULD NOT FIT THROUGH THE HOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15267207 ↗
  • MalfunctionAug 19, 2022Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE THREAD ON THE SCREW IS STRIPPED. OCCURRED BEFORE CASE STARTED. INSTRUMENT WAS USED IN CASE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15258261 ↗
  • MalfunctionOct 28, 2021Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS NOTED THAT THE SCREW ON THE IM TIBIA JIG WAS STUCK AND WOULD NOT SCREW OR UNSCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12711863 ↗
  • MalfunctionSep 29, 2021Break; Break

    IT WAS REPORTED THAT THE BLUE KNOBS ON IM TIBIA JIG BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12546994 ↗
  • MalfunctionSep 29, 2021Break; Break

    IT WAS REPORTED THAT THE BLUE KNOBS ON IM TIBIA JIG BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12546992 ↗
  • MalfunctionAug 4, 2021Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE HEIGHT ADJUSTMENT KNOB ON THE PFC IM TIBIAL RESECTION GUIDE WAS NOT ENGAGING. IT APPEARS THE THREADS ON THE SCREW ARE NOT ENGAGING WITH THE GROOVES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12274087 ↗
  • MalfunctionDec 17, 2020Material Integrity Problem; Material Integrity Problem

    BLUE THUMB SCREW ON INTRAMEDULLARY JIG IS FAULTY. USED EXTRAMEDULLARY ALIGNMENT GUIDE AS THE ALTERNATIVE. NO SURGICAL DELAY. PROCEDURE COMPLETED WITHOUT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11027428 ↗
  • MalfunctionMay 28, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE KNOB THAT TIGHTENS THE SIGMA IM TIBIAL JIG WAS STRIPPED. NO FURTHER INFORMATION WAS AVAILABLE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10097833 ↗
  • MalfunctionDec 5, 2019Material Twisted/Bent; Naturally Worn; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE BOLT WAS STRIPPED ON DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9418238 ↗
  • MalfunctionSep 3, 2019Break; Break

    IT WAS REPORTED THAT THE BLUE SCREW KNOB WAS BROKEN OF THE IM TIBIA JIG DURING SURGERY. IT WAS RETRIEVED AND WAS DISCARDED BY OPERATION ROOM. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8957803 ↗
  • MalfunctionJul 2, 2019Break; Break

    IT WAS REPORTED THAT THE BLUE SCREW WAS BROKE OFF IM TIBIA JIG. BOTH PIECES RETRIEVED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8754714 ↗
  • MalfunctionJun 24, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    CLIENT REQUESTS THE CHANGE OF THE EXTRA MEDULLARY GUIDE OF TIBIAL CUT CODE (B)(4), SINCE IT IS LOCKED IN ITS USE AND THAT MODIFIES THE TIBIAL CUT. ALREADY NEED REPLACEMENT FOR SMOOTH PINS, WHICH WEAR AND TWIST DURING SURGERIES. HP IM TIBIA JIG WAS STUCK.

    Patient problems: Not Applicable

    MAUDE report 8726698 ↗
  • MalfunctionJun 24, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    CLIENT REQUESTS THE CHANGE OF THE EXTRA MEDULLARY GUIDE OF TIBIAL CUT CODE (B)(4), SINCE IT IS LOCKED IN ITS USE AND THAT MODIFIES THE TIBIAL CUT. ALREADY NEED REPLACEMENT FOR SMOOTH PINS, WHICH WEAR AND TWIST DURING SURGERIES. HP IM TIBIA JIG WAS STUCK.

    Patient problems: Not Applicable

    MAUDE report 8726696 ↗
  • MalfunctionMay 22, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    LOCKING SCREWS ON THE HP IM JIG ARE STICKING.

    Patient problems: Not Applicable

    MAUDE report 8631770 ↗
  • MalfunctionMar 28, 2019Use of Device Problem

    IT WAS REPORTED THAT THE TIBIAL JIG HAS TO MUCH TO PLAY IN IT WHEN SETTING THE SLOPE. WHEN COMPARED TO OTHER JIGS THERE WAS A SIGNIFICANT DIFFERENCE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8460642 ↗
  • MalfunctionMar 21, 2019Break; Break

    IT WAS REPORTED THAT THE BLUE SCREW KNOB OF THE IM TIBIA JIG WAS BROKEN OFF DURING SURGERY. IT WAS RETRIEVED. DISCARDED BY HOSPITAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8441145 ↗
  • MalfunctionOct 11, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IM TIBIAL JIG BLUE KNOB WOULD NOT TIGHTEN/LOOSEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7957546 ↗
  • MalfunctionOct 2, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    BLUE KNOB ON THE FRONT WASN'T LOCKING INTO PLACE WHEN ADJUSTING THE LEVEL OF RESECTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7927329 ↗
  • MalfunctionSep 24, 2018Incorrect Measurement; Appropriate Device Problem Term/Code Not Available; Incorrect Measurement; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE HP, IM TIBIAL JIG IS NOT WORKING , YOU CAN NOT ADJUST THE DEGREE OF SLOPE. THE PART #/LOT # ARE WORN OFF CAN NOT READ WHAT THEY ARE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7902925 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page