Skip to content

SIGMA LCS 9505-02-016Recalls & reported problems

Depuy Orthopaedics, Inc. · SIGMA HIGH PERFORMANCE LCS HIGH PERFORMANCE MBT PUNCH CEMENTED SZ 1-1.5 · Class 1

All models of SIGMA LCS (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jan 8, 2016Terminated

    Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space. Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.

    Z-0787-2016 · Root cause: Component design/selection · Recall details · FDA record ↗

Other PUNCH, SURGICAL devices

All PUNCH, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page