SIGMA LCS recallZ-0787-2016 · DePuy Orthopaedics, Inc. · initiated Jan 8, 2016
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space. Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.
FDA root cause category: Component design/selection
What the company is doing
On 1/8/2016, URGENT INFORMATION RECALL NOTICE for Specific Lots of SIGMA High Performance (HP) MBT Non-Keel Punch Knee Instrument notifications were sent to the affected distributors and Medical Professionals with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc.s Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. 5 p.m. EDT). For questions about the device recall information provided, please contact Kim Earle, Senior Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. 5 p.m. EDT).
Recall details
- Recall number
- Z-0787-2016
- Status
- Terminated
- Initiated by the company
- Jan 8, 2016
- Posted by the FDA
- Feb 9, 2016
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 351
- Product code
- LRY
- Distribution
- Worldwide Distribution-US Nationwide, Armenia, Australia, Austria, Belgium, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, P R China, Portugal, Poland, Russia, Singapore, Slovenia, South Africa, Spain, Sweden, South Korea, Switzerland, Thailand, Turkey, UAE, UK, and Brazil.
Product description
SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction.
Code information (models, lots, serial numbers)
Cat. No. GTIN Lot Description 950502016 10603295227366 ABB92402 SIGMA HP MBT Cemented Punch size 1-1.5 950502017 10603295227373 ABB84491 SIGMA HP MBT Cemented Punch size SZ 2-3 950502018 10603295227380 ABB92358 SIGMA HP MBT Cemented Punch size SZ 4-7 950502020 10603295227403 ABC33125 SIGMA HP MBT Noncemented Punch size 2-3 950502020 10603295227403 ABB81733 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502020 10603295227403 ABB81732 SIGMA HP MBT Noncemented Punch size SZ 2-3 950502021 10603295227410 ABB39952 SIGMA HP MBT Noncemented Punch size SZ 4-7 950502021 10603295227410 ABC33123 SIGMA HP MBT Noncemented Punch size SZ 4-7
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- SIGMA LCS 9505-02-018Depuy Orthopaedics, Inc. · PUNCH, SURGICAL1 recall10 reports
- SIGMA LCS 9505-02-017Depuy Orthopaedics, Inc. · PUNCH, SURGICAL1 recall1 report
- SIGMA LCS 9505-02-016Depuy Orthopaedics, Inc. · PUNCH, SURGICAL1 recall0 reports
- SIGMA LCS 9505-02-020Depuy Orthopaedics, Inc. · PUNCH, SURGICAL1 recall0 reports
- SIGMA LCS 9505-02-021Depuy Orthopaedics, Inc. · PUNCH, SURGICAL1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page