SIGMA LCS 9505-02-070Recalls & reported problems
Depuy (Ireland) · SIGMA HIGH PERFORMANCE LCS HIGH PERFORMANCE PIN IMPACTOR/EXTRACTOR · Class 1
All models of SIGMA LCS (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10603295227861), received between May 7, 2018 and Nov 30, 2023 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Device-Device Incompatibility5
- Loose or Intermittent Connection3
- Naturally Worn2
- Connection Problem1
- Detachment of Device or Device Component1
- Difficult to Open or Close1
- Fitting Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient8
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 30, 2023Detachment of Device or Device Component; Difficult to Open or Close; Detachment of Device or Device Component; Difficult to Open or Close
ADDITIONAL INFORMATION RECEIVED STATES THAT IT IS UNSURE EXACTLY WHICH PART OF THE INSTRUMENT HAS FAILED BUT AS PER BELOW THERE IS A SMALL SCREW MISSING FROM THE HANDLE SO DOES NOT FUNCTION AS INTENDED AS DOES NOT CLOSE AROUND PINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18237146 ↗MalfunctionJul 14, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN IMPACTOR IS BROKEN. NO LONGER HOLDS ON TO PIN SECURELY. NO DELAY TO SURGERY OR HARM TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15014552 ↗MalfunctionFeb 17, 2022Device-Device Incompatibility; Device-Device Incompatibility
PIN EXTRACTOR NO LONGER HOLDS THE PINS FOR EXTRACTING. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13553208 ↗MalfunctionFeb 1, 2022Naturally Worn; Naturally Worn
INSTRUMENTS ARE DULL. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13408388 ↗MalfunctionFeb 1, 2022Naturally Worn; Naturally Worn
INSTRUMENTS ARE DULL. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13408208 ↗MalfunctionOct 4, 2021Break; Break
1. WAS IT BROKE INTO 2 OR MORE PIECES? NO. 2. WHICH PART OF THE HP PIN IMPACTOR/EXTRACTOR WAS BROKEN? THE PORTION WHICH IS A METAL STRIP THAT ACTS AS A SPRING WHEN GRASPING A PIN & RELEASING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12571869 ↗MalfunctionSep 24, 2021Break; Break
THE SPRING HANDLE NO LONGER WORKS, ALTERNATIVE USED. NO PATIENT IMPACT OR DELAY TO SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12528266 ↗MalfunctionSep 10, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE TIBIA PUNCH HANDLE WILL NOT RETAIN PUNCH. PIN PULLER INNER RESISTANCE WILL NOT STAY TOGETHER. ALL INSTRUMENT PIECES INTACT. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12452418 ↗MalfunctionJun 24, 2021Break; Break
IT WAS REPORTED AS "RESISTANCE PIECE OF PIN PULLER BROKE OFF. ALL PIECES WERE RETRIEVED." NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12060992 ↗MalfunctionApr 29, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE DEPUY SYNTHES PIN EXTRACTOR IN OUR INSTRUMENT SET WOULD NOT REMOVE A PIN. THEY OPENED ANOTHER SET AND WITHOUT ANY DELAY CONTINUED THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11748044 ↗MalfunctionApr 29, 2021Break; Break
IT WAS REPORTED THAT THE PIN PULLER SPRING MECHANISM BROKE. THE INSTRUMENT BROKE IN 2 PIECES AND ALL PIECES WERE RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11747957 ↗MalfunctionApr 29, 2019Device-Device Incompatibility; Device-Device Incompatibility
THE INSTRUMENT IS NO LONGER ABLE TO EXTRACT THE PIN. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8562224 ↗MalfunctionApr 8, 2019Break; Fitting Problem; Break; Fitting Problem
THE INSTRUMENT IS NO LONGER POSSIBLE TO EXTRACT THE PIN. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8490917 ↗MalfunctionFeb 19, 2019Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT WHILE THE SURGEON IS IMPACTING THE TIBIA BASE PINS IN WITH THE HP PIN IMPACTOR/EXTRACTOR, THE PINS GET STUCK IN THE IMPACTOR AND CANNOT BE REMOVED. THE SIGMA SIZE 4 PS BOX TRIAL SCREW IS STRIPPED AND DOESN¿T ALLOW THE SCREW DRIVER TO TIGHTEN SCREW DOWN ON FEMUR TRIAL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8349991 ↗MalfunctionAug 22, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ALL THREE PIN PULLERS OPENED ON A CASE AND AREN'T TIGHT ENOUGH TO GRAB THE PIN. NO SURGICAL DELAY AND ALL PIECES RETRIEVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806778 ↗MalfunctionAug 22, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ALL THREE PIN PULLERS OPENED ON A CASE AND AREN'T TIGHT ENOUGH TO GRAB THE PIN. NO SURGICAL DELAY AND ALL PIECES RETRIEVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806730 ↗MalfunctionAug 22, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ALL THREE PIN PULLERS OPENED ON A CASE AND AREN'T TIGHT ENOUGH TO GRAB THE PIN. NO SURGICAL DELAY AND ALL PIECES RETRIEVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806705 ↗MalfunctionJun 14, 2018Break; Break
IT WAS REPORTED THAT THE INSTRUMENT NO LONGER WORKS DUE TO SPRING BEING BROKEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7602727 ↗MalfunctionMay 7, 2018Connection Problem; Connection Problem
THE CLIP IN THE MID OF THE INSTRUMENT DOES NOT HOLD ANY LONGER. THIS WAS SEEN IN CSSD. NO SURGERY DELAY, NO ADVERSE CONSEQUENCES FOR ANYBODY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7490242 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page