Skip to content

NA 9626-07-000Recalls & reported problems

Depuy International Ltd · CEMENTED CUP PUSHER · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295243311), received between Mar 12, 2020 and May 2, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Crack” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack6
  • Contamination /Decontamination Problem1
  • Device-Device Incompatibility1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 2, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED: I HAVE NOT BEEN MADE AWARE OF ANY ADVERSE EVENT TO A PATIENT ARISING BECAUSE OF THIS MATTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19229378 ↗
  • MalfunctionDec 20, 2022Crack; Crack

    I¿VE HAD AN EMAIL FROM BMI LINCOLN REPORTING THAT ONE OF THEIR CEMENTED CUP PUSHERS HAS A CRACK IN ITS HANDLE AFTER REPEATED REPROCESSING SO THEIR TSSU WILL NOT LONGER PROCESS. THERE WAS NO IMPACT TO SURGERY OR THE PATIENT AND THE ITEM HAS BEEN DISPOSED OF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16013587 ↗
  • MalfunctionApr 13, 2022Crack; Crack

    IT WAS REPORTED THAT CRACK IN THE HANDLE, DISCOVERED DURING INSTRUMENTS CHECK BY SALES REPRESENTATIVE. NO DELAY TO SURGERY BECAUSE OF THE ABOVE. I NOTED IT, NOT THE ACCOUNT, SO NOT REPORTED BY THEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14102314 ↗
  • MalfunctionNov 25, 2020Crack; Crack

    THE TWO INSTRUMENTS ARRIVED THE HOSPITAL WITH CRACK IN THE PLASTIC WHERE / WHEN DID THE EVENT OCCUR? OTHER ARRIVED WITH CRACK WHAT ACTION WAS TAKEN/REQUIRED TO MANAGE THE PROBLEM DURING THE PROCEDURE HAS NOT BEEN IN USE WAS A PATIENT INVOLVED: NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10900500 ↗
  • MalfunctionNov 25, 2020Crack; Crack

    THE TWO INSTRUMENTS ARRIVED THE HOSPITAL WITH CRACK IN THE PLASTIC. WHERE / WHEN DID THE EVENT OCCUR? OTHER ARRIVED WITH CRACK. WHAT ACTION WAS TAKEN/REQUIRED TO MANAGE THE PROBLEM DURING THE PROCEDURE HAS NOT BEEN IN USE. WAS A PATIENT INVOLVED: NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10900386 ↗
  • MalfunctionNov 25, 2020Crack; Contamination /Decontamination Problem; Crack; Contamination /Decontamination Problem

    HAS PRODUCT BEEN POTENTIALLY CONTAMINATED OR IS CONTAMINATED WITH BODILY FLUIDS? YES, CRACK IN PLASTIC. THIS COMPLAINT IS ABOUT AN INSTRUMENT, WHICH ONLY HAS BEEN USED IN THE WASHING MACHIN, AND A CRACK APPEARED IN THE PLASTIC. WHERE / WHEN DID THE EVENT OCCUR? POST - OP. WAS A PATIENT INVOLVED? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10899700 ↗
  • MalfunctionMar 12, 2020Crack; Crack

    A COMPLAINT ON AN INSTRUMENT WHICH HAS NOT BEEN USED. CRACK IN THE PLASTIC DURING WASHING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9824112 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page