TRUMATCH 42-0906Recalls & reported problems
Depuy (Ireland) · TruMatch PERSONALIZED SOLUTIONS FEMUR RESECTION GUIDE TIBIA PIN GUIDE SET - CT KNEE FOR USE WITH: LEFT HYBRID · Class 2
All models of TRUMATCH (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295384816), received between Jul 17, 2020 and Apr 17, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement3
- Device-Device Incompatibility2
- No Apparent Adverse Event1
- Nonstandard Device1
Patient problems
- Insufficient Information3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Device-Device Incompatibility; No Apparent Adverse Event
IT WAS REPORTED THAT THE SCRUB TECHS WERE HAVING ISSUES ATTACHING THE 2-PRONGED ATTACHMENT PIECE TO THE TIBIAL PIN BLOCKS BECAUSE THE HOLES SEEM TO BE MANUFACTURED SMALLER THAN PREVIOUSLY. IT HAS ALSO CREATED ISSUES FOR THE SURGEONS NOT BEING ABLE TO VISUALIZE THEIR ALIGNMENT AND SLOPE BECAUSE THE METAL PIECE WON¿T SIT FLUSH. A COUPLE OF BLOCKS HAVE BROKEN AT THAT JUNCTURE, AS THE SCRUB TECHS USE METAL INSTRUMENTS ATTEMPTING TO TAP IT IN FLUSH.
Patient problems: Insufficient Information
MAUDE report 24916573 ↗MalfunctionApr 10, 2026Nonstandard Device; Device-Device Incompatibility
IT WAS REPORTED THAT THE SCRUB TECHS WERE HAVING ISSUES ATTACHING THE 2-PRONGED ATTACHMENT PIECE TO THE TIBIAL PIN BLOCKS BECAUSE THE HOLES SEEM TO BE MANUFACTURED SMALLER THAN PREVIOUSLY. IT HAS ALSO CREATED ISSUES FOR THE SURGEONS NOT BEING ABLE TO VISUALIZE THEIR ALIGNMENT AND SLOPE BECAUSE THE METAL PIECE WON¿T SIT FLUSH. A COUPLE OF BLOCKS HAVE BROKEN AT THAT JUNCTURE, AS THE SCRUB TECHS USE METAL INSTRUMENTS ATTEMPTING TO TAP IT IN FLUSH.
Patient problems: Insufficient Information
MAUDE report 24840930 ↗MalfunctionFeb 2, 2021Incorrect Measurement; Incorrect Measurement
DURING BILATERAL TKA, THERE WAS A MAJOR ISSUE WITH THE EXTERNAL ROTATION ON BOTH FEMORAL BLOCKS. THE BLOCKS WERE EXTERNALLY ROTATED. AFTER NOTICING THERE WAS TOO MUCH EXTERNAL ROTATION, SURGEON HAD TO USE THE REGULAR SIZING DEVICE TO DIAL IT BACK DOWN TO NORMAL, DELAYING EACH KNEE BY ABOUT 10 MINUTES."
Patient problems: Insufficient Information
MAUDE report 11265643 ↗MalfunctionJul 17, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED AS " EVENT DESCRIPTIONS: TMK23137, KK, NOT ENOUGH EXTERNAL ROTATION IN SURGERY TMK23228, MT, TOO MUCH EXTERNAL ROTATION IN SURGERY TMK23258, JG, LEFT KNEE, FOLLOWING UP ON ROTATION ISSUE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10290243 ↗MalfunctionJul 17, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED AS " EVENT DESCRIPTIONS: TMK23137, KK, NOT ENOUGH EXTERNAL ROTATION IN SURGERY. TMK23228, MT, TOO MUCH EXTERNAL ROTATION IN SURGERY. TMK23258, JG, LEFT KNEE, FOLLOWING UP ON ROTATION ISSUE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10290230 ↗
Other Knee arthroplasty implantation system devices
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR111111102Conformis, Inc. · Knee Arthroplasty Implantation System390 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR111SPEC02Conformis, Inc. · Knee Arthroplasty Implantation System11 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR121SPEC02Conformis, Inc. · Knee Arthroplasty Implantation System11 reports
- iTotal® Identity Cruciate Retaining Knee Replacement System TCR301111301Conformis, Inc. · Knee Arthroplasty Implantation System3 reports
- iTotal Identity CR RCR301111301Conformis, Inc. · Knee Arthroplasty Implantation System1 report
- ITOTAL IDENTITY IMPACTOR ADAPTER 1080-213Conformis, Inc. · Knee Arthroplasty Implantation System1 report
- VISIONAIRE V0200128Smith & Nephew, Inc. · Knee arthroplasty implantation system1 report
- TRUMATCH 420902Depuy Orthopaedics, Inc. · Knee arthroplasty implantation system1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page