Skip to content

ACTIS 2010-07-120Recalls & reported problems

Depuy (Ireland) · ACTIS CURVED INSERTER SHAFT · Class 1

All models of ACTIS (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Actis curved inserter shaft is a surgical instrument used during hip replacement surgery. The 49 reports mostly describe the instrument not working with other parts of the DePuy system, sticking or jamming so pieces would not come apart after cleaning, and bending, chipping or breaking during surgery, sometimes causing a delay of a few minutes. No deaths or injuries are described. There are no recalls naming this device, and reports have fallen from 9 to 4 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 10603295394006), received between Sep 25, 2018 and Sep 30, 2026 — classified as 50 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (41 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility41
  • Break5
  • Material Twisted/Bent3
  • Material Deformation2
  • Mechanical Jam2
  • Physical Resistance/Sticking2
  • Appropriate Device Problem Term/Code Not Available1
  • Corroded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • No Consequences Or Impact To Patient7
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Device-Device Incompatibility

    It was reported that the emphasys impactor threaded part no longer threads.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946330 ↗
  • MalfunctionSep 15, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781191 ↗
  • MalfunctionAug 18, 2026Device-Device Incompatibility

    It was reported that after the instruments went through decontamination process the 2 piece Actis stem inserter and handle would not come apart.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50299073 ↗
  • MalfunctionMay 8, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT UNIVERSAL HANDLE AND CURVED STEM INSERTER WERE STUCK TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25123289 ↗
  • MalfunctionMar 31, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DOES NOT FIT IN THE UNIVERSAL INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742066 ↗
  • MalfunctionOct 20, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP REPLACEMENT. THE TIP CHIPPED OFF OF THE INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23340519 ↗
  • MalfunctionSep 10, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE PATIENT HAS A LEFT TOTAL HIP. THE 2 PIECE ACTIS STEM INSERTER WAS STUCK TOGETHER AND IT WOULD NOT SEPARATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23014896 ↗
  • MalfunctionAug 12, 2025Break; Break

    IT WAS REPORTED THAT THE IMPACTOR FITTING BROKEN DURING SURGERY. BROKE WHEN ONE PIECE FIT INTO ANOTHER. BOTH PIECES BROKEN INTO 2 PIECES. NO FRAGMENTS WERE GENERATED. A NEW INSTRUMENT WAS OPENED TO USE. THERE WAS A 5 MINUTES OF SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22786542 ↗
  • MalfunctionAug 11, 2025Corroded; Material Deformation; Corroded; Material Deformation

    IT WAS REPORTED THAT RETRACTORS BENTS. THERE WAS METAL PITTING ON INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22769992 ↗
  • MalfunctionJul 30, 2025Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE C-CLAMP ON THE NECK BROKE OFF, THE TIP OF THE STEM INSERTER HAS BEEN DISTORTED AND WILL NOT ENGAGE IN STEM. THIS EVENT DID NOT HAPPEN IN SURGERY. ALL AVAILABLE INFORMATION HAS BEEN FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22664129 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED THAT THERE WERE BROKEN INSTRUMENTS DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22338510 ↗
  • MalfunctionMar 17, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE CURVED INSERTER SEIZED WITH INSERTER HANDLE, WILL NOT DISCONNECT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21621123 ↗
  • MalfunctionFeb 11, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS IMPACTOR AND UNIVERSAL HANDLE WAS FROZEN TOGETHER AND THEY WERE UNABLE TO SEPARATE IN POST-OP TO WASH AND SANITIZE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356244 ↗
  • MalfunctionNov 14, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSERTER AND INSERTER HANDLE ARE COLD WELDED TOGETHER, AND COULD NOT COME APART.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20688185 ↗
  • MalfunctionSep 5, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE.

    Patient problems: Insufficient Information

    MAUDE report 20143981 ↗
  • MalfunctionJun 4, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE UNIVERSAL HANDLE AND CURVED INSERTER BECAME COLD WELDED AND COULD NOT BE REMOVED FROM EACH OTHER. THIS DID NOT DELAY IN THE SURGERY. NO OTHER INFORMATION WAS AVAILABLE AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 19458708 ↗
  • MalfunctionJan 29, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS STEM INSERTER IS STUCK IN THE UNIVERSAL STEM INSERT HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18601154 ↗
  • MalfunctionJan 29, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT STEM INSERTER SHAFT AND HANDLE BECAME FUSED AFTER IMPACTION. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO, WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN, WERE FRAGMENTS GENERATED? UNKNOWN, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES NO, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18600239 ↗
  • MalfunctionJan 4, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS STEM INSERTER AND HANDLE GOT STUCK AND COULD NOT BE PULLED APART AFTER THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444061 ↗
  • MalfunctionDec 19, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT IMPACTOR HANDLE WAS BROKEN FROM BACK AT STRIKE PLATE. ACTIS CURVED INSERTER SHAFT & UNIVERSAL STEM INSERT HANDLE STUCK TOGETHER, COULD NOT SEPARATE. AND WHITE HANDLE WAS CRACKED.

    Patient problems: Insufficient Information

    MAUDE report 18354647 ↗
  • MalfunctionDec 4, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE TIPS ON THESE INSERTERS ARE WORN BEYOND USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18261080 ↗
  • MalfunctionAug 25, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS OFFSET INSERTER IS HOLDING ON TO PROSTHESIS AND NOT RELEASING. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17626123 ↗
  • MalfunctionApr 24, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE CALCAR PLANAR DISC WOULD NOT THREAD ONTO THE CALCAR PLANAR BASE. IT LOOKED LIKE THE THREADS ON THE BASE WERE WORN OR MIGHT HAVE BEEN CROSS THREADED. THE STEM INSERTER AND HANDLE WERE ALSO STUCK AFTER THE CASE AND COULD NOT BE SEPARATED BY TECH OR REP. IT WAS UNKNOWN ABOUT THE SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16795303 ↗
  • MalfunctionApr 11, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THE STEM INSERTER HAND AND ANTERIOR TIP WERE STUCK TOGETHER. DISCOVERY WAS MADE AFTER SURGERY DURING DECONTAMINATION. NO ADVERSE AFFECTS ON THAT PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16715702 ↗
  • MalfunctionDec 13, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TWO PIECE INSERTER WILL NOT SEPARATE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15968684 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page