Skip to content

ATTUNE 2545-00-153Recalls & reported problems

Depuy (Ireland) · ATTUNE A-P CHAMFER BLOCK SIZE 3 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10603295395348), received between Sep 25, 2019 and Jun 27, 2025 — classified as 19 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation16
  • Naturally Worn4
  • Corroded3
  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient9

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE BROKEN. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22350041 ↗
  • MalfunctionMay 27, 2025Material Deformation; Material Deformation

    METAL ON SIZE 3 CUTTING BLOCK HAS BEEN DAMAGED DUE TO OVERUSE. STERILE PROCESSING DEPARTMENT WILL NOT PROCESS DUE TO DAMAGE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22090913 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19095795 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORRMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19095793 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19095792 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19095791 ↗
  • MalfunctionApr 12, 2024Material Deformation; Material Deformation

    AT A GENERAL CHECK OF THE ATTUNE AP CUTTING BLOCKS THE CENTRAL STERILIZATION DEPARTMENT DETECTED SUBSTANTIAL SHARP EDGES AT THE CORNERS OF THE CUTTING SLOTS. THERE WERE NO IMPLICATIONS DURING ANY SURGERY! NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19095779 ↗
  • MalfunctionNov 13, 2023Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE A-P CHAMFER BLOCK SIZE 3 HAS A SHARP EDGE AND NEEDS REPLACING NO EVENT. WEAR AND TEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18118592 ↗
  • MalfunctionAug 12, 2022Break; Break

    IT WAS REPORTED THAT THE SIZE 3 4-1 CUTTING BLOCK METAL GENERATED FRAGMENTS FROM SAW BLADE DURING PROCESSING THE INSTRUMENT BY SPD. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15216678 ↗
  • MalfunctionSep 29, 2021Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE CUTTING BLOCKS HAVE BEEN WORN OVER TIME BY THE SAW BLADE, MAKING THEM DIFFICULT TO USE FOR THE SURGEON IN THE CASE. HE HAS REQUESTED NEW BLOCKS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12549994 ↗
  • MalfunctionMar 24, 2020Corroded; Material Deformation; Corroded; Material Deformation

    THE PROBLEM WAS THAT THE METAL QUALITY WAS INFERIOR AND THIS LEADS TO SHARP METAL EDGES IN THE CUTTING SLOTS AND EVEN RUST/CORROSION FORMATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9871183 ↗
  • MalfunctionMar 24, 2020Corroded; Material Deformation; Corroded; Material Deformation

    THE PROBLEM WAS THAT THE METAL QUALITY WAS INFERIOR AND THIS LEADS TO SHARP METAL EDGES IN THE CUTTING SLOTS AND EVEN RUST/CORROSION FORMATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9871182 ↗
  • MalfunctionMar 24, 2020Corroded; Material Deformation; Corroded; Material Deformation

    THE PROBLEM WAS THAT THE METAL QUALITY WAS INFERIOR AND THIS LEADS TO SHARP METAL EDGES IN THE CUTTING SLOTS AND EVEN RUST/CORROSION FORMATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9871176 ↗
  • MalfunctionOct 25, 2019Material Deformation; Material Deformation

    BURRS STICKING UP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9238390 ↗
  • MalfunctionOct 24, 2019Material Deformation; Material Deformation

    BURRS STICKING UP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9233072 ↗
  • MalfunctionOct 24, 2019Material Deformation; Material Deformation

    BURRS STICKING UP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9233070 ↗
  • MalfunctionOct 24, 2019Material Deformation; Material Deformation

    BURRS STICKING UP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9233068 ↗
  • MalfunctionSep 25, 2019Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE 4 IN 1 BLOCKS WERE FRAYED ON EDGES WERE TEARING GLOVES. NEED TO BE DULLED DOWN OR REPLACED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119223 ↗
  • MalfunctionSep 25, 2019Material Deformation; Naturally Worn; Material Deformation; Naturally Worn

    IT WAS REPORTED THAT THE 4 IN 1 BLOCKS WERE FRAYED ON EDGES WERE TEARING GLOVES. NEED TO BE DULLED DOWN OR REPLACED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119220 ↗

Other INSTRUMENT, CUTTING, ORTHOPEDIC devices

All INSTRUMENT, CUTTING, ORTHOPEDIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page