Skip to content

ATTUNE 2545-00-163Recalls & reported problems

Depuy (Ireland) · ATTUNE TIBIAL DRILL STOP SIZE 3-5 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10603295422525), received between Apr 25, 2018 and Aug 5, 2026 — classified as 22 malfunctions.

The most frequently reported specific device problem is “Break” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break13
  • Device-Device Incompatibility9
  • Detachment of Device or Device Component1
  • Illegible Information1
  • Naturally Worn1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient4
  • Insufficient Information2

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50139523 ↗
  • MalfunctionJan 7, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRILL STOP WOULD NOT CLICK ONTO THE BEARINGS. THE OTHER DRILL STOPPED WORKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23990330 ↗
  • MalfunctionNov 24, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT IT WOULD NOT SNAP ON THE DRILL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630354 ↗
  • MalfunctionJul 28, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT RED RING SLIPS OFF TOO EASY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22636579 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT WAS BROKEN AFTER USE DURING A TOTAL KNEE REPLACEMENT. ALL PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 21925148 ↗
  • MalfunctionApr 2, 2025Break; Break

    IT WAS REPORTED THAT A SIZE 3-5 DRILL STOP FOR THE TIBIAL DRILL WAS FOUND BROKEN WHEN THE TRAY WAS OPENED FOR A PROCEDURE. THE PIECE THAT HAS BEEN BROKEN OFF THE DRILL WAS NOT IN THE TRAY. THE SURGEON USED THE TIBIAL DRILL WITHOUT THE DRILL STOP, USING THE LINES MARKED ON THE DRILL TO GUIDE DEPTH. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21752469 ↗
  • MalfunctionMar 4, 2025Break; Break

    IT WAS REPORTED THAT THE DEVICES WERE BROKEN. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21515122 ↗
  • MalfunctionMar 4, 2025Break; Break

    IT WAS REPORTED THAT THE DEVICES WERE BROKEN. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514963 ↗
  • MalfunctionJul 9, 2024Break; Break

    IT WAS REPORTED THAT THE PIECES HAVE BROKEN. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19700401 ↗
  • MalfunctionJan 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PINS WERE USED AND AFTER THEY CAME OUT OF THE WASH, THESE WERE MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18552628 ↗
  • MalfunctionJan 19, 2024Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    ATTUNE DRILL STOP HAS BECOME LOOSE AND IS NO LONGER GRIPPING THE DRILL. IT CAUSED NO IMPACT TO SURGERY OR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18548027 ↗
  • MalfunctionSep 22, 2023Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17799041 ↗
  • MalfunctionJun 5, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE RED 3-5 DRILL STOP WON¿T CLICK INTO THE REAMER. SO IT WOULD FALL AS WE REAMED. THEY DISCARDED THE RING. NO DELAYED IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17060061 ↗
  • MalfunctionMar 21, 2023Device-Device Incompatibility; Naturally Worn; Illegible Information; Device-Device Incompatibility; Naturally Worn; Illegible Information

    IT WAS REPORTED THAT THE RED DRILL STOP DOESN'T STAY ON THE DRILL BIT. IT IS NOT BROKEN INTO PIECES, AND NO INSTRUMENT OR INSTRUMENT PARTS WERE LEFT IN THE PATIENT. ALL INSTRUMENTS WERE REMOVED FROM THE PATIENT. NO SURGICAL DELAY. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16582733 ↗
  • MalfunctionFeb 16, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT THE SURGERY. IN THE SURGERY, THE ENGAGEMENT BETWEEN THE ATTUNE TIBIAL DRILL TOWER IN QUESTION AND THE ATTUNE TIBIAL DRILL IN QUESTION WAS NOT GOOD. SO, REAMING OVER THE TOWER WAS NOT POSSIBLE, SO THE SURGEON REAMED MANUALLY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16386866 ↗
  • MalfunctionJan 11, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE 3-5 DEPTH STOP, NO LONGER ATTACHES TO THE TOP OF THE REAMER. THIS WAS SHOWN BY THE SCRUB TECH PRIOR TO THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16137228 ↗
  • MalfunctionJun 22, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRILL STOP WAS NOT SECURELY FITTING ON DRILL. NO PIECES WERE BROKE OFF OF DRILL STOP. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12041771 ↗
  • MalfunctionMay 19, 2021Break; Break

    ATTUNE 3-5 DRILL STOP BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11851910 ↗
  • MalfunctionJul 14, 2020Break; Break

    IT WAS REPORTED THAT THE 3-5 TIBIA DEPTH STOP ATTACHMENT ARM BROKE OFF. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10272285 ↗
  • MalfunctionMay 15, 2020Break; Break

    IT WAS REPORTED THAT THE 3-5 DEPTH STOP CLIP BROKE OFF AND THE FEMUR AND TIBIA IMPACTORS ARE CRACKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10061757 ↗
  • MalfunctionDec 12, 2019Break; Break

    IT WAS REPORTED THAT THE ITEM WAS BROKEN DURING SURGERY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9459781 ↗
  • MalfunctionApr 25, 2018Break; Break

    IT WAS REPORTED THAT THE ATTUNE 3-5 DRILL STOPPER PIECES BROKE IN HALF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7459249 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates